Surgify Halo | 4.0 mm | Extendable recallZ-2559-2026 · Surgify Medical Oy · initiated Apr 30, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Surgify Medical Oy or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS).
FDA root cause category: Under Investigation by firm
What the company is doing
Surgify issued a Medical Device Correction Notice - Surgify Halo - Restriction of Use notice to its consignees on 4/30/2026 via email. The notice explained the issue, potential risk, and requested the following: "What steps must the user take? Please review your current practice and ensure the device is only used in accordance with the updated safety information. Instruction to Customers: The customers have been instructed to do the following: - Do not use this device in endoscopic surgical procedures. If endoscopic access is required, an alternative instrument should be selected. Specifically: . Do not use this device in endoscopic surgical procedures. If endoscopic access is required, an alternative instrument should be selected. . Ensure that all relevant clinical staff and surgical team members are made aware of this notice. . Please retain this notice for your records and attach it to the relevant product documentation. Please acknowledge receipt of this Medical Device Correction Notice by replying to [email protected] within 30 business days, confirming that the notice has been read, understood, and communicated to all relevant users within your organization. Please include your facility name, your name, and your role. If you have previously used the device in endoscopic procedures without incident, no patient follow-up is required; however, this application should now be discontinued in line with the updated safety information." For questions regarding this matter, please contact us by phone at +358 105 176 310 or via email at [email protected].
Recall details
- Recall number
- Z-2559-2026
- Status
- Open, Classified
- Initiated by the company
- Apr 30, 2026
- Posted by the FDA
- Jun 24, 2026
- Recalling firm
- Surgify Medical Oy
- Product code
- HBE
- FDA submissions
- K253627
- Distribution
- US Nationwide distribution in the states of MN, MA, IN, LA, CA, NY, FL.
Product description
Surgify Halo, 4.0 mm, Extendable, Model/Catalog Number: 40.000.SEE.U1; drills, burrs, trephines & accessories (simple, powered)
Code information (models, lots, serial numbers)
UDI: 06429811532281; All lots until the IFU update has been implemented
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Surgify Halo | 4.0 mm | Extendable 40.000.SEE.U1Surgify Medical Oy · Drills, Burrs, Trephines & Accessories (Simple, Powered)1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 98918).
- Z-2555-2026 · Open, ClassifiedSurgify Halo, 3.0 mm, Extendable, Model/Catalog Number: 30.000.SEE.U2; drills, burrs, trephines & accessories (simple, powered)
- Z-2556-2026 · Open, ClassifiedSurgify Halo, 3.0 mm, Short, Model/Catalog Number: 30.070.NVG.U2; drills, burrs, trephines & accessories (simple, powered)
- Z-2557-2026 · Open, ClassifiedSurgify Halo, 3.0 mm, Long, Model/Catalog Number: 30.125.NVG.U2; drills, burrs, trephines & accessories (simple, powered)
- Z-2558-2026 · Open, ClassifiedSurgify Halo, 4.0 mm, Extendable, Model/Catalog Number: 40.000.SEE.H1; drills, burrs, trephines & accessories (simple, powered)
- Z-2560-2026 · Open, ClassifiedSurgify Halo, 4.0 mm, Short, Model/Catalog Number: 40.070.NVG.H1; drills, burrs, trephines & accessories (simple, powered)
- Z-2561-2026 · Open, ClassifiedSurgify Halo, 4.0 mm, Short, Model/Catalog Number: 40.070.NVG.U1; drills, burrs, trephines & accessories (simple, powered)
- Z-2562-2026 · Open, ClassifiedSurgify Halo, 4.0 mm, Short, Model/Catalog Number: 40.125.NVG.H1; drills, burrs, trephines & accessories (simple, powered)
- Z-2563-2026 · Open, ClassifiedSurgify Halo, 40 mm, Long, Model/Catalog Number: 40.125.NVG.U1; drills, burrs, trephines & accessories (simple, powered)
- Z-2564-2026 · Open, ClassifiedSurgify Halo, 5.4 mm, Extendable, Model/Catalog Number: 54.000.SEE.H1; drills, burrs, trephines & accessories (simple, powered)
- Z-2565-2026 · Open, ClassifiedSurgify Halo, 5.4 mm, Extendable, Model/Catalog Number: 54.000.SEE.U1; drills, burrs, trephines & accessories (simple, powered)
- Z-2566-2026 · Open, ClassifiedSurgify Halo, 5.4 mm, Short, Model/Catalog Number: 54.070.NVG.H1; drills, burrs, trephines & accessories (simple, powered)
- Z-2567-2026 · Open, ClassifiedSurgify Halo, 5.4 mm, Short, Model/Catalog Number: 54.070.NVG.U1; drills, burrs, trephines & accessories (simple, powered)
- Z-2568-2026 · Open, ClassifiedSurgify Halo, 5.4 mm, Medium, Model/Catalog Number: 54.085.SHD.H1; burrs, trephines & accessories (simple, powered)
- Z-2569-2026 · Open, ClassifiedSurgify Halo, 5.4 mm, Medium, Model/Catalog Number: 54.085.SHD.U1; drills, burrs, trephines & accessories (simple, powered)
- Z-2570-2026 · Open, ClassifiedSurgify Halo, 5.4 mm, Long, Model/Catalog Number: 54.125.NVG.H1; drills, burrs, trephines & accessories (simple, powered)
- Z-2571-2026 · Open, ClassifiedSurgify Halo, 5.4 mm, Long, Model/Catalog Number: 54.125.NVG.U1; drills, burrs, trephines & accessories (simple, powered)
- Z-2572-2026 · Open, ClassifiedSurgify Halo, 5.4 mm, Long, Model/Catalog Number: 54.140.SHD.H1; drills, burrs, trephines & accessories (simple, powered)
- Z-2573-2026 · Open, ClassifiedSurgify Halo, 5.4 mm, Long, Model/Catalog Number: 54.140.SHD.U1; drills, burrs, trephines & accessories (simple, powered)
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page