Halyard PERC TRAY kit. Model Numbers: SLPC34-01, SLPC34-02 recallZ-1977-2026 · AVID Medical, Inc. · initiated Mar 13, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from AVID Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

FDA root cause category: Under Investigation by firm

What the company is doing

Owens and Minor/Avid Medical notified consignees on about 03/13/2026 and then sent an updated letter on about 03/18/2026 with additional kits added to the affected product list. The notice explained the issue was initiated by Medline Industries, risk associated with syringe disconnection and loose connection, and instructed consignees to identify, segregate and quarantine all affected product, and to apply warning labels to affected AVID kits to ensure all users are aware of the actions requested by Medline. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. Per Medline, due to the risk of serious injury or death, customers are directed to remove and destroy all Namic RA Syringes. If use is unavoidable because failure to proceed would result in patient harm, the Namic RA syringe must be used with extreme caution and vigilance, including manual stabilization of syringe-to-manifold connections and continuous, 100% monitoring during setup and throughout the entire procedure. All usage guidelines and the Instructions for Use are provided in Medline s recall letter. Consignees were also instructed to complete and return the Recall Response Form as well as to notify customers if affected kits were further distributed to ensure all end-users are appropriately notified and maintain records of effectiveness.

Recall details

Recall number
Z-1977-2026
Status
Open, Classified
Initiated by the company
Mar 13, 2026
Posted by the FDA
May 1, 2026
Recalling firm
AVID Medical, Inc.
Quantity
6,052 kits
Product code
LRO
Distribution
US distribution to FL, IL, MO, MS, NC, NE, TX.

Product description

Halyard PERC TRAY kit. Model Numbers: SLPC34-01, SLPC34-02.

Code information (models, lots, serial numbers)

1. Model Number: SLPC34-01. UDI-DI (Kit Lot Numbers): 10809160450963 (1640169, 1641409, 1641410, 1641411, 1641412, 1641413, 1641414, 1641415, 1645684); 10809160459904 (1649576, 1660882, 1660883, 1660884, 1663896, 1665768). 2. Model Number: SLPC34-02. UDI-DI (Kit Lot Numbers): DHR 108916049515 NAV 10809160511213 (1668809, 1668810, 1668812, 1668813, 1668814, 1668815, 1668816).

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 98686).