SIGNA Premier systems recallZ-1767-2026 · GE Healthcare LLC · initiated Feb 27, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Healthcare LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Whole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous (brass) fittings on the magnet rear and if system is accessed to perform service while the magnet is ramped, the magnetic field could attract the ferrous fittings potentially resulting in personnel injury.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On 2/27/2026, correction notices were mailed to Director of Clinical/Radiology Risk Manager/Hospital Administrators informing them to take the following actions: 1) Ensure all potential servicing personnel in your facility are made aware of this safety notification and the recommended actions. 2) If the device requires service at the rear of the magnet, ensure that all servicing personnel are aware of this potential issue. 3) If servicing personnel have accessed the rear of the magnet with covers removed and identified that ferrous fittings are installed, do NOT attempt to remove or disconnect them. Call your GE HealthCare representative immediately. To help identify the parts: -The incorrect fittings (ferrous) are steel/silver in color -The correct fittings (non-ferrous) are brass/gold in color 4) Complete and return the acknowledgement form electronically via (https://gehealthcare-svc.my.site.com/publicForm/s/?formId=aGjUr000002kiJR) or print, fill out manually, scan and email to [email protected] If you have any questions or concerns regarding this notification, please contact the firm's HealthCare Service at 1-800-437-1171 or your local Service Representative.

Recall details

Recall number
Z-1767-2026
Status
Open, Classified
Initiated by the company
Feb 27, 2026
Posted by the FDA
Apr 2, 2026
Recalling firm
GE Healthcare LLC
Quantity
41
Product code
LNH
FDA submissions
K171128, K183231, K193282
Distribution
Worldwide - US Nationwide distribution in the states of CA, MA, NJ, NY, OH, SD, TX, VA, PR and the countries of Sweden, Switzerland, United Kingdom, Vietnam, China, Norway, Portugal, Spain, France, Germany, Greece, Korea, Republic of

Product description

SIGNA Premier systems

Code information (models, lots, serial numbers)

UDI-DI/GTIN: 00840682135269, 00195278010797. System ID: 415353UMR2, 415723SHMR19, 413253MR3T, 908725MR, 609652MR3T, M48823, 7168453T, 513584MR5, 513636EASTMR, 513636T1MR, 605333MRPRM, 832522CRMR1, 703858PRE2, 804675VA3T, 082427110162, 082427220143, 082427020147, 082427200242, 082427160461, 082427250139, 082427140382, M4216469, 353745MR01, GON5655515, KRMR56474971, KRMR56203791, KRMR55433661, KRMR56344161, KRMR55774051, KRMR55914661, NOMRI26833321, 15974045, GON1882380, ESMRI27056983, ESMRI26960129, A203MR06, 690120MR01, 11524261, VNMRI26792525, VNMRI26592248, VNMRI26809677