Skip to content

Allia recallZ-2318-2026 · GE Medical Systems, LLC · initiated Apr 20, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Medical Systems, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.

FDA root cause category: Device Design

What the company is doing

An URGENT MEDICAL DEVICE CORRECTION dated 4/20/26 was sent to customers. Actions to be taken by Customer/User You can continue to use your system. As a reminder, qualified service personnel must follow the Service Manual applicable to your system and use the appropriate personal protective equipment (PPE) for electrical hazards when performing any activity inside the cabinet while power is ON. If any abnormal condition is observed on the Iris Power Cable during servicing, such as loose or visibly exposed wires, qualified service personnel must follow the Electrical Lockout/Tagout Procedure as described in Service Manual. Please ensure that all potential users and service personnel in your facility are made aware of this safety notification and the recommended actions and post this letter in a visible location near the C-FRT cabinet. Please retain this document for your records. Please complete and return the attached acknowledgement form electronically via FMI 18014 Digital Response Form or print, fill out manually, scan, and email to [email protected].

Recall details

Recall number
Z-2318-2026
Status
Open, Classified
Initiated by the company
Apr 20, 2026
Posted by the FDA
Jun 5, 2026
Recalling firm
GE Medical Systems, LLC
Quantity
3 units
Product code
OWB
FDA submissions
K251199
Distribution
Worldwide - US Nationwide distribution.

Product description

Allia Moveo angiographic X-ray system

Code information (models, lots, serial numbers)

UDI-DI: 00198953052488; Serial Numbers: C4-25-001, C4-25-002, C4-26-001, C4-26-002, C4-26-004, C4-26-005

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 98880).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page