Medline Convenience Kits containing select SKUs of 10mL… recallZ-1881-2026 · Medline Industries, LP · initiated Feb 27, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Unapproved design changes to the products outside of the 510(k) clearance.

FDA root cause category: Component change control

What the company is doing

On February 27, 2026, the firm began notifying customers via email and first-class mail. Customers were instructed to request over-labels to place on affected kits containing recalled syringes instructing them to remove the recalled component and replace with product from their supply.

Recall details

Recall number
Z-1881-2026
Status
Open, Classified
Initiated by the company
Feb 27, 2026
Posted by the FDA
Apr 17, 2026
Recalling firm
Medline Industries, LP
Quantity
270,311 total
Product code
FDE
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Panama, Barbados.

Product description

Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes BARIATRIC DYNJ905153M, DYNJ905153N GENERAL LAPAROSCOPY PACK DYNJ81602 LAPAROSCOPY DYNJ905157K, DYNJ905157L LAPAROTOMY PACK DYNJ46675G MAJOR LITHOTOMY PACK DYNJ81607 MAJOR PLUS PACK DYNJ905160O MINOR GENERAL PACK DYNJ81599

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

DYNJ905153M UDI-DI 10195327556761 (EA) 40195327556762 (CS) LOT 24CMG485 DYNJ905153N UDI-DI 10198459017797 (EA) 40198459017798 (CS) LOT 25DMC254 DYNJ81602 UDI-DI 10195327123857 (EA) 40195327123858 (CS) LOT 24ABP080 DYNJ905157K UDI-DI 10195327556921 (EA) 40195327556922 (CS) LOTS 24DMF349 24EMC297 DYNJ905157L UDI-DI 10198459017711 (EA) 40198459017712 (CS) LOTS 24FMD083 24HMD400 24IME332 24JMA234 24KMG588 25AMA393 25AMJ541 25BMI202 25CMD909 25CMH371 DYNJ46675G UDI-DI 10195327419066 (EA) 40195327419067 (CS) LOT 24ABS814 DYNJ81607 UDI-DI 10195327123918 (EA) 40195327123919 (CS) LOT 24ABP097 DYNJ905160O UDI-DI 10198459017742 (EA) 40198459017743 (CS) LOTS 24HMG093 24IME714 24JMI924 24LMB556 25AME613 25BMJ076 DYNJ81599 UDI-DI 10195327124021 (EA) 40195327124022 (CS) LOT 24BBB847 ***Updated 5/20/26 - The following lots were included in the customer letter but not in the FDA submission.*** DYNJ905153M lots 24BMH968 24EMG682; DYNJ905153N lots 24HMG087 24IMA442 24IMG245; DYNJ905157K lots 23LME674 24AMH261 24AMJ205 24BME886 24CMD057

Devices named in this recall

9 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 98601).