epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis… recallZ-2005-2026 · Siemens Healthcare Diagnostics Inc · initiated Apr 2, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.
FDA root cause category: Under Investigation by firm
What the company is doing
On April 2, 2026, Urgent Medical Device Recall letters were sent to customers. Actions to be taken: Perform the instructions provided below: " Discontinue use of and discard the kit lots listed in the table above (Products Section). " Please review this letter with your Medical Director/POC Coordinator to determine the appropriate course of action, including for any previously generated results, if applicable. " Complete and return the Field Correction Effectiveness Check Form and Product Replacement Form attached to this letter within 30 days. " Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax. " Please retain this letter with your laboratory records and forward this letter to those who may have received this product. " For any questions, you may call Siemens Healthineers Technical Solutions Center at 877- 229-3711, available daily from 7:00 AM to 11:00 PM EST.
Recall details
- Recall number
- Z-2005-2026
- Status
- Open, Classified
- Initiated by the company
- Apr 2, 2026
- Posted by the FDA
- Apr 29, 2026
- Recalling firm
- Siemens Healthcare Diagnostics Inc
- Quantity
- 71725 (38025 US, 33700 OUS)
- Product code
- CHL
- FDA submissions
- K200107
- Distribution
- Worldwide distribution - US Nationwide and the countries of Italy, Spain, Austria, Portugal, Ireland, France, Romania, Canada, India, Philippines, Mexico.
Product description
epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;
Code information (models, lots, serial numbers)
Siemens Material Number (SMN): 10736515; UDI-DI: 00809708121860; Lot numbers: 01-26024-70, 01-26026-70, 01-26028-70, 01-26030-70 (OUS only), 02-26027-80, 02-26028-80 (OUS only);
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- EPOC BGEM BUN TEST CARD 10736515Epocal Inc6 recalls1 report
Other devices under the same FDA submission
1 device cleared under K200107 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- epoc NXS Host 11413475Epocal Inc · ACID, LACTIC, ENZYMATIC METHOD2 recalls6 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page