GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L… recallZ-2246-2026 · GE Medical Systems Information Technologies Inc · initiated Apr 24, 2026
Class II Open, Classified All recalls posted in May 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Medical Systems Information Technologies Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
FDA root cause category: Software design
What the company is doing
GE HealthCare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 4/24/2026 by letter using a traceable means. The notice explained the the safety issue pertaining to the persistent NO COMM or OFF NETWORK condition, and provided instructions for the safe use of the device should either issue arise. "Actions to be taken by Customer/User Pending corrections from GE HealthCare, you can continue to use your ApexPro CTS with Central Station for telemetry only or combination monitoring unless you observe a persistent (more than a few seconds) NO COMM or OFF NETWORK notification. If you do observe a persistent NO COMM or OFF NETWORK notification, follow the instructions below. 1. Use alternative bedside patient monitoring to ensure continuous clinical surveillance until the NO COMM or OFF NETWORK condition is resolved by GE HealthCare. NOTE: If you are using combination monitoring, it is recommended to switch ECG monitoring to the bedside monitor. 2. Immediately notify your Facility s Biomedical Engineering, IT or Service Department and contact GE HealthCare Service. 3. Verify monitoring is restored at the Central Station(s) and/or bedside monitor(s). Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please post this notification by your Central Station(s). Please retain this document for your records." "GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction." For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Recall details
- Recall number
- Z-2246-2026
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Apr 24, 2026
- Posted by the FDA
- May 28, 2026
- Recalling firm
- GE Medical Systems Information Technologies Inc
- Quantity
- 2366 units
- Product code
- MHX
- FDA submissions
- K103142
- Distribution
- Worldwide distribution.
Product description
GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICA, REF #5697603-902, Product Code KSA; used in conjunction wtih ApexPro CARESCAPE Telemetry Server (CTS) v6.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Code information (models, lots, serial numbers)
Show the full code information (28k characters)
No UDI, Serial Number or Sales Order Number: KSA7536146, KSA7536158, KSA7536159, KSA7550156, KSA7550158, KSA7550159, KSA7667317, KSA7754389, KSA7769597, KSA7769610, KSA7769612, KSA7773420, KSA7773431, KSA7773440, KSA7120704, KSA7120767, KSA6964886, KSA6808842, KSA6808852, KSA6808866, KSA6808868, KSA6808881, KSA6808883, KSA6808912, KSA6808928, KSA6866974, KSA6664007, KSA6866963, KSA7120766, KSA7162586, KSA6623197, KSA6623216, KSA6623223, KSA6623226, KSA6623239, KSA6623248, KSA6623270, KSA6623278, KSA7391460, KSA6477094, KSA6477095, KSA6477135, KSA6477155, KSA6477159, KSA6477206, KSA6954245, KSA6954269, KSA6954276, KSA6954277, KSA6954284, KSA6954291, KSA6477096, KSA6477154, KSA7702196, KSA7702198, KSA7702199, KSA7702208, KSA7702222, KSA6623191, KSA6623194, KSA6623237, KSA6623251, KSA6623267, KSA6626626, KSA6626632, KSA6626639, KSA6626666, KSA6626670, KSA6626716, KSA6536442, KSA6536443, KSA6536454, KSA6536460, KSA6536469, KSA6536484, KSA6536485, KSA6536490, KSA6536496, KSA6536504, KSA6536526, KSA6536528, KSA6536532, KSA7120699, KSA7120728, KSA7120764, KSA7120773, KSA7162570, KSA7162593, KSA7162610, KSA7162622, KSA7162640, KSA6623198, KSA6623202, KSA6623232, KSA6623236, KSA6623275, KSA6623280, KSA6626622, KSA6626625, KSA6626635, KSA6626636, KSA6626648, KSA6626655, KSA6626659, KSA6626661, KSA6536520, KSA6536525, KSA7304431, KSA6866921, KSA6866926, KSA6866927, KSA6866983, KSA6808882, KSA6649195, KSA6649250, KSA7162575, KSA7162588, KSA7162608, KSA7162614, KSA7162656, KSA6899015, KSA6899022, KSA6808879, KSA5646278, KSA6017318, KSA6017321, KSA6017325, KSA6017333, KSA6017336, KSA6017352, KSA6017360, KSA6017361, KSA6017367, KSA6017368, KSA6017370, KSA6017371, KSA6017377, KSA6017378, KSA6017382, KSA6017383, KSA6017384, KSA6017385, KSA6017387, KSA6017388, KSA6017391, KSA6017404, KSA6017405, KSA6017406, KSA6017407, KSA6017409, KSA6017410, KSA6017416, KSA6017417, KSA6017424, KSA6017445, KSA6017450, KSA6017458, KSA6017471, KSA6017504, KSA6899026, KSA7676874, KSA7676878, KSA7676893, KSA7013925, KSA6808863, KSA6808864, KSA7304413, KSA7304437, KSA6945205, KSA6945208, KSA6945219, KSA6945223, KSA6945230, KSA6945236, KSA6945296, KSA6945314, KSA6945321, KSA6945340, KSA6945350, KSA6945353, KSA6945382, KSA6945387, KSA6899004, KSA6664048, KSA6664049, KSA6664054, KSA6664059, KSA6664075, KSA6664037, KSA6664045, KSA6664065, KSA6664069, KSA6623192, KSA6626641, KSA6626649, KSA6626687, KSA6626691, KSA6626701, KSA6626706, KSA6626707, KSA6626715, KSA6626718, KSA7120720, KSA7162574, KSA7162580, KSA7162632, KSA7162655, KSA6419987, KSA6419993, KSA6420030, KSA6420053, KSA6420080, KSA6420081, KSA6420082, KSA6420102, KSA6664002, KSA6664005, KSA6664036, KSA6664052, KSA6664057, KSA6664078, KSA6664090, KSA6664091, KSA6664006, KSA6866903, KSA6866924, KSA6866928, KSA6866933, KSA6866942, KSA6866977, KSA6866981, KSA6866987, KSA6866992, KSA6866996, KSA6899000, KSA6899060, KSA6945146, KSA6945185, KSA6945190, KSA6945191, KSA6945224, KSA6945250, KSA6945278, KSA6945294, KSA6945311, KSA6945391, KSA6945199, KSA6945204, KSA6945226, KSA6945244, KSA6945258, KSA6945302, KSA6945303, KSA6945319, KSA6945355, KSA6945358, KSA6945359, KSA6945384, KSA7013899, KSA7013902, KSA7013903, KSA7013905, KSA7013912, KSA7013914, KSA7013918, KSA7013921, KSA7013932, KSA7013940, KSA7013955, KSA7013956, KSA7013957, KSA7013963, KSA7013964, KSA7013973, KSA7013975, KSA7013986, KSA6569809, KSA6569821, KSA6569833, KSA6569840, KSA6569867, KSA6569890, KSA6569898, KSA6623189, KSA6623195, KSA6626637, KSA6626651, KSA6626690, KSA6626699, KSA6626713, KSA6626719, KSA6898999, KSA6899011, KSA6899013, KSA6899018, KSA6899021, KSA6899028, KSA6899029, KSA6899033, KSA6899034, KSA6899037, KSA6899051, KSA6899054, KSA6899059, KSA6899067, KSA6899070, KSA6899072, KSA6899074, KSA6899078, KSA6899080, KSA6899084, KSA6899085, KSA6899086, KSA6899092, KSA6899093, KSA6899097, KSA6945099, KSA6945100, KSA6945108, KSA6945113, KSA6945130, KSA6945140, KSA6945171, KSA6945175, KSA6945177, KSA6945183, KSA6945187, KSA6945189, KSA6945254, KSA6945262, KSA6945283, KSA6945288, KSA6945298, KSA6945310, KSA6945328, KSA6945331, KSA6945370, KSA6945377, KSA6945213, KSA6649214, KSA6649226, KSA6649261, KSA6649267, KSA6649272, KSA7120679, KSA7120683, KSA7120693, KSA7120725, KSA7120750, KSA7120754, KSA6689288, KSA6689302, KSA6689309, KSA6689313, KSA6689322, KSA6689364, KSA6689373, KSA6689375, KSA6689382, KSA7442815, KSA7676896, KSA7754382, KSA7754398, KSA7754402, KSA7769599, KSA7769600, KSA7769603, KSA7773423, KSA7773430, KSA7773436, KSA7773442, KSA7744947, KSA7744960, KSA7754384, KSA7754387, KSA7754395, KSA7769605, KSA7769607, KSA7769618, KSA7769622, KSA7773422, KSA7773428, KSA7773444, KSA7239656, KSA7239665, KSA7239668, KSA7239676, KSA7239677, KSA7239678, KSA7239683, KSA7239684, KSA7239686, KSA7239688, KSA7239694, KSA7239696, KSA7239697, KSA7239703, KSA7239704, KSA7239707, KSA7239708, KSA7239711, KSA7239712, KSA7239725, KSA7239728, KSA7239729, KSA7239736, KSA7239737, KSA7239741, KSA7239742, KSA7239743, KSA7239748, KSA6808841, KSA6808859, KSA6808902, KSA6808935, KSA6808936, KSA6623190, KSA6623209, KSA6623225, KSA6808870, KSA6808880, KSA7120744, KSA7120755, KSA6623204, KSA6623268, KSA6899005, KSA6899045, KSA6899050, KSA6899055, KSA6899089, KSA6623199, KSA6623214, KSA6626624, KSA6590128, KSA6590137, KSA6590143, KSA6590163, KSA6590170, KSA6590184, KSA6590210, KSA6590221, KSA6626663, KSA6626676, KSA6420092, KSA6420117, KSA6536467, KSA6536501, KSA6536510, KSA6536515, KSA6569806, KSA6569864, KSA6569903, KSA7724020, KSA7724027, KSA7724037, KSA7724040, KSA7727679, KSA7727686, KSA7727699, KSA7732509, KSA7732526, KSA7754383, KSA7773427, KSA6649211, KSA6649233, KSA6649263, KSA6569808, KSA6569839, KSA6569863, KSA6569871, KSA6569876, KSA6569888, KSA6808839, KSA6808853, KSA6808860, KSA6808893, KSA6808905, KSA6808919, KSA6808921, KSA6808924, KSA6808926, KSA6808938, KSA6409191, KSA6409209, KSA6409245, KSA6420002, KSA6420011, KSA6420108, KSA6477136, KSA6477153, KSA6477164, KSA6477184, KSA6477193, KSA6477196, KSA6477212, KSA6477225, KSA7550168, KSA7676876, KSA7724025, KSA7724045, KSA7727682, KSA7120701, KSA7120751, KSA7120713, KSA7120721, KSA7120731, KSA7120738, KSA7120741, KSA7120776, KSA7120777, KSA7013919, KSA7013934, KSA7013987, KSA6664013, KSA6664014, KSA6664016, KSA6664021, KSA6664083, KSA6664085, KSA6689285, KSA6689330, KSA6689353, KSA7391451, KSA7391479, KSA7442788, KSA7442812, KSA7442857, KSA7442859, KSA7442877, KSA7442879, KSA7442888, KSA7442895, KSA6866906, KSA6866919, KSA6866950, KSA6866970, KSA6866975, KSA6866976, KSA6866980, KSA6808847, KSA6808848, KSA6808849, KSA6808869, KSA6808888, KSA6808895, KSA6808914, KSA6808915, KSA6808918, KSA6808925, KSA7520462, KSA7520480, KSA6945229, KSA6623203, KSA6623229, KSA6623247, KSA6623264, KSA6626633, KSA6626634, KSA6626695, KSA6626717, KSA6623210, KSA6664055, KSA6664064, KSA7724021, KSA7724030, KSA7724033, KSA7732497, KSA7536149, KSA7120761, KSA7162595, KSA6945157, KSA6945235, KSA6945249, KSA6945275, KSA6945247, KSA6945290, KSA6945291, KSA6945327, KSA6945365, KSA7754377, KSA7769601, KSA7773419, KSA7773421, KSA7773429, KSA7773435, KSA6664003, KSA6866989, KSA7162566, KSA7162571, KSA7162585, KSA7162591, KSA7162594, KSA7162597, KSA7162607, KSA7162611, KSA7162615, KSA7162617, KSA7162620, KSA7162627, KSA7162633, KSA7162642, KSA7162643, KSA7162644, KSA7162647, KSA7162658, KSA7391429, KSA7391469, KSA7391472, KSA7096912, KSA7096925, KSA7096926, KSA7096929, KSA7096934, KSA7096936, KSA7096937, KSA7096938, KSA7096944, KSA7096948, KSA7096962, KSA7096973, KSA7096984, KSA7120708, KSA6945097, KSA6945109, KSA6945118, KSA6945128, KSA6945129, KSA6945132, KSA6945141, KSA6945181, KSA6017431, KSA6017433, KSA6017477, KSA6017495, KSA6017393, KSA6945394, KSA6899002, KSA6899019, KSA6899020, KSA6899030, KSA6899091, KSA6899048, KSA6899082, KSA6623185, KSA6623238, KSA6623261, KSA6649178, KSA6649227, KSA6649269, KSA6945147, KSA6945173, KSA6945277, KSA6945336, KSA6945120, KSA6945127, KSA6945148, KSA6945149, KSA6945156, KSA6945163, KSA6945168, KSA6945178, KSA6945179, KSA6945193, KSA6945260, KSA6945312, KSA6945341, KSA6945361, KSA6945395, KSA6964884, KSA6964914, KSA6964918, KSA6964930, KSA6964931, KSA6964941, KSA6626689, KSA7013890, KSA7013891, KSA7013893, KSA7013913, KSA7013923, KSA7013924, KSA7013942, KSA7013948, KSA7013954, KSA6945154, KSA6945215, KSA6945273, KSA6945300, KSA6945333, KSA6945352, KSA6945381, KSA6626630, KSA6626631, KSA6626657, KSA6626660, KSA6626665, KSA6626667, KSA6626672, KSA6626694, KSA6626697, KSA6626700, KSA6626709, KSA6964900, KSA6964911, KSA6964912, KSA6964922, KSA6964924, KSA6899044, KSA7239669, KSA7239670, KSA7239687, KSA7239698, KSA7239705, KSA7239714, KSA7239727, KSA7239730, KSA7239733, KSA7239746, KSA6569873, KSA6623228, KSA6623243, KSA6623266, KSA6623276, KSA6626640, KSA6626642, KSA6626643, KSA6626650, KSA6626653, KSA6626673, KSA6626675, KSA6626678, KSA6626711, KSA7096993, KSA7120702, KSA7120703, KSA7120705, KSA7120710, KSA7120759, KSA7120769, KSA7162573, KSA7162599, KSA7162601, KSA7162635, KSA7162651, KSA7162657, KSA6945165, KSA6945315, KSA6477102, KSA6477115, KSA6477145, KSA6477150, KSA6945209, KSA6945212, KSA6945232, KSA6945271, KSA6945276, KSA6409175, KSA6409178, KSA6409192, KSA6409193, KSA6409204, KSA6409207, KSA6409210, KSA6409212, KSA6409226, KSA6409232, KSA6409236, KSA7162604, KSA7162645, KSA7744936, KSA7744945, KSA7744948, KSA7744954, KSA7744962, KSA7744963, KSA7754380, KSA7754388, KSA7754394, KSA7754403, KSA7754406, KSA7769595, KSA7769604, KSA7769616, KSA7769620, KSA7773418, KSA7773426, KSA7773437, KSA7773441, KSA7442825, KSA7442839, KSA7676899, KSA7727681, KSA7744940, KSA7744942, KSA7744949, KSA7744950, KSA7754378, KSA7754379, KSA7754381, KSA7754392, KSA7754393, KSA7754397, KSA7754399, KSA7754404, KSA7769594, KSA7769611, KSA7769615, KSA7769619, KSA7773425, KSA7773432, KSA7773438, KSA6623255, KSA6626621, KSA6626629, KSA6626664, KSA6626714, KSA7442808, KSA6899036, KSA7724028, KSA7239735, KSA7013974, KSA6623222, KSA6626644, KSA6626692, KSA7120717, KSA7120745, KSA7162629, KSA7162648, KSA7120687, KSA7162567, KSA7162598, KSA7239659, KSA7239660, KSA7239671, KSA7239680, KSA7239715, KSA7239717, KSA7239720, KSA7239731, KSA7239732, KSA7304420, KSA6945182, KSA6945211, KSA7120718, KSA7013941, KSA7574789, KSA6477112, KSA6477120, KSA6477152, KSA6477179, KSA6477194, KSA6866946, KSA6866969, KSA6477157, KSA6477176, KSA6477185, KSA6477192, KSA6477228, KSA6899008, KSA6899056, KSA6945266, KSA6945323, KSA7013889, KSA7013898, KSA7013904, KSA7013908, KSA7013909, KSA7013911, KSA7013920, KSA7013922, KSA7013927, KSA7013929, KSA7013930, KSA7013931, KSA7013936, KSA7013945, KSA7013951, KSA7013960, KSA7013961, KSA7013965, KSA7013966, KSA7013970, KSA7013981, KSA7013988, KSA6477127, KSA6477174, KSA6477226, KSA6477245, KSA6477252, KSA6569835, KSA6664030, KSA6664033, KSA6664086, KSA6866930, KSA6866949, KSA6866982, KSA6477200, KSA6477117, KSA6477146, KSA6477190, KSA6477229, KSA6536456, KSA6945124, KSA6945139, KSA6945180, KSA6945216, KSA6945263, KSA6945284, KSA6017386, KSA6017421, KSA6017437, KSA6017510, KSA6945227, KSA6945242, KSA6945256, KSA6945264, KSA6945299, KSA6945308, KSA6945338, KSA6536461, KSA6536464, KSA6536468, KSA6536475, KSA6536477, KSA6536483, KSA6536491, KSA7162564, KSA7162578, KSA7162636, KSA7162652, KSA6664017, KSA6664018, KSA6866905, KSA6945119, KSA6945143, KSA6945150, KSA6945164, KSA6945239, KSA6945243, KSA6945282, KSA6945316, KSA6945354, KSA6945373, KSA6623193, KSA6623196, KSA6623211, KSA6623213, KSA6623215, KSA6623217, KSA6623224, KSA6623227, KSA6623252, KSA6623254, KSA6623257, KSA6623271, KSA6626620, KSA6626647, KSA6626669, KSA5646290, KSA6477220, KSA6649208, KSA6866960, KSA6808916, KSA6536519, KSA6409201, KSA6945162, KSA6649220, KSA6866902, KSA6866916, KSA6866937, KSA6649202, KSA6649249, KSA6649262, KSA6623283, KSA6590123, KSA6590124, KSA6590126, KSA6590138, KSA6590139, KSA6590140, KSA6590141, KSA6590161, KSA6590172, KSA6590174, KSA6590177, KSA6590188, KSA6590191, KSA6590197, KSA6590199, KSA6590208, KSA6590211, KSA6623207, KSA6623233, KSA6623256, KSA6623260, KSA6623263, KSA6623277, KSA6623279, KSA6623281, KSA6623284, KSA6664092, KSA6808871, KSA6808890, KSA6808894, KSA6808903, KSA6808907, KSA6808917, KSA6409157, KSA6409237, KSA6409247, KSA6477109, KSA6477114, KSA6477144, KSA6477151, KSA6477181, KSA6477197, KSA6477231, KSA6477240, KSA6477250, KSA6536529, KSA7550151, KSA7550173, KSA7574767, KSA7574784, KSA7590252, KSA7590258, KSA7590260, KSA7590263, KSA7590266, KSA7667307, KSA7667318, KSA7667321, KSA7667323, KSA7667332, KSA6664088, KSA7590242, KSA7590264, KSA7667311, KSA7667325, KSA7667326, KSA7667328, KSA7667333, KSA7667334, KSA7667306, KSA7667314, KSA7667315, KSA7667319, KSA7667327, KSA7667335, KSA6945159, KSA6808899, KSA7442791, KSA7442841, KSA7667309, KSA7667313, KSA7667329, KSA7676872, KSA7676873, KSA7676880, KSA7676891, KSA7676892, KSA7676894, KSA7676897, KSA7676898, KSA7676900, KSA7702195, KSA7702197, KSA7702202, KSA7702204, KSA7702205, KSA7702206, KSA7702207, KSA7702209, KSA7702211, KSA7702212, KSA7702213, KSA7702215, KSA7702216, KSA7702217, KSA7702218, KSA7702221, KSA6536441, KSA7574786, KSA7590241, KSA7590269, KSA7590246, KSA7590247, KSA7590267, KSA7667308, KSA7667312, KSA7667316, KSA7667322, KSA7676875, KSA7676877, KSA7676879, KSA7676881, KSA7676882, KSA7676884, KSA7676885, KSA7676887, KSA7676890, KSA7676895, KSA7702194, KSA7702200, KSA7702201, KSA7702203, KSA7702210, KSA7702214, KSA7702219, KSA7702220, KSA7702223, KSA7590249, KSA7590251, KSA7590257, KSA7590262, KSA7667310, KSA7667324, KSA7667330, KSA6623231, KSA6626671, KSA6626684, KSA6626703, KSA6664038, KSA6866912, KSA7120682, KSA7120732, KSA7120733, KSA7120749, KSA7120756, KSA7120774, KSA7162628, KSA7550171, KSA7574788, KSA6945255, KSA7520488, KSA6964960, KSA6899039, KSA6954214, KSA6954225, KSA6954228, KSA6954231, KSA6954233, KSA6954238, KSA6954239, KSA6954248, KSA6954258, KSA6954259, KSA6954262, KSA6954264, KSA6954266, KSA6954285, KSA6954289, KSA6954300, KSA6954304, KSA6954309, KSA6945362, KSA6649242, KSA6536463, KSA7727680, KSA7732508, KSA7732519, KSA6689341, KSA6689348, KSA6689362, KSA6689367, KSA6649189, KSA6649222, KSA6649224, KSA6649236, KSA6649237, KSA6649243, KSA6649253, KSA6649259, KSA6649264, KSA6649265, KSA7442807, KSA7727685, KSA7732514, KSA6689298, KSA6689308, KSA6689336, KSA6689340, KSA6689345, KSA6689357, KSA7520470, KSA7520472, KSA7536139, KSA7536145, KSA7536152, KSA7536154, KSA7536163, KSA7536165, KSA7536166, KSA7536168, KSA7442783, KSA7442795, KSA7442803, KSA7442806, KSA7442809, KSA7442820, KSA7442824, KSA7442826, KSA7442831, KSA7442832, KSA7442833, KSA7442835, KSA7442842, KSA7442860, KSA7442864, KSA7442875, KSA7442890, KSA6649229, KSA6569831, KSA6569893, KSA6569895, KSA6569899, KSA6536447, KSA6536451, KSA6536455, KSA6536462, KSA6536476, KSA6536478, KSA6536489, KSA6536498, KSA6536506, KSA6536508, KSA6536509, KSA6536513, KSA6536518, KSA6536533, KSA6536540, KSA6623186, KSA6623240, KSA6623269, KSA6649201, KSA6649215, KSA6649235, KSA6649260, KSA6649270, KSA6866901, KSA6664023, KSA7120689, KSA7120757, KSA6626627, KSA5646249, KSA5646255, KSA5646273, KSA5646279, KSA7391402, KSA7391434, KSA7391441, KSA7391457, KSA7391465, KSA7391474, KSA7391484, KSA7391488, KSA7391494, KSA7391497, KSA7391500, KSA7442776, KSA7442785, KSA7442793, KSA7442794, KSA7442796, KSA7442802, KSA7442805, KSA7442811, KSA7442823, KSA7442834, KSA7442844, KSA7442846, KSA7442855, KSA7442868, KSA7442869, KSA7442872, KSA7442884, KSA7442885, KSA7442889, KSA7442893, KSA7509842, KSA6945152, KSA7485765, KSA6649176, KSA6649180, KSA6649192, KSA6649194, KSA6649199, KSA6649200, KSA6649207, KSA6649248, KSA6866953, KSA6866958, KSA7120685, KSA7120697, KSA7120709, KSA7120711, KSA7120714, KSA7120726, KSA7120753, KSA7120765, KSA7120775, KSA6649174, KSA6649181, KSA6649187, KSA6649188, KSA6649190, KSA6649209, KSA6649210, KSA6649213, KSA6649231, KSA6649238, KSA6649239, KSA6649240, KSA6649245, KSA6649254, KSA6649271, KSA6623206, KSA6623219, KSA6623235, KSA6623245, KSA6945111, KSA6945135, KSA6945155, KSA6945186, KSA6945198, KSA6945339, KSA6945369, KSA6945380, KSA6899035, KSA7391416, KSA7391476, KSA7485740, KSA7590243, KSA6664015, KSA6664020, KSA6664027, KSA7239662, KSA7239666, KSA7239692, KSA7239719, KSA7239723, KSA7239724, KSA7239726, KSA7304375, KSA7304382, KSA7304401, KSA7304412, KSA7304419, KSA7304423, KSA7304429, KSA7304433, KSA7304436, KSA7304439, KSA7304446, KSA7304458, KSA7667320, KSA7676888, KSA7724022, KSA7724024, KSA7727684, KSA7727689, KSA7727690, KSA7727692, KSA7727695, KSA7727696, KSA7727700, KSA7727702, KSA7732506, KSA7732515, KSA7732520, KSA7732521, KSA7732522, KSA6954220, KSA6954230, KSA6954241, KSA6954274, KSA6954281, KSA6954287, KSA6954302, KSA7120680, KSA7120681, KSA7120686, KSA7120690, KSA7120706, KSA7120712, KSA7120722, KSA7120727, KSA7120729, KSA7120746, KSA7120760, KSA7120763, KSA7120772, KSA7096906, KSA7096913, KSA7096954, KSA7096991, KSA7096992, KSA6954270, KSA7162606, KSA7162646, KSA7162649, KSA7239650, KSA7239651, KSA7239658, KSA7239661, KSA7239663, KSA7239664, KSA7239681, KSA7239685, KSA7239745, KSA7304362, KSA7304399, KSA7304402, KSA7304408, KSA7304428, KSA7304445, KSA7485736, KSA7485739, KSA7485747, KSA7485764, KSA7495810, KSA7495818, KSA7495822, KSA7509828, KSA7509829, KSA7509832, KSA7509833, KSA7509835, KSA7509838, KSA7509840, KSA7509844, KSA7509847, KSA7509849, KSA7509854, KSA6569828, KSA6569855, KSA6569859, KSA6569878, KSA6569885, KSA6569902, KSA6409160, KSA6409166, KSA6409168, KSA6409182, KSA6409184, KSA6409185, KSA6409189, KSA6409190, KSA6409199, KSA6409200, KSA6409214, KSA6409218, KSA6409224, KSA6409229, KSA6409243, KSA6409250, KSA6409252, KSA6409254, KSA6569827, KSA6569832, KSA6569858, KSA6569869, KSA6409158, KSA6477103, KSA6477106, KSA6477107, KSA6477108, KSA6477119, KSA6477122, KSA6477125, KSA6477129, KSA6477133, KSA6477137, KSA6477140, KSA6477147, KSA6477149, KSA6477156, KSA6477167, KSA6477172, KSA6477180, KSA6477188, KSA6477189, KSA6477195, KSA6477202, KSA6477204, KSA6477211, KSA6477213, KSA6477215, KSA6477216, KSA6477217, KSA6477218, KSA6477230, KSA6477232, KSA6477233, KSA6477236, KSA6477238, KSA6477239, KSA6477241, KSA6477242, KSA6477243, KSA6477247, KSA6477249, KSA6536446, KSA6536448, KSA6536450, KSA6536453, KSA6536457, KSA6536465, KSA6536466, KSA6536473, KSA6536479, KSA6536481, KSA6536486, KSA6536495, KSA6536497, KSA6536499, KSA6536507, KSA6536539, KSA6569820, KSA6569824, KSA6569834, KSA6569836, KSA6569843, KSA6569857, KSA6569865, KSA6569870, KSA6569886, KSA6569892, KSA6590179, KSA7239693, KSA6419983, KSA6419992, KSA6419995, KSA6420013, KSA6420014, KSA6420021, KSA6420029, KSA6420032, KSA6420041, KSA6420044, KSA6420045, KSA6420051, KSA6420063, KSA6420070, KSA6420071, KSA6420079, KSA6420084, KSA6420095, KSA6420107, KSA6420109, KSA6420111, KSA6420119, KSA6420120, KSA6419986, KSA6420010, KSA6420025, KSA6420035, KSA6420042, KSA6420043, KSA6420046, KSA6420048, KSA6420050, KSA6420056, KSA6420060, KSA6420065, KSA6420072, KSA6420085, KSA6420090, KSA6420097, KSA6420098, KSA6420101, KSA6420118, KSA6477166, KSA6477201, KSA6419988, KSA6420005, KSA6420008, KSA6420009, KSA6420015, KSA6420016, KSA6420017, KSA6420018, KSA6420031, KSA6420034, KSA6420036, KSA6420037, KSA6420040, KSA6420047, KSA6420052, KSA6420055, KSA6420059, KSA6420062, KSA6420068, KSA6420069, KSA6420073, KSA6420086, KSA6420087, KSA6420093, KSA6420099, KSA6420105, KSA6420112, KSA6420113, KSA6420114, KSA6420116, KSA6420121, KSA6420122, KSA6477099, KSA6477100, KSA6477118, KSA6477123, KSA6477130, KSA6477132, KSA6477134, KSA6477138, KSA6477141, KSA6477160, KSA6477171, KSA6477173, KSA6477198, KSA6477209, KSA6477219, KSA6477221, KSA6477223, KSA6477224, KSA6477235, KSA6664073, KSA6419985, KSA6419989, KSA7536151, KSA7550153, KSA7550154, KSA7550165, KSA7550167, KSA7550175, KSA6569826, KSA6569841, KSA6569842, KSA6569845, KSA6569860, KSA6569882, KSA6569889, KSA6569894, KSA6419984, KSA6419994, KSA6419996, KSA6420004, KSA6420007, KSA6420038, KSA6420061, KSA6420067, KSA6420075, KSA6420076, KSA6420104, KSA6569817, KSA6569822, KSA6569825, KSA6569830, KSA6569844, KSA6569848, KSA6569849, KSA6569852, KSA6569887, KSA6569901, KSA6623244, KSA7536143, KSA7536147, KSA7536150, KSA7536153, KSA7536161, KSA7536164, KSA7536167, KSA7391420, KSA7391433, KSA7391438, KSA7391442, KSA7391445, KSA7391466, KSA6866990, KSA7590245, KSA7550149, KSA7550150, KSA7550174, KSA7574769, KSA7574773, KSA7574775, KSA7574779, KSA7574781, KSA7574785, KSA7574790, KSA7590240, KSA7590256, KSA7590259, KSA6866900, KSA6866920, KSA6866929, KSA6866931, KSA6866945, KSA6866956, KSA6866957, KSA6866962, KSA6866973, KSA6866985, KSA6866995, KSA6866997, KSA6866999, KSA6945102, KSA6945107, KSA6945121, KSA6945160, KSA6945214, KSA6945225, KSA6945228, KSA6945238, KSA6945307, KSA6945326, KSA6945349, KSA6945360, KSA6945375, KSA6964880, KSA6964895, KSA6964896, KSA6964901, KSA6964910, KSA6964926, KSA6964932, KSA6964933, KSA6964953, KSA6964964, KSA6964966, KSA6689327, KSA6689344, KSA6689352, KSA6689363, KSA6689383, KSA6689300, KSA6689305, KSA6689306, KSA6689347, KSA6689355, KSA6689371, KSA6590142, KSA6590145, KSA6590152, KSA6590168, KSA6590186, KSA6590195, KSA6590206, KSA6590209, KSA5646264, KSA6017314, KSA6017324, KSA6017394, KSA6017398, KSA6017403, KSA6017408, KSA6017435, KSA6017508, KSA6689293, KSA6689323, KSA6689331, KSA6689346, KSA6689358, KSA6689360, KSA6689381, KSA6808867, KSA6808840, KSA6808845, KSA6808856, KSA6808858, KSA6808862, KSA6808865, KSA6808872, KSA6808875, KSA6808877, KSA6808885, KSA6808886, KSA6808889, KSA6808901, KSA6808906, KSA6808908, KSA6808911, KSA6808922, KSA6808929, KSA6808930, KSA7495817, KSA7520477, KSA7550177, KSA7120707, KSA7120762, KSA6689314, KSA6689350, KSA6866911, KSA6866914, KSA6866915, KSA6866918, KSA6866936, KSA6866959, KSA6866972, KSA6866978, KSA7442845, KSA7485737, KSA7485748, KSA7485760, KSA7485762, KSA6649177, KSA6649185, KSA6649232, KSA7574766, KSA7574780, KSA7724038, KSA7724039, KSA7485756, KSA7495832, KSA6649204, KSA6649223, KSA6945188, KSA6945217, KSA6945246, KSA6945251, KSA6945257, KSA6945305, KSA6945317, KSA6945322, KSA6945329, KSA6945343, KSA6945344, KSA6945351, KSA6945366, KSA6945368, KSA6945376, KSA6945378, KSA6945383, KSA6945392, KSA6945396, KSA7727674, KSA7769598, KSA7676883, KSA6689289, KSA6689294, KSA6689301, KSA6689303, KSA6689319, KSA6689320, KSA6689324, KSA6689325, KSA6689334, KSA6689335, KSA6689337, KSA6689376, KSA7520483, KSA6663994, KSA6663997, KSA6663998, KSA6664019, KSA6664022, KSA7096907, KSA7096917, KSA7096921, KSA7096951, KSA7096952, KSA7096967, KSA7096978, KSA7096988, KSA7096996, KSA6569805, KSA6569810, KSA6569818, KSA6569829, KSA6569837, KSA6569847, KSA6569850, KSA6569851, KSA6569861, KSA6569881, KSA6689286, KSA7096914, KSA7096919, KSA7096920, KSA7096930, KSA7096949, KSA7096956, KSA7096969, KSA7096982, KSA7096987, KSA7096990, KSA7096997, KSA6964913, KSA6964920, KSA6964938, KSA7391427, KSA7442849, KSA7536144, KSA7096902, KSA7096909, KSA7096911, KSA7096915, KSA7096916, KSA7096923, KSA7096933, KSA7096939, KSA7096960, KSA7096965, KSA7096966, KSA7096974, KSA7096975, KSA7096981, KSA7096983, KSA7096989, KSA7096995, KSA6964890, KSA6945133, KSA6964878, KSA6964885, KSA6964887, KSA6964894, KSA6964936, KSA6964969, KSA7120768, KSA7162639, KSA6899003, KSA7304377, KSA7304397, KSA7304400, KSA7391409, KSA7391411, KSA7391431, KSA7391444, KSA7391454, KSA7391458, KSA7391459, KSA7391467, KSA7391470, KSA7391485, KSA6664082, KSA6945166, KSA6945388, KSA6964875, KSA6964882, KSA6964921, KSA6964925, KSA6964955, KSA7120715, KSA7239652, KSA7239667, KSA7239674, KSA7239675, KSA7239679, KSA7239682, KSA7239691, KSA7239695, KSA7239699, KSA7239700, KSA7239701, KSA7239716, KSA7239721, KSA7239722, KSA7239739, KSA7239744, KSA7239749, KSA7304369, KSA7304374, KSA7304376, KSA7304384, KSA7304385, KSA7304404, KSA7304405, KSA7304407, KSA7304410, KSA7304417, KSA7304438, KSA7304442, KSA7304451, KSA6964906, KSA6964907, KSA6964923, KSA6964927, KSA6964928, KSA6964934, KSA6964937, KSA6964945, KSA6964956, KSA6964963, KSA7676889, KSA6866908, KSA6866944, KSA6866979, KSA6898998, KSA6899001, KSA6899006, KSA6899007, KSA6899009, KSA6899016, KSA6899025, KSA6899027, KSA6899031, KSA6899032, KSA6899046, KSA6899047, KSA6899052, KSA6899053, KSA6899057, KSA6899058, KSA6899061, KSA6899062, KSA6899064, KSA6899066, KSA6899071, KSA6899075, KSA6899077, KSA6899079, KSA6899081, KSA6899087, KSA6899090, KSA6899094, KSA6899095, KSA6899096, KSA7120692, KSA7120716, KSA7120739, KSA7120748, KSA7120691, KSA7120730, KSA7120734, KSA7120736, KSA7120742, KSA7120747, KSA7162568, KSA7162569, KSA7162579, KSA6964883, KSA6689379, KSA7120684, KSA7120698, KSA7162613, KSA7162654, KSA7162659, KSA6590127, KSA6590129, KSA6590130, KSA6590149, KSA6590151, KSA6590155, KSA6590173, KSA6590181, KSA6590185, KSA6590201, KSA6590202, KSA6590212, KSA6590214, KSA6590215, KSA6590222, KSA6536449, KSA7013928, KSA7013944, KSA7013958, KSA7013979, KSA7724018, KSA7727683, KSA7727693, KSA7727694, KSA7727697, KSA7727698, KSA7732498, KSA7732502, KSA7732504, KSA7732505, KSA7732511, KSA7732517, KSA7732518, KSA7732523, KSA7732524, KSA6689290, KSA6689304, KSA6689318, KSA6689354, KSA6689361, KSA6689368, KSA6689372, KSA6689377, KSA6623234, KSA6626628, KSA6626638, KSA6626683, KSA6626688, KSA6590132, KSA6590134, KSA6590135, KSA6590144, KSA6590146, KSA6590147, KSA6590156, KSA6590157, KSA6590158, KSA6590159, KSA6590160, KSA6590166, KSA6590169, KSA6590175, KSA6590176, KSA6590187, KSA6590192, KSA6590194, KSA6590198, KSA6590200, KSA6590203, KSA6590204, KSA6590205, KSA6590207, KSA6590213, KSA6590218, KSA6590219, KSA6420058, KSA6569807, KSA6569815, KSA6569819, KSA6569854, KSA6569872, KSA6569874, KSA6569877, KSA6569897, KSA6536470, KSA6536480, KSA6536482, KSA6536531, KSA6536536, KSA6623200, KSA6623201, KSA6623218, KSA6626652, KSA6626656, KSA6626662, KSA6626674, KSA6626677, KSA6626680, KSA6626693, KSA6420000, KSA6420006, KSA6420027, KSA6420049, KSA6420054, KSA6420066, KSA6420094, KSA6420100, KSA6866910, KSA6623242, KSA6945101, KSA6945196, KSA7485742, KSA7485746, KSA7485751, KSA7509836, KSA7520461, KSA7520464, KSA7520465, KSA7520479, KSA7520484, KSA7536140, KSA7536141, KSA7536148, KSA7536155, KSA7536156, KSA7536157, KSA7536162, KSA7550155, KSA6945114, KSA6945116, KSA6945123, KSA6945125, KSA6945126, KSA6945131, KSA6945134, KSA6945145, KSA6945346, KSA6945357, KSA6945363, KSA6945385, KSA7724031, KSA7727678, KSA7536160, KSA6536512, KSA6536517, KSA6536530, KSA6664072, KSA6649193, KSA6649196, KSA6649197, KSA6649198, KSA6649216, KSA6649218, KSA6649219, KSA6649225, KSA6649228, KSA6649230, KSA6649234, KSA6649252, KSA6649256, KSA6649273, KSA6664029, KSA6664031, KSA6664039, KSA6664041, KSA6664053, KSA6664071, KSA6664077, KSA6664079, KSA6664081, KSA6664084, KSA6664087, KSA6419999, KSA6808844, KSA6808854, KSA6808855, KSA6808913, KSA6808927, KSA6808934, KSA6649191, KSA6954215, KSA6954223, KSA6954226, KSA6954232, KSA6954237, KSA6954242, KSA6954244, KSA6954247, KSA6954251, KSA6954253, KSA6954257, KSA6954260, KSA6954261, KSA6954268, KSA6954273, KSA6954280, KSA7162563, KSA6964940, KSA6945265, KSA6954222, KSA6954229, KSA6954249, KSA6954252, KSA6954282, KSA6954293, KSA6954294, KSA6954298, KSA6954299, KSA6954305, KSA7120719, KSA7391405, KSA7391417, KSA7391421, KSA7391423, KSA7391425, KSA7391426, KSA7391448, KSA7391471, KSA7391487, KSA7391499, KSA6664050, KSA6664068, KSA7013971, KSA7013982, KSA7013983, KSA7724034, KSA6409162, KSA6409170, KSA6409174, KSA6409194, KSA6409196, KSA6409239, KSA6409246, KSA6409248, KSA7485741, KSA7442798, KSA7442840, KSA7442865, KSA7442894, KSA6477101, KSA6477105, KSA6477110, KSA6477121, KSA6477131, KSA6477158, KSA6477162, KSA6477165, KSA6477169, KSA6477170, KSA6477178, KSA6477182, KSA6477186, KSA6477187, KSA6477203, KSA6477207, KSA6477214, KSA6477227, KSA6477244, KSA6477246, KSA6477251, KSA6477253, KSA6569862, KSA6569883, KSA6569884, KSA6649186, KSA6649217, KSA6945318, KSA6590153, KSA6945112, KSA6945151, KSA6945222, KSA6945252, KSA6945274, KSA6945279, KSA6945335, KSA6945389, KSA6017323, KSA6017357, KSA6017436, KSA6017457, KSA6017464, KSA6017466, KSA6017472, KSA6017473, KSA6017476, KSA6017480, KSA6017486, KSA6017488, KSA6017489, KSA6017492, KSA6017501, KSA6017506, KSA6945106, KSA6945269, KSA6945287, KSA7744935, KSA7744937, KSA7744938, KSA7744941, KSA7744943, KSA7744946, KSA7744951, KSA7744952, KSA7744959, KSA7744964, KSA7239657, KSA7239653, KSA7239689, KSA6945248, KSA7727677, KSA6945202, KSA6945210, KSA6945231, KSA6945241, KSA6945245, KSA6945281, KSA6945306, KSA6663996, KSA6664004, KSA6664009, KSA6664010, KSA6664012, KSA6664028, KSA6664032, KSA6664035, KSA6664056, KSA6664061, KSA6664080, KSA6664011, KSA6664034.
Other devices under the same FDA submission
2 devices cleared under K103142 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Ultraview SL Ultraview SL®Spacelabs Healthcare (Washington), Inc · SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE2 recalls2 reports
- Ultraview SL UV SL Command ModuleSpacelabs Healthcare (Washington), Inc · DETECTOR AND ALARM, ARRHYTHMIA1 recall3 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98938).
- Z-2242-2026 · Open, ClassifiedGE HealthCare CARESCAPE TELEMETRY SERVER V5 ATO MODEL, REF #2063702-101, Product Code SS9; used in conjunction with ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiolo…
- Z-2243-2026 · Open, ClassifiedGE HealthCare APEXPRO v5.0 - ARK2250 - Software Disaster Recovery USB Kit - Orderable by GE Service only, REF #2108523-01; ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, P…
- Z-2244-2026 · Open, ClassifiedGE HealthCare IOU ASSY ARK2250 CARESCAPE TELEMETRY SERVER, REF #2108729-901, Product Code SS9; ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With A…
- Z-2245-2026 · Open, ClassifiedGE HealthCare CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICA, REF #5697603, Product Code KSA; used in conjunction wtih ApexPro CARESCAPE Telemetry Server (CTS) v6.0; Monitor,…
- Z-2247-2026 · Open, ClassifiedGE HealthCare ApexPro v6.0 eDelivery Software, REF #5700026, Product Code SUV; used in conjunction wtih ApexPro CARESCAPE Telemetry Server (CTS) v6.0; Monitor, Physiological, Patie…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page