Adapter for Head Positioning, Model/Catalog Number:… recallZ-2638-2026 · Baxter Healthcare Corporation · initiated Apr 21, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The Adapter for Head Positioning accessory may become loose during patient positioning and unintentionally move when used in conjunction with the side rails of the Upper Back Section H V U component on Baxter operating tables.
FDA root cause category: Device Design
What the company is doing
An URGENT PRODUCT RECALL notification letter dated 4/20/26 was sent to customers. Actions to be Taken by Customers 1. Please notify all users of the product affected within your facility, especially all operating room staff. The affected Adapters may continue to be used only when mounted directly to the side rails of the operating table back plate. Do not use the Adapter on the Upper Back Section H V U side rails (component). Using the Upper Back Section H V U (separate component which is attached to the operating table s back plate) in conjunction with the Adapter accessory must be stopped/ceased, until a replacement takes place. 2. Once the newly designed Adapters are available Baxter will contact customers with instructions for return and replacement of the impacted Adapters. 3. If you received this communication directly from Baxter, acknowledge receipt by following the within the body of the corresponding email. Acknowledging receipt of this notification will prevent you from receiving repeated notices. If you do not complete the acknowledgement, you will receive a phone call from Accenture LLP on behalf of Baxter to confirm your receipt of this notification. 4. If you purchased this product from a distributor or wholesaler, please contact your supplier to arrange for return of the affected product. Please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to them according to their instructions. 5. Please forward a copy of this communication to the director of nursing, facility risk manager, patient safety, biomedical engineering department, director of surgical services, operating room staff, and any other departments within your institution who use the affected product. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please conduct a user-level recall of
Recall details
- Recall number
- Z-2638-2026
- Status
- Open, Classified
- Initiated by the company
- Apr 21, 2026
- Posted by the FDA
- Jul 1, 2026
- Recalling firm
- Baxter Healthcare Corporation
- Quantity
- 13 units
- Product code
- GDC
- Distribution
- Worldwide - US Nationwide distribution in the states of Minnesota, Maryland, and Pennsylvania and the countries of Canada, Hong Kong, India, Japan, Korea, Philippines, Taiwan, Thailand, Vietnam, Austria, Belgium, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Iraq, Israel, Italy, Jordan, Kenya, Lebanon, Libya, Malta, Morocco, Netherlands, Norway, Oman, Pakistan, Poland, Portugal, Qatar, Romania, Saudi Arabia, Senegal, Serbia, Slovakia, Spain, Sweden, Switzerland, Tunisia, Turkey, United Arabia Emrites, Ukraine, Argentina, Brazil, Chile, Colombia, Costa Rica, Mexico, and Peru.
Product description
Adapter for Head Positioning, Model/Catalog Number: 1739994, when used in conjunction with the side rails of the Upper Back Section H V U component (product code 1909813) on Baxter operating tables
Code information (models, lots, serial numbers)
Lot Code: Product Code: 1739994 UDI-DI: 00887761972780 Serial Numbers: 109931797, 109931798, 109931799, 109943588, 109943590, 109948550, 109964963, 109964964, 109964969, 109964970, 109964971, 109980700, and 110006694
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Head-Positioning accessories 1739994Baxter Medical Systems GmbH + Co. KG · TABLE, OPERATING-ROOM, ELECTRICAL1 recall1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.