Skip to content

Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 45cm.… recallZ-1854-2026 · Olympus Corporation of the Americas · initiated Mar 23, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for detachment of a distal tip component of the device during use.

FDA root cause category: Device Design

What the company is doing

Olympus notified consignees on about 03/23/2026 via emailed letter. Consignees were instructed to immediately cease use of all THUNDERBEAT II devices, examine inventory for affected product and quarantine any on hand, and complete, return the Reply Form provided and arrange for the return of all affected units in inventory. Consignees were also instructed to notify other users if the affected products were transferred or further distributed.

Recall details

Recall number
Z-1854-2026
Status
Open, Classified
Initiated by the company
Mar 23, 2026
Posted by the FDA
Apr 17, 2026
Recalling firm
Olympus Corporation of the Americas
Quantity
490 units
Product code
GEI
FDA submissions
K252150
Distribution
International distribution in the countries of Australia, Japan, Hong Kong, and Europe.

Product description

Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 45cm. Model Number: TB2-0545FC.

Code information (models, lots, serial numbers)

Model Number: TB2-0545FC. UDI-DI: 04953170440021. All Lot Numbers.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

3 devices cleared under K252150 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 98550).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page