Skip to content

EVA NEXUS Posterior 23G TDC Pack recallZ-2322-2026 · D.O.R.C. Dutch Opthalmic Research Center Intl B.V. · initiated Apr 30, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from D.O.R.C. Dutch Opthalmic Research Center Intl B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

An increase in incidence in complaints of a loose outer knife and reports of the outer knife becoming loose during use.

FDA root cause category: Process design

What the company is doing

On April 30, 2026 Urgent Medical Device Recall letters were sent to customers. Actions to be taken: Review this Product information as soon as possible with relevant members of your staff. 1. Pass this FSN as soon as possible on to all those who need to be aware of it, including other locations, organizations and/or customers who might have received the impacted Product(s) through you. 2. If you have the affected Products and the affected lots (see attachment below) in your inventory, remove and return them to DORC following the instructions in the attachment. 3. Follow the instructions on how to complete and submit the Field Action Acknowledgement Form that is provided with this FSN.

Recall details

Recall number
Z-2322-2026
Status
Open, Classified
Initiated by the company
Apr 30, 2026
Posted by the FDA
Jun 5, 2026
Recalling firm
D.O.R.C. Dutch Opthalmic Research Center Intl B.V.
Quantity
314 boxes, totaling 1884 units
Product code
HQE
FDA submissions
K213467
Distribution
Worldwide - US Nationwide and the countries of AUSTRALIA, AUSTRIA, BELGIUM, BULGARIA, CZECHIA, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, HONG KONG, IRELAND, ITALY, JAPAN, NETHERLANDS, NEW ZEALAND, POLAND, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, TAIWAN, TURKEY.

Product description

TDC VELOCE", 23G Two Dimensional Cutter - 20k CPM; Article Numbers: 9268.VIT23, 9311.23G01, 9311.23G02, 9311.23K01, 9311.23K02, 9311.23K03, 9623.G0201, 9623.G0202, 9623.K0201, 9311.23G01-00;

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

1. Article Number: 9268.VIT23; UDI-DI: 08717872031866; 2. Article Number: 9311.23G01; UDI-DI: 08717872035109; 3. Article Number: 9311.23G02; UDI-DI: 08717872035338; 4. Article Number: 9311.23K01; UDI-DI: N/A; 5. Article Number: 9311.23K02; UDI-DI: N/A; 6. Article Number: 9311.23K03; UDI-DI: N/A; 7. Article Number: 9623.G0201; UDI-DI: N/A; 8. Article Number: 9623.G0202; UDI-DI: N/A; 9. Article Number: 9623.K0201; UDI-DI: N/A; 10. Article Number: 9311.23G01-00; UDI-DI: 08717872035307; Lot Numbers: 100637001, 57682-*-*-1, 64312-*-*-1, 69017-*-*-1, 73237-*-*-1, 74225-*-*-1, 75735-*-*-1, 79460-*-*-1, 80338-*-*-1, 80963-*-*-1, 81335-*-*-1, 81679-*-*-1, 82213-*-*-1, 82325-*-*-1, 82742-*-*-1, 82779-*-*-1, 83584-*-*-1, 84580-*-*-1, 85451001, 86069001, 86798001, 86799001, 87205001, 88231001, 91957001, 92629001, 92630001, 93336001, 94351001, 95436001, 95462001, 96097001, 96123001, 97279001, 97789001, 98643001, 99130001, 99185001, 99697001;

Devices named in this recall

4 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

10 devices cleared under K213467 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 98814).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page