Skip to content

Medline medical convenience kit labeled as EYE TRAY-LF… recallZ-2260-2026 · Medline Industries, LP · initiated Apr 10, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Specific Medline Kits contain Cardinal Health Monoject 1mL Luer Lock Syringes, Tuberculin Print, Sterile. Cardinal Health has recalled these syringes because the outer carton and blister pack are labeled as a 1 mL Luer Lock Tuberculin Syringe (Product Code 1180100777), but the syringes inside are U-100 insulin syringes.

FDA root cause category: Under Investigation by firm

What the company is doing

Medline Industries, LP issued an URGENT MEDICAL DEVICE RECALL notice to it consignees on 4/10/2026 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed on the enclosed response form. Quarantine all affected product. 2. Please complete and return the enclosed response form listing the quantity of affected product you have in inventory. Even if you do not have any affected product in inventory, please complete and return the response form. 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product." For questions, contact the Recall Department at 866-359-1704 or [email protected].

Recall details

Recall number
Z-2260-2026
Status
Open, Classified
Initiated by the company
Apr 10, 2026
Posted by the FDA
May 28, 2026
Recalling firm
Medline Industries, LP
Quantity
8 kits
Product code
OJK
Distribution
US Nationwide distribution in the states of California.

Product description

Medline medical convenience kit labeled as EYE TRAY-LF, Medline Kit SKU DYNJ21627P

Code information (models, lots, serial numbers)

UDI/DI 10198459151569 (each), 40198459151560 (case), Lot Numbers: 26BMC538

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates