Skip to content

ZYMUTEST HIA IgG, in vitro diagnostic ELISA kit, Ref. RK040A recallZ-2171-2026 · Aniara Diagnostica LLC · initiated Apr 1, 2026

Class II Open, Classified All recalls posted in May 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Aniara Diagnostica LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Kits from lot FD1265 had been packaged on 17 October 2024 and put on the market with an outdated flyer (8 October 2024), containing incorrect values for the positive control.

FDA root cause category: Labeling Change Control

What the company is doing

A Field Corrective Action notification letter dated 3/30/26 was sent to customers. Actions implemented by HYPHEN BioMed: -Confirmation of the existence of a flyer discrepancy affecting 95 kits from lot FD1265. - Request for quarantine of the affected kits in US agent warehouse. -Coordination with US agent for notification of the impacted end users. -investigation to define the root cause and to establish corrective and preventive actions. Actions to be implemented by the end user: 1. Check the TAV flyer included in their kit. - If the flyer matches the correct version (dated 18 October 2024) - No action is required. - If the flyer corresponds to the incorrect version (dated I October 2024) - proceed to step 2 2. Destroy the incorrect TAV flyer. The outdated flyer included in the kit must be immediately removed and destroyed, as well as any paper or electronic copy that could have been done in end-user facility, to prevent further use. 3. Replace the incorrect flyer with the corrected version. Use the corrected TAV flyer provided with this notice for all future testing 4. Return the completed and signed form to your local distributor before April 15th, 2026. For any question or information regarding this notification, please contact vigilance team, by email at [email protected].

Recall details

Recall number
Z-2171-2026
Recall class
Class II
Status
Open, Classified
Initiated by the company
Apr 1, 2026
Posted by the FDA
May 12, 2026
Recalling firm
Aniara Diagnostica LLC
Quantity
359 units
Product code
LCO
FDA submissions
K071255
Distribution
US Nationwide distribution in the states of CA, GA, MN, MO, NC, NM, NY, VA, WA, and WI

Product description

ZYMUTEST HIA IgG, in vitro diagnostic ELISA kit, Ref. RK040A

Code information (models, lots, serial numbers)

UDI-DI: 03663537018763; Lot Number: FD1265

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

2 devices cleared under K071255 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page