Skip to content

Titanium/Fluoroplastic Piston 0.6mm Dia X 3.75mm L recallZ-2308-2026 · Grace Medical, Inc. · initiated Apr 3, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Grace Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to incorrect functional length on device labeling.

FDA root cause category: Process control

What the company is doing

On 04/03/2026, the firm emailed a "RECALL NOTICE" to it single customer informing them that the firm determined the pistons were manufactured with an incorrect functional length. Devices are labeled as 3.75 mm but have an actual functional length of 4.00 mm. The customer was instructed to: " Immediately discontinue use and quarantine affected product " Return affected product using the provided RMA " If any devices have been used, provide the date(s) of surgery on the response form " Complete and return the enclosed Recall Response Form For questions - contact Grace Medical at 901-386-0990 or [email protected].

Recall details

Recall number
Z-2308-2026
Status
Open, Classified
Initiated by the company
Apr 3, 2026
Posted by the FDA
Jun 4, 2026
Recalling firm
Grace Medical, Inc.
Quantity
7
Product code
ETB
FDA submissions
K060518
Distribution
U.S Nationwide distribution in the state of MD.

Product description

TITANIUM/FLUOROPLASTIC PISTON 0.6MM DIA X 3.75MM L Model/Catalog Number: 409-375 Product Description: Stapes Prosthesis

Code information (models, lots, serial numbers)

Model/Catalog Number: 409-375 UDI code: (01)00844505000772(10)122927(17)310301 Lot Number: 122927

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K060518 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.