Geneva recallZ-2712-2026 · Elekta, Inc. · initiated Apr 3, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Elekta, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to safety concerns regarding the detachment of the tip of the intrauterine tube (IU tube) that may remain within a patient.
FDA root cause category: Process control
What the company is doing
On 04/03/2026, an "IMPORTANT FIELD SAFETY NOTIFICATION/URGENT: MEDICAL DEVICE CORRECTION" Letter was sent via email informing customers of a safety concern regarding detachment of the tip of an Intrauterine tube (IU tube) with a 4 mm diameter. Customers are instructed to: To prevent any potential health risks related to a detached tip, we emphasize the importance of adhering to the Instructions for Use: " Do not use an IU tube beyond its 3-year life expectancy or 300 reprocessing cycles, whichever comes first. " During the treatment preparation phase, before insertion of the applicator: Visually inspect parts for indication of damage, weakening, bending, cracks, corrosion, or general deterioration. " Inspect joints and connections on play, strength or other forms of degradation. In addition, we recommend the following: " Inspect the end of the IU tube to ensure that there is no gap between the tip and the fiber tube and that the tip is not detached. " During inspection, insertion and removal: avoid the use of excessive force that could damage the IU tube. " After removal: Ensure that all applicator parts have been removed from the patient and that the tip is not detached. In case you find a damaged or detached IU tube tip, remove it from clinical use and contact Elekta for a replacement. For Questions- Region America [email protected] Tel: +1 855 6935358 https://www.elekta.com Region Asia Pacific [email protected] Tel: + 852 2891 2208 https://www.elekta.com Japan [email protected] Tel: + 81 3 6722 3800 Fax: +81 3 6436 4231 https://www.elekta.com Region China [email protected] Tel : + 86 800 810 2550 https://www.elekta.com Region Europe [email protected] Tel : + 46 8 587 254 00 https://www.elekta.com Region Turkey, India & Middle East [email protected] Turkey Tel: +90 216 444 6374 India Tel: +1-800-103-7454 Middle East Tel: +00 800 4000 5000 https://www.elekta.com
Recall details
- Recall number
- Z-2712-2026
- Status
- Open, Classified
- Initiated by the company
- Apr 3, 2026
- Posted by the FDA
- Jul 9, 2026
- Recalling firm
- Elekta, Inc.
- Quantity
- 1317
- Product code
- JAQ
- FDA submissions
- K201272
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TX, UT, VA, WI, and WV. The countries of Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Botswana, Brazil, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran (Islam Republic of), Iraq, Ireland, Italy, Japan, Kazakhstan, Kenya, Korea (Republic of), Kosovo, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Rwanda, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Togo, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Viet Nam, and Zambia.
Product description
Brand Name: Geneva Product Name: Brachytherapy Applicator Product Description: The device is intended for intracavitary brachytherapy for cancer treatment of the cervix and the endometrium. Optional needles can be placed for interstitial therapy.
Code information (models, lots, serial numbers)
Lot Code: 08717213056039, 08717213056053, 08717213056077, 08717213056091, 08717213056398, 08717213056411, 08717213056435, 08717213056459, 08717213056473, 08717213056497, 08717213056510,08717213056534, 08717213056558, 08717213056572, 08717213056596, 08717213056619, 08717213056633, 08717213056657
Devices named in this recall
17 devices in the FDA device registry whose identifier appears in the recall's code information.
- Geneva 152007-00Nucletron B.V. · System, applicator, radionuclide, remote-controlled1 recall0 reports
- Geneva 152008-00Nucletron B.V. · System, applicator, radionuclide, remote-controlled1 recall0 reports
- Geneva 152009-00Nucletron B.V. · System, applicator, radionuclide, remote-controlled1 recall0 reports
- Geneva 152010-00Nucletron B.V. · System, applicator, radionuclide, remote-controlled1 recall0 reports
- Geneva 152011-00Nucletron B.V. · System, applicator, radionuclide, remote-controlled1 recall0 reports
- Geneva 152012-00Nucletron B.V. · System, applicator, radionuclide, remote-controlled1 recall0 reports
- Geneva 152013-00Nucletron B.V. · System, applicator, radionuclide, remote-controlled1 recall0 reports
- Geneva 152014-00Nucletron B.V. · System, applicator, radionuclide, remote-controlled1 recall0 reports
- Geneva 152015-00Nucletron B.V. · System, applicator, radionuclide, remote-controlled1 recall0 reports
- Geneva 152016-00Nucletron B.V. · System, applicator, radionuclide, remote-controlled1 recall0 reports
- Geneva 152017-00Nucletron B.V. · System, applicator, radionuclide, remote-controlled1 recall0 reports
- Geneva 152019-00Nucletron B.V. · System, applicator, radionuclide, remote-controlled1 recall0 reports
- Geneva 152020-00Nucletron B.V. · System, applicator, radionuclide, remote-controlled1 recall0 reports
- Geneva 152021-00Nucletron B.V. · System, applicator, radionuclide, remote-controlled1 recall0 reports
- Geneva 152022-00Nucletron B.V. · System, applicator, radionuclide, remote-controlled1 recall0 reports
- Geneva 152023-00Nucletron B.V. · System, applicator, radionuclide, remote-controlled1 recall0 reports
- Geneva 152024-00Nucletron B.V. · System, applicator, radionuclide, remote-controlled1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 98704).
- Z-2710-2026 · Open, ClassifiedBrand Name: Fletcher CT/MR Shielded Applicator Set Product Name: Intracavitary Remote Afterloading Applicator Product Description: The Fletcher CT/MR Shielded Applicator Set is int…
- Z-2711-2026 · Open, ClassifiedBrand Name: Brachytherapy Applicator Product Name: Advanced Gynecological Applicator Product Description: The intended use of the Advanced Gynecological Applicator is intracavitary…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page