Brand Name: Genius" Review Station recallZ-1949-2026 · Hologic, Inc · initiated Mar 23, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Hologic, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
It was identified that certain devices were being operated outside the validated and FDA-cleared display configuration due to user modification of the manufacturer-installed settings. Modifications included disabling Barco Application Appearance Manager (AAM), altering calibration and compliance parameters, and adjusting additional display-related settings. This field correction is being conducted to restore affected system to the validated configuration and reinforce proper use in accordance with the Operator's Manual.
FDA root cause category: Software design
What the company is doing
On March 23, 2026, Urgent Medical Device Recall letters were sent to customers. Actions to be taken: If you are receiving this notification, our records indicate that you have received a Genius" Review Station Monitor (CMP-01669). As this field correction applies to all monitors, please follow the instructions below and complete the Customer Response Form (CRF) using the link or QR code provided below. If you are a distributor or otherwise supply Genius" Review Station Monitors to downstream customers, you are responsible for promptly forwarding this notification to all affected sub-accounts and ensuring completion of the CRF for each end-user site. Do not change monitor settings " Continue to use the Genius" Review Station only with the original monitor settings configured by Hologic at installation. " Do not: " Disable Barco Application Appearance Manager (AAM) " Change Barco QAWeb calibration or compliance test settings " Adjust or otherwise modify the monitor configuration parameters outside of the original configuration by Hologic at installation. " Only Hologic, or its authorized provider may modify Genius" Review Station monitor configuration settings. If monitor settings have been altered " If you are aware that monitor display or calibration settings have been changed from the original configuration, do not use the Genius" Review Station monitor for clinical review until the monitor has been returned to the original configuration by Hologic or its authorized service provider. " Complete the Customer Response Form (CRF) so Hologic and its field service partner, IQVIA MedTech, can coordinate restoration of the approved configuration. Please promptly fill out the Customer Response Form (CRF) to acknowledge you have received this notification. Distributors should ensure that all downstream customers complete the CRF or provide the required information on their behalf.
Recall details
- Recall number
- Z-1949-2026
- Status
- Open, Classified
- Initiated by the company
- Mar 23, 2026
- Posted by the FDA
- Apr 24, 2026
- Recalling firm
- Hologic, Inc
- Quantity
- 1200 units
- Product code
- QYV
- Distribution
- Worldwide distribution - US Nationwide and the countries of Asia Pacific, Europe, and the Middle East.
Product description
Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: CMP-01669 Software Version: N/A - not software version specific Component: Yes, the Genius" Review Station is a component of the Genius" Digital Diagnostics System with the Genius" Cervical AI Algorithm cleared under DEN210035.
Code information (models, lots, serial numbers)
Model Number: CMP-01669; UDI-DI: 15420045514201; Serial Number: All;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- DIGITAL DIAGNOSTICS SYSTEM DISPLAY, 27IN, 8MPHologic, Inc. · Digital cervical cytology slide imaging system with artificial intelligence algorithm1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.