SIGNA Premier systems recallZ-1767-2026 · GE Healthcare LLC · initiated Feb 27, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Healthcare LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Whole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous (brass) fittings on the magnet rear and if system is accessed to perform service while the magnet is ramped, the magnetic field could attract the ferrous fittings potentially resulting in personnel injury.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On 2/27/2026, correction notices were mailed to Director of Clinical/Radiology Risk Manager/Hospital Administrators informing them to take the following actions: 1) Ensure all potential servicing personnel in your facility are made aware of this safety notification and the recommended actions. 2) If the device requires service at the rear of the magnet, ensure that all servicing personnel are aware of this potential issue. 3) If servicing personnel have accessed the rear of the magnet with covers removed and identified that ferrous fittings are installed, do NOT attempt to remove or disconnect them. Call your GE HealthCare representative immediately. To help identify the parts: -The incorrect fittings (ferrous) are steel/silver in color -The correct fittings (non-ferrous) are brass/gold in color 4) Complete and return the acknowledgement form electronically via (https://gehealthcare-svc.my.site.com/publicForm/s/?formId=aGjUr000002kiJR) or print, fill out manually, scan and email to [email protected] If you have any questions or concerns regarding this notification, please contact the firm's HealthCare Service at 1-800-437-1171 or your local Service Representative.
Recall details
- Recall number
- Z-1767-2026
- Status
- Open, Classified
- Initiated by the company
- Feb 27, 2026
- Posted by the FDA
- Apr 2, 2026
- Recalling firm
- GE Healthcare LLC
- Quantity
- 41
- Product code
- LNH
- FDA submissions
- K171128, K183231, K193282
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, MA, NJ, NY, OH, SD, TX, VA, PR and the countries of Sweden, Switzerland, United Kingdom, Vietnam, China, Norway, Portugal, Spain, France, Germany, Greece, Korea, Republic of
Product description
SIGNA Premier systems
Code information (models, lots, serial numbers)
UDI-DI/GTIN: 00840682135269, 00195278010797. System ID: 415353UMR2, 415723SHMR19, 413253MR3T, 908725MR, 609652MR3T, M48823, 7168453T, 513584MR5, 513636EASTMR, 513636T1MR, 605333MRPRM, 832522CRMR1, 703858PRE2, 804675VA3T, 082427110162, 082427220143, 082427020147, 082427200242, 082427160461, 082427250139, 082427140382, M4216469, 353745MR01, GON5655515, KRMR56474971, KRMR56203791, KRMR55433661, KRMR56344161, KRMR55774051, KRMR55914661, NOMRI26833321, 15974045, GON1882380, ESMRI27056983, ESMRI26960129, A203MR06, 690120MR01, 11524261, VNMRI26792525, VNMRI26592248, VNMRI26809677
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- NA SIGNA PremierGE Medical Systems, L.L.C. · Coil, magnetic resonance, specialty3 recalls0 reports
- NA SIGNA PremierGE Medical Systems, L.L.C. · Coil, magnetic resonance, specialty3 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.