Convenience kits containing Medline Namic Angiographic… recallZ-1981-2026 · American Contract Systems Inc · initiated Mar 19, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from American Contract Systems Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential risk for biohazard exposure, blood loss, infection, or air embolism.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Firm notified distributors with an "URGENT: MEDICAL DEVICE RECALL NOTIFICATION" dated March 19, 2026, via email. Distributors were instructed to notify end-users. An updated letter with additional kits was sent to the customer on 3/20/26. Customers should quarantine all affected product. Customers may add warning labels to affected ACS kits to ensure all users are aware of prescribed actions that were identified in the Medline recall notice. The warning labels state, "Per Medline, due to the risk of serious injury or death, customers are directed to remove and destroy all Namic RA Syringes. If use is unavoidable because failure to proceed would result in patient harm, the Namic RA syringe must be used with extreme caution and vigilance, including manual stabilization of syringe-to-manifold connections and continuous, 100% monitoring during setup and throughout the entire procedure." Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. If you have any questions, contact the Regulatory Affairs Manager, by email: [email protected] or at 419-455-2090.

Recall details

Recall number
Z-1981-2026
Status
Open, Classified
Initiated by the company
Mar 19, 2026
Posted by the FDA
May 6, 2026
Recalling firm
American Contract Systems Inc
Quantity
7,311 kits
Product code
LRO
Distribution
US distribution to MN. No OUS distribution.

Product description

Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers ANCA80AP ANCA80AQ

Code information (models, lots, serial numbers)

ANCA80AP UDI-DI 191072232168 Lot 42-8146111 ANCA80AQ UDI-DI 191072236678 Lots 42-8207211 42-8267711 42-8311311 42-8337611 42-8353311 42-8379211 42-8407111 42-8510711 42-8575111 42-8611511 42-8718911 42-8718912 42-8722611 42-8805311 42-8820611 42-8870911

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.