3mensio Workstation (Vascular Fenestrated) software recallZ-2767-2026 · PIE Medical Imaging B.V. · initiated May 7, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from PIE Medical Imaging B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

User measurement value reported by diagnostic bioimaging software intended to measure/visualize cardiovascular structures may be larger (0.8% to 34.4%) than the actual anatomical dimension, if all the following occur: affected software is used, and the user has access only to 3D Echo OR Mitral 3D Echo, and the dataset being analyzed is a Philips 3D echo dataset containing Color Doppler information

FDA root cause category: Software change control

What the company is doing

On 5/7/2026, correction notices were emailed to customers who were asked to do the following: 1) Do not perform measurements on Philips 3D echo datasets containing the doppler information using 3mensio Workstation 10.7, 10.7 SP1 SP5. 2) Ensure that this notice is distributed to all relevant personnel within your organization and to any organization where the affected product may have been transferred to or is in use. 3) PMI recommends follow-up to verify if results of ordered devices are correctly analyzed/calculated. 4) Complete and return the enclosed Customer Acknowledgment Form and return via email to [email protected] This issue has been resolved in 3mensio Workstation 10.7 SP6. PMI will support users in upgrading to this version.

Recall details

Recall number
Z-2767-2026
Status
Open, Classified
Initiated by the company
May 7, 2026
Posted by the FDA
Jul 21, 2026
Recalling firm
PIE Medical Imaging B.V.
Quantity
165
Product code
QIH
FDA submissions
K250330
Distribution
Worldwide - US Nationwide distribution in the states of CA, NY, MN, MI, TX and the countries of United Kingdom, Australia, France, Italy, Hong Kong SAR, Thailand, Taiwan, Netherlands, Ireland, Spain, Greece, Japan, Belgium.

Product description

3mensio Workstation (Vascular Fenestrated) software

Code information (models, lots, serial numbers)

Software Version 10.7. UDIs: (01)08056304455925(11)250423(8012)10.7, (01)08056304455925(11)250523(8012)10.7 SP1, (01)08056304455925(11)250714(8012)10.7 SP2, (01)08056304455925(11)250729(8012)10.7 SP3, (01)08056304455925(11)251016(8012)10.7 SP4, (01)08056304455925(11)251103(8012)10.7 SP5

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K250330 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page