Gentuity HF-OCT Imaging System Model G10-01 recallZ-1823-2026 · Gentuity, LLC · initiated Mar 23, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Gentuity, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The potential for repeated frames during HF-OCT pullback when using the Gentuity HF-OCT Console (software version 21.11 to 23.3.13) with the Vis-Rx and Vis-Rx Prime catheters. This correction is intended to ensure that users are fully informed of how to recognize, avoid, and mitigate this condition to maintain accurate longitudinal (length) measurements during OCT imaging.
FDA root cause category: Device Design
What the company is doing
On March 23, 2026, URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: The customers will be required to follow the updated instructions to correct for repeated frames per the revised Operator Manual to address longitudinal measurement accuracy until the updated version of the software that addresses this issue is installed on the console within 6 months.
Recall details
- Recall number
- Z-1823-2026
- Status
- Open, Classified
- Initiated by the company
- Mar 23, 2026
- Posted by the FDA
- Apr 16, 2026
- Recalling firm
- Gentuity, LLC
- Quantity
- 184 units (28 US, 156 OUS)
- Product code
- DQO
- FDA submissions
- K242966
- Distribution
- Domestic - US Nationwide distribution in the states of CA, FL, GA, IL, MA, MN, NJ, NY, OH, PA, TX; International: United Arab Emirates, Belgium, Switzerland, Cyprus, Denmark, Spain, Finland, France, Greece, Italy, Japan, Kuwait, Latvia, Poland, Saudi Arabia, Slovakia;
Product description
Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13;
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
Model Number: G10-01; UDI-DI: 00859910007032; Lot/Serial Number: 23C0113, 22J0101, 20A0202, 23F0101, 22F0113, 22L0100, 26B0203, 26B0200, 26B0204, 26B0202, 26B0201, 25L0302, 22E0204, 22J0102, 24H0101, 24H0102, 23K0201, 20A0203, 22K0306, 20A0201, 25J0200, 23C0110, 22K0305, 25E0100, 22F0110, 22K0304, 22K0302, 24B0107; OUS Lot numbers: 21G1900, 21G1901, 21G1902, 21G1903, 21G1904, 21H1101, 21H1102, 21H1103, 21H1104, 21H1105, 22C2300, 22C2301, 22C3000, 22D2600, 22D2700, 22D2701, 22D2702, 22D2705, 22E0500, 22E0501, 22F2700, 22F2701, 22F2702, 22F2703, 22F2704, 22G2600, 22G2601, 22G2700, 22H2900, 22H2901, 24B2100, 24B2101, 24B2102, 24B2103, 24B2104, 24C1301, 24C1302, 24C1303, 24C1304, 24C1305, 24D1504, 24D1505, 24D1506, 24D1507, 24D1508, 24E0800, 24E0801, 24E0802, 24E0803, 24E0804, 24F2408, 24F2409, 24F2410, 24F2411, 24F2412, 24G2205, 24G2206, 24G2207, 24G2208, 24G2209, 24H1400, 24H1401, 24H1402, 24H1403, 24H1404, 24M1000, 22L3000, 22L3001, 23A1904, PDA23-001, 23C2202, 23C2203, 23E0900, 23E0901, 23E0902, 23F2704, 23F2705, 23G2601, 23G2602, 23G2603, 23G2604, 23G0700, 23G0701, 23H3000, 23H3001, 23H3002, 23H3003, 23M1102, 23M1103, 23M1104, 23M1301, 24B2105, 24B2106, 24C2200, 24C2201, 24H1900, 24H1901, 24H1902, 24J0500, 24J0501, 24J0502, 24M1101, 24M1102, 24M1103, 24M1104, 24M1105, 25C0320, 25C0321, 25C0322, 25C0323, 25C0324, 25F2500, 25F2501, 25F2502, 25G3103, 25G3104, 25G3105, 25J0800, 25J0801, 25J0802, 25J2300, 25K0909, 25K0910, 25L0306, 25L0307, 25M0203, 25M0201, 25M0202, 26B0310, 26B0311, 26B2300, 26B1900, 26B1901, 26B1902;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Gentuity HF-OCT Imaging System SW 23.3.13Gentuity LLC · System, Imaging, Optical Coherence Tomography (Oct)1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.