Brand Name: EXENT Analyser recallZ-2362-2026 · The Binding Site Group, Ltd. · initiated Apr 16, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from The Binding Site Group, Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

It was reported that plates 2 and 3 were marked as passed , although it appeared that QC had not been performed on these plates, despite samples being measured. The potential clinical hazards include false or missed detection of M-protein (false positive); and falsely elevated or decreased M-protein concentration.

FDA root cause category: Under Investigation by firm

What the company is doing

Firm notified customers via "Urgent Medical Device Correction" letter dated April 14, 2026. Customers were advised to confirm QC results are present in the QC results screen for the corresponding plate barcode after the completion of each plate measurement. If no QC result is displayed, do not release results from that plate and repeat the run ensuring QC material is properly loaded. The customer should also contact US Technical Support. When performing multi-plate runs while loading a previously used QC vials, ensure newly opened QCs are loaded onto Reagent Runner 2 as well, or discard and load newly opened QC vials into Reagent Runner 1. Retrospective review of previously released results is not considered necessary by the recalling firm. Please maintain awareness of the issue until Thermo Fisher Scientific identifies and implements a software enhancement that addresses the issue.

Recall details

Recall number
Z-2362-2026
Status
Open, Classified
Initiated by the company
Apr 16, 2026
Posted by the FDA
Jun 9, 2026
Recalling firm
The Binding Site Group, Ltd.
Quantity
1
Product code
OTA
FDA submissions
K250159
Distribution
US Nationwide distribution.

Product description

Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20 Product Description: The EXENT Analyser is an automated analyser intended for the qualitative and quantitative in vitro measurement of analytes in human body fluids used in conjunction with the EXENT assays.

Code information (models, lots, serial numbers)

Lot Code: Model No; IE800A; EXENT Analyser; UDI; 05051700020886, Serial Number; 101150, Software version; v1.0.20.