Bicarby recallZ-2321-2026 · Fresenius Medical Care Holdings, Inc. · initiated May 12, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Fresenius Medical Care Holdings, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Given the increased number of leaks reported by users, it was determined to be the appropriate course of action to notify users of the potential for the occurrence of Bicarby luer-lock leaks during manipulation of the frangible and tubing when preparing the product for use for treatment. This issue presents the potential of a slip and fall hazard associated with fluid on the floor.
FDA root cause category: Process design
What the company is doing
On 5/12/2026 an IMPORTANT MEDICAL DEVICE INFORMATION letters were sent to customers to inform them that manipulation of the luer-lock location during treatment preparation may result in fluid spillage onto the user and floor. This occurs prior to treatment and renders the product unusable. Direct contact with the fluid does not present any risk to the user as frangible manipulation only occurs after the peal seam has been broken and the two fluid compartments have been mixed. The potential risk associated with the affected Bicarby Dialysate product is a slip and fall hazard due to the presence of fluid on the floor.
Recall details
- Recall number
- Z-2321-2026
- Status
- Open, Classified
- Initiated by the company
- May 12, 2026
- Posted by the FDA
- Jun 5, 2026
- Recalling firm
- Fresenius Medical Care Holdings, Inc.
- Quantity
- 1,756,108 units
- Product code
- KPO
- FDA submissions
- K233950, K243786
- Distribution
- US Nationwide distribution in the states of NY, TN and TX.
Product description
Bicarby" Dialysate; Model Numbers: RFP-400-G, RFP-401-G, RFP-402-G, RFP-404-G, RFP-407-G, and RFP-456-G;
Code information (models, lots, serial numbers)
Model Number: RFP-400-G; UDI-DI: 00840861102839; 2. Model Number: RFP-401-G; UDI-DI: 00840861102853; 3. Model Number: RFP-402-G; UDI-DI: 00840861102822; 4. Model Number: RFP-404-G; UDI-DI: 00840861102914; 5. Model Number: RFP-407-G; UDI-DI: 00840861102846: 6. Model Number: RFP-456-G; UDI-DI: 00840861102945; All lots produced from January 13, 2025, through lot 26AG02019 on January 13, 2026
Devices named in this recall
6 devices in the FDA device registry whose identifier appears in the recall's code information.
- Bicarby RFP-400-GFresenius Medical Care Renal Therapies Group, LLC · Dialysate Concentrate For Hemodialysis (Liquid Or Powder)2 recalls0 reports
- Bicarby RFP-402-GFresenius Medical Care Renal Therapies Group, LLC · Dialysate Concentrate For Hemodialysis (Liquid Or Powder)1 recall0 reports
- Bicarby RFP-407-GFresenius Medical Care Renal Therapies Group, LLC · Dialysate Concentrate For Hemodialysis (Liquid Or Powder)1 recall0 reports
- Bicarby RFP-401-GFresenius Medical Care Renal Therapies Group, LLC · Dialysate Concentrate For Hemodialysis (Liquid Or Powder)1 recall0 reports
- Bicarby RFP-404-GFresenius Medical Care Renal Therapies Group, LLC · Dialysate Concentrate For Hemodialysis (Liquid Or Powder)1 recall0 reports
- Bicarby RFP-456-GFresenius Medical Care Renal Therapies Group, LLC · Dialysate Concentrate For Hemodialysis (Liquid Or Powder)1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page