Skip to content

Straumann n!ce Zr, HT, Full-arch Restoration. Article… recallZ-1995-2026 · Straumann USA LLC · initiated Mar 27, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Straumann USA LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Includes an incorrect screw seat interface.

FDA root cause category: Under Investigation by firm

What the company is doing

Straumann USA notified consignees on about 03/27/2026 via telephone call, mailed letter, or email. Consignees were instructed to examine inventory for affected product, cease use and quarantine any affected units on hand, arrange for the return of all affected units and complete and return the provided Customer Confirmation Form. Consignees were also instructed to no contact Straumann contact Jennifer Jackson to organize a remake of the units, if any were used. Once the new restoration is available, schedule the patient to replace the restoration.

Recall details

Recall number
Z-1995-2026
Status
Open, Classified
Initiated by the company
Mar 27, 2026
Posted by the FDA
Apr 29, 2026
Recalling firm
Straumann USA LLC
Quantity
6 units
Product code
NHA
FDA submissions
K222836
Distribution
US distribution to AL, AZ, FL, HI, LA, MD, OH, VA.

Product description

Straumann n!ce Zr, HT, Full-arch Restoration. Article 010.0158. Endosseous Dental Implant Abutments

Code information (models, lots, serial numbers)

Article 010.0158. UDI: (01)07630031759764(11)YYMMDD(10)XXXXX. Project Numbers: 28000US_00622_00_b00, 28000US_00431_00_b00, 28000US_00489_00_b00, 28000US_00792_00_b00, 28000US_00908_00_b00, 28000US_00908_01_b00.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 98681).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page