da Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number:… recallZ-2722-2026 · Intuitive Surgical, Inc. · initiated May 12, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Intuitive Surgical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Robotic-assisted surgical system arm drapes may detach from the instrument arm due to an issue involving the drape rings located at the top of the product, which may expose a section of instrument arm, which may result in procedure delay or conversion to an alternate minimally invasive approach. Retrospectively reported recall.
FDA root cause category: Process control
What the company is doing
On 5/12/2025, correction notices were emailed to customers informing them of the following: 1) Conduct pre-use inspection steps outlined prior to each procedure, see below. 2) If you encounter a drape that does not pass the inspection steps, contact firm customer service. 3) Complete and return the acknowledgement form via email to [email protected] or [email protected] If you need further information or support concerning this Safety Notice, please contact your Clinical Sales Representative or contact firm Customer Service at the numbers listed below: - North and South America: (800) 876-1310, Option 3 or mail: [email protected]. - Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 or [email protected] - South Korea: 02-3271-3200or [email protected] - Japan: 0120-56-5635 or 03-5575-1362 or [email protected] " Taiwan: +0800-86-8181 Pre-Use Inspection Steps: 1. Fully drape the da Vinci SP system according to user manual. 2. Press and hold the Instrument Arm Clutch button. Top Instrument arm clutch, not recommended. 3. While holding the Instrument Arm Clutch button, manually rotate the instrument drive cluster in one direction, 360-degrees. Then, rotate the instrument drive at least 360-degrees in the other direction. Stop immediately and press "deploy for draping" again to rehome the instrument drive cluster to replace the drape if you observe any of the following behaviors: a. Inability to rotate the instrument drive 360-degrees in both directions b. VSC message that drape is not fully attached (VSC message: "Check drape is secure to all 4 instrument arm clips" or "Insecure or missing drape. The drape may not be properly secured and could fall. Ensure the drape is properly connected to all 4 mounts to continue"). c. Drape outer ring falls off silver section of system. d. Any of the 4 dark gray marks disengage from the dark gray connectors during the rotation.
Recall details
- Recall number
- Z-2722-2026
- Status
- Open, Classified
- Initiated by the company
- May 12, 2025
- Posted by the FDA
- Jul 10, 2026
- Recalling firm
- Intuitive Surgical, Inc.
- Quantity
- 11113
- Product code
- NAY
- FDA submissions
- K182371, K173906
- Distribution
- Worldwide - US Nationwide distribution in the states of NJ, OR, NY, FL, AL, SD, TX, VA, WV, PA, CA, WA, CT, MA, IL, OK, OH, MI, NC, MO, KY, WI, AZ, CO, GA, MD, SC, IN, NV, TN, MS, KS, IA, UT, MN, LA, NE, ME and the countries of Germany, Sweden, United Kingdom, Italy, Belgium, Switzerland, Denmark, Japan, South Korea, Hong Kong, Taiwan, France, Ireland, Spain.
Product description
da Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-15
Code information (models, lots, serial numbers)
UDI-DI: 00886874113486, Lot: 3074LA0500, 1124LA0200, 2124LA0400, 1134LA0700, 1144LA0500, 7154LA0200, 6164LA0800, 1084LA0300, 3164LAA800, 4084LA0500, 4124LA0600, 1274LA0500, 2294LA0200, 1274LA0400, 4284LA0800, 6324LA0200, 1324LA0300, 4154LA0400, 2284LA0100, 4144LA1300, 3324LA0500, 2354LA0500, 7144LA0500, 4294LA0700, 3374LA0400, 5264LA0700, 3224LA0500, 3424LA0200, 1394LA0100
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- None 430013Intuitive Surgical, Inc. · System, surgical, computer controlled instrument2 recalls139 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page