eTRAX Needle System Starter Kit recallZ-1734-2026 · Civco Medical Instruments Co. Inc. · initiated Mar 2, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Civco Medical Instruments Co. Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
FDA root cause category: Software change control
What the company is doing
An URGENT: MEDICAL DEVICE RECALL notification letter dated 3/2/26 was sent to customers. Actions to be taken by the Customer/User: CIVCO requests that you check your inventory for any eTRAX Needle Sensor for Aurora Trackers and complete the supplied response letter for the appropriate action. Please contact your distribution partner to return any existing inventory, or report its destruction by completing the attached response form. Please complete the enclosed response form even if you do not have any eTRAX Needle Sensor for Aurora Trackers remaining in your inventory to assist in our reconciliation process. Response forms can also be sent via email to [email protected]. For questions, contact CIVCO at # 800-445-6741, or 319-248-6757. Type of Action by the Company: This notice is being communicated to all customers who have purchased an eTRAX Needle Sensor for Aurora Trackers from affected lots. CIVCO requests you provide this notification to the appropriate personnel within your facility. If the affected product was distributed outside of your organization, please notify those locations down to the medical facility level.
Recall details
- Recall number
- Z-1734-2026
- Status
- Open, Classified
- Initiated by the company
- Mar 2, 2026
- Posted by the FDA
- Apr 1, 2026
- Recalling firm
- Civco Medical Instruments Co. Inc.
- Quantity
- 20 units
- Product code
- IYO
- FDA submissions
- K092619
- Distribution
- US States: OH, PA, WA. China.
Product description
eTRAX Needle System Starter Kit 18G (for Aurora Trackers), Part Number 667-152
Code information (models, lots, serial numbers)
UDI-DI: 00841436111065; Lot Number: A233347, A228677, A268116, A268238, A272513, A279457, A280388 , A281840
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- eTRAX™ Needle System Starter Kit 667-152Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
Product lines
Other devices under the same FDA submission
11 devices cleared under K092619 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- VirtuTRAX Instrument Navigator 5350925Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- VirtuTRAX Instrument Navigator 610-1080Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- eTRAX™ Needle System Starter Kit 667-149Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- eTRAX™ Needle System Starter Kit 667-150Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- eTRAX™ Needle System Starter Kit 667-151Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- eTRAX™ Needle Sensor 667-156Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- eTRAX™ Needle Sensor 667-157Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- eTRAX™ Needle Sensor 667-158Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- eTRAX™ Needle Sensor 667-159Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- VirtuTRAX Instrument Navigator 610-1059Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- VirtuTRAX Instrument Navigator 610-1153Civco Medical Instruments CO., Inc. · System, imaging, pulsed echo, ultrasonic1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98513).
- Z-1731-2026 · Open, ClassifiedeTRAX Needle System Starter Kit 12G (for Aurora Trackers), Part Number 667-149
- Z-1732-2026 · Open, ClassifiedeTRAX Needle System Starter Kit 14G (for Aurora Trackers), Part Number 667-150
- Z-1733-2026 · Open, ClassifiedeTRAX Needle System Starter Kit 16G(for Aurora Trackers), Part Number 667-151
- Z-1735-2026 · Open, ClassifiedeTRAX Needle Sensor - 12G (for Aurora Trackers), Part Number 667-156
- Z-1736-2026 · Open, ClassifiedeTRAX Needle Sensor - 14G(for Aurora Trackers), Part Number 667-157
- Z-1737-2026 · Open, ClassifiedeTRAX Needle Sensor - 16G(for Aurora Trackers), Part Number 667-158
- Z-1738-2026 · Open, ClassifiedeTRAX Needle Sensor - 18G (for Aurora Trackers), Part Number 667-159
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page