Mint Lesion recallZ-2068-2026 · Mint Medical GmbH · initiated Jan 7, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Mint Medical GmbH or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
If the connection between a mint Lesion workstation and the mint Lesion server is interrupted while performing a read, in rare cases some information may be lost or incorrectly linked to other patients.
FDA root cause category: Under Investigation by firm
What the company is doing
On or about January 7, 2026 Urgent Field Safety Notice letters were emailed to customers. Customers were instructed to read the notice and check if any affected product is available for use. If that is the case, the malfunction may have occurred and may occur in your system. Please be aware that the malfunction may occur. Your IT department must ensure that no temporary network interruptions occur in your clinical network. Also, your IT department must ensure that the mint Lesion server is not restarted while users work with mint Lesion application (i.e., the server may be restartet, but only outside of working hours). Please verify correctness of the information of prior time-points. Be attentive to potential inconsistencies in lesion attributes when opening existing cases for follow up. Cross check lesion location and categorization with prior radiological reports and image data. If you suspect that the problem may have occurred on your mint Lesion system, please contact Mint Medical support via [email protected] or any other contact details for technical analysis and investigation. Mint Medical Support will analyze if the malfunction appeared at your mint Lesion system in the past. If you believe that this failure has occurred, please review the potentially affected radiological reports in your reporting application and take the necessary steps to correct them. Pass this notice on to all users of mint Lesion" within your organization. Please maintain awareness of this notice and resulting action for an appropriate period to ensure the effectiveness of the corrective action.
Recall details
- Recall number
- Z-2068-2026
- Status
- Open, Classified
- Initiated by the company
- Jan 7, 2026
- Posted by the FDA
- May 6, 2026
- Recalling firm
- Mint Medical GmbH
- Quantity
- 101 systems
- Product code
- LLZ
- FDA submissions
- K142647
- Distribution
- Worldwide distribution - US Nationwide and the countries of Germany, Switzerland, Austria, Belgium, Italy, France, Romania, Netherlands, United Kingdom, Spain, Czech Republic, Ireland, Poland;
Product description
Mint Lesion; Software Versions: 3.4.0 up to 3.9.5.;
Code information (models, lots, serial numbers)
Software Versions: 3.4.0 up to 3.9.5. UDI-DI: 04260495880341, 04260495880358, 04260495880365, 04260495880372, 04260495880389, 04260495880396.
Devices named in this recall
6 devices in the FDA device registry whose identifier appears in the recall's code information.
- mintLesion™ 3.9.0Mint Medical GmbH · System, Image Processing, Radiological2 recalls0 reports
- mintLesion® 3.4.0Mint Medical GmbH · System, Image Processing, Radiological1 recall0 reports
- mintLesion® 3.5.0Mint Medical GmbH · System, Image Processing, Radiological1 recall0 reports
- mintLesion® 3.6.0Mint Medical GmbH · System, Image Processing, Radiological1 recall0 reports
- mintLesion® 3.7.0Mint Medical GmbH · System, Image Processing, Radiological1 recall0 reports
- mintLesion™ 3.8.0Mint Medical GmbH · System, Image Processing, Radiological1 recall0 reports
Other devices under the same FDA submission
1 device cleared under K142647 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- mintLesion™ 3.10.0Mint Medical GmbH · System, Image Processing, Radiological1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page