Tandem pumps are battery-operated infusion pumps capable of… recallZ-1636-2026 · Tandem Diabetes Care, Inc. · initiated Jan 30, 2026
Class II Open, Classified All recalls posted in March 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Tandem Diabetes Care, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The Czech language user guide contained multiple translation errors. The most significant error was that it incorrectly instructs users to verify that the infusion set is connected to the body, when it should instead direct users to confirm that the infusion set is not connected to the body. Following this instruction could result in unintended insulin delivery, potentially leading to a hypoglycemic event.
FDA root cause category: Process change control
What the company is doing
On January 30, 2026, Tandem Diabetes Care issued a Urgent Medical Device Recall Notice to affected consignees via E-Mail. Tandem ask consignees to take the following actions: 1. Discard your existing User Guide. 2. Refer to the corrected User Guide version 7.8.1 which is available online at www.tandemdiabetes.com/cs-cz/support/resources/documents for accurate instructions on using your t:slim X2 Insulin Pump. 3. Physical copies of the corrected User Guide are being produced to replace the incorrect versions. If you would like a physical copy of the corrected User Guide, you may contact A. Import s customer service line to arrange for a replacement. 4.As requested above, please complete the online form acknowledging receipt of this notice. 5. Should you experience any device-related serious incidents, please report them to your distributor, a local contact point, or the State Institute for Drug Control.
Recall details
- Recall number
- Z-1636-2026
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Jan 30, 2026
- Posted by the FDA
- Mar 25, 2026
- Recalling firm
- Tandem Diabetes Care, Inc.
- Quantity
- 1,380 units
- Product code
- QFG
- FDA submissions
- K232380, K232382, K243823, K250792, K250798
- Distribution
- International distribution in the countries of Czech Republic, Slovakia.
Product description
Tandem pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion set, into a patient s subcutaneous tissue. As with current insulin infusion pumps on the U.S. market, the desired timing and quantity of the insulin delivery is programmed by the user (i.e., the patient).
Code information (models, lots, serial numbers)
Version 7.8.1/ UDI: (01)008500066132625
Other devices under the same FDA submission
2 devices cleared under K232380, K232382, K243823, K250792, K250798 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Tandem t:slim Mobile Application, Android 1014706Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump2 recalls0 reports
- Pharmacy Kit, Starter Pack, Tandem t:slim X2, Control IQ+ 1017064Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump8 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗