Skip to content

Viper recallZ-1938-2026 · Diversatek Healthcare · initiated Mar 18, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Diversatek Healthcare or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The catheter inflation tag attached to the device may contain an incorrect part number with incorrect balloon diameter sizes and inflation pressures for each stage.

FDA root cause category: Labeling Change Control

What the company is doing

Diversatek Healthcare issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 3/18/2026 via USPS. The notice explained the problem, risk to health, and requested the consignee discard any affected product. Complete and return the attached MEDICAL DEVICE RECALL RETURN RESPONSE Acknowledgement and Receipt Form and provide Signature Confirming Customer Receipt to Laura Boll, VP Regulatory Affairs & Quality at [email protected] or 414-755-4806 to receive a replacement order. For questions about this issue, please contact Laura Boll, Vice President - Regulatory Affairs & Quality, at 414-755-4806 or [email protected].

Recall details

Recall number
Z-1938-2026
Status
Open, Classified
Initiated by the company
Mar 18, 2026
Posted by the FDA
Apr 23, 2026
Recalling firm
Diversatek Healthcare
Quantity
15 units
Product code
KNQ
Distribution
US: MO, CA, FL, TX, AZ, PA, RI, NC, IL, MS, KS, NY, OH, VA, LA, IA, MI, NJ, AL, Brazil, Belgium

Product description

Diversatek Healthcare Viper 3-Stage Fixed Wire Balloon Dilator, 6mm-7mm-8mm, REF 1204-08

Code information (models, lots, serial numbers)

UDI/DI 00816734021699, Lot Numbers: 005983

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 98649).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page