ARROW recallZ-1917-2026 · Arrow International, LLC · initiated Mar 11, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Arrow International, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to receiving a notice from a supplier that the liquid adhesive was incorrectly manufactured
FDA root cause category: Process control
What the company is doing
On 03/18/2026, the firm sent via FedEx-2-day an "URGENT MEDICAL DEVICE RECALL" Letter informing customers that Ferndale Laboratories issued a Field Safety Notice (FSN) to inform users of multiple complaints reporting the butyrate tubing of the Mastisol Liquid Adhesive vial cracked during actuation. According to the Ferndale Laboratories FSN, one lot of MASTISOL Liquid Adhesive was exposed to a double sterilization cycle which may have contributed to the butyrate tubing becoming brittle and susceptible to cracking during actuation. Medical Facilities should: 1. Immediately check inventory for product within the scope of this recall and cease use and distribution of affected product and immediately quarantine the affected product. 2. For impacted product, mark the applicable checkbox on the Acknowledgement Form (Appendix 1) and contact Teleflex Customer Service utilizing one of the following methods: a. Fax, including Attn: Customer Service , or email customer service using the point of contact details outlined on Page 3 with a copy of Appendix 1 OR b. Using the online FSN Portal, complete the acknowledgement form virtually using the following link: https://fca.myteleflex.com/en/recall/000606 Note: Option B is solely applicable to customers who received product direct from Teleflex. 3. For impacted product, mark the applicable checkbox on the Acknowledgement Form (Appendix 1) and return the form to Teleflex at the contact details provided. 4. Teleflex (or your local dealer) will issue a credit note upon receipt of the returned affected product. Distributor should: 1. Provide a copy of this recall notice to all customers who have received impacted product. Each customers is then required to complete the Acknowledgement Form and return it to Distributor. 2. Request immediate check of inventory, cease use and distribution of, and quarantine, affected products and returned to Teleflex. For questions-contact customer service at 1-866-396-2111 or email Recalls@telefle
Recall details
- Recall number
- Z-1917-2026
- Status
- Open, Classified
- Initiated by the company
- Mar 11, 2026
- Posted by the FDA
- Apr 22, 2026
- Recalling firm
- Arrow International, LLC
- Quantity
- 95
- Product code
- OFC
- Distribution
- U.S. Nationwide distribution in the states of AZ, CA, GA, MA, ME, MI, OR, PA, and VA.
Product description
Arrowg+ard Blue MAC" Two-Lumen Central Venous Access Kit for use with 7.5 - 8 Fr. Catheters REF ASK-21142-DMC UDI code: (01)10801902218930(17)270630(11)251211(10)33F25M0671 The MAC Multi-Lumen Central Venous Access Device with Arrowg+ard Blue technology permits short-term (<30 days) venous access and catheter introduction to the central circulation. It may be inserted into the jugular, subclavian, or femoral veins. The Arrowg+ard technology is intended to help provide protection against catheterrelated infections. Clinical data have not been collected that demonstrate the use of the Arrowg+ard antimicrobial surface in decreasing catheter-related infections for this device. It is not intended to be used as a treatment for existing infections, nor is it indicated for long-term use.
Code information (models, lots, serial numbers)
Product Code: ASK-21142-DMC UDI code: (01)10801902218930(17)270630(11)251211(10)33F25M0671 Batch Number: 33F25M0671
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ARROW IPN929229Teleflex Incorporated · Introducer, catheter2 recalls0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 98618).
- Z-1910-2026 · Open, ClassifiedArrow Echogenic Introducer Needle REF ASK-04001-BWK2 UDI code: (01)10801902220315(17)261231(11)251210(10)33F25J0347 This device is intended for injection or aspiration of fluids. …
- Z-1911-2026 · Open, ClassifiedARROW RADIAL ARTERY CATHETERIZATION KIT REF ASK-04220-UCL1 UDI code: (01)10801902159127(17)270331(11)251203(10)33F25J0352 REF ASK-04020-EUH2 UDI code: (01)10801902213379(17)27073…
- Z-1912-2026 · Open, ClassifiedPERCUTANEOUS SHEATH INTRODUCER with 7.5 - 8 Fr. Catheter REF ASK-09903-MGH3 UDI code: (01)10801902159127(17)270331(11)251203(10)33F25J0352 The Arrow percutaneous sheath introducer…
- Z-1913-2026 · Open, ClassifiedFLEXTIP PLUS EPIDURAL CATHETER;FLEXBLOCK CONTINOUS PERIPHERAL NERVE BLOCK CATHETER KIT/SET REF ASK-05502-BSM UDI code: (01)10801902209853(17)270831(11)251210(10)33F25K0409 The Arr…
- Z-1914-2026 · Open, ClassifiedARROW, ARROWG+ARD AND ARROWG+ARD BLUE PLUS PRESSURE INJECTABLE CENTRAL VENOUS CATHETER REF ASK-42802-PLH UDI code: (01)10801902133110(17)261231(11)251211(10)33F25L0140 REF ASK-4…
- Z-1915-2026 · Open, ClassifiedPressure Injectable Arrowg+ard Blue Plus Four-Lumen CVC Kit REF ASK-45854-PUPM1 UDI code: (01)10801902159578(17)270531(11)251205(10)33F25M0011 The Arrowg+ard technology is intende…
- Z-1916-2026 · Open, ClassifiedMAC" Two-Lumen Central Venous Access Kit for use with 7.5 - 8 Fr. Catheters REF ASK-11142-LC2 UDI code: (01)10801902206807(17)270531(11)251211(10)33F25M0670 The MAC Multi-Lumen Ce…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page