BD GasPak EZ Campy Pouch System. Catalog Number: 260685.… recallZ-1983-2026 · Becton Dickinson & Co. · initiated Apr 7, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Becton Dickinson & Co. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
BD identified an issue affecting certain lots of BD GasPak EZ CO2 and BD GasPak EZ Campy Gas Generating Pouch Systems, where gas generating sachets may produce carbon dioxide levels below specification.
FDA root cause category: Under Investigation by firm
What the company is doing
On 04/07/2026, BD sent URGENT: Medical Device Correction letters to customers and distributors. Consignees were instructed that there are no recommendations for retrospective chart review or repeat patient testing. They are to verify incubation atmospheres by incorporating the use of appropriate gas indicators as part of routine laboratory quality control practices. Each BD GasPak EZ Gas Generating Pouch System should be tested periodically for its ability to provide adequate conditions for the growth of target organisms. For CO2 incubation, laboratories may use BD CO2 Indicators (Catalog No. 271055), along with the impacted catalog 260684, to verify sufficient CO2 presence in the environment. If conditions are not confirmed or are suspected to be out of specification, laboratories should repeat incubation using alternative validated methods or established laboratory procedures to ensure organism recovery. Additionally, consignees were instructed to ensure the contents of the notification are read and understood, share this notification with the facility network and forward it to any customers to whom product may have been distributed, and complete and return the Customer Response Form. Distributors were instructed to ensure the notification is read and understood, identify all customers that purchased any affected product, and provide a copy of the attached customer letter to all customers who received the affected lots to advise them of this Product Advisory on BD's behalf. BD is investigating the root cause and will implement appropriate corrective actions to prevent recurrence of this issue. BD GasPak catalog numbers 260684 and 260685 are currently on backorder due to seal integrity challenges identified in some batches following a recent overhaul of supplier manufacturing equipment. BD anticipates resumption of product shipments by the end of March 2026, with full recovery expected by June 2026.
Recall details
- Recall number
- Z-1983-2026
- Status
- Open, Classified
- Initiated by the company
- Apr 7, 2026
- Posted by the FDA
- Apr 29, 2026
- Recalling firm
- Becton Dickinson & Co.
- Quantity
- 2,200 units
- Product code
- KZJ
- Distribution
- Worldwide distribution - US Nationwide and the country of Hong Kong.
Product description
BD GasPak EZ Campy Pouch System. Catalog Number: 260685. for microbiologic use
Code information (models, lots, serial numbers)
Catalog Number: 260685; UDI-DI: 30382902606858; Lot Number: 5329863.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- BD GasPak™ EZ Campy Gas Generating Pouch System 260685Becton, Dickinson And Company · DEVICE, GAS GENERATING1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98657).
- Z-1982-2026 · Open, ClassifiedBD GasPak EZ CO2 Pouch System. Catalog Number: 260684. for microbiologic use
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page