Multi-Snare Set: 5 mm x 125 cm, REF: 147305 recallZ-2526-2026 · Pfm Medical Inc. · initiated Apr 21, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Pfm Medical Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A snare catheter including luer lock, indicated for removal or manipulation of a foreign body in the vascular system, failed to meet the biocompatibility requirements for in-vitro-cytotoxicity, which if in contact with a patient's vasculature could result in localized inflammation and tissue irritation.

FDA root cause category: Under Investigation by firm

What the company is doing

On 4/21/2026, recall notices were emailed to customers who were asked to do the following: 1. Inform all customers (hospitals and clinics) and device users (physicians) who have received or used affected batches. This notification should also be shared with any organizations where the potentially affected devices have been transferred. 2. Return unused devices from the affected batches to PFM Medical, Inc. 3. Complete the enclosed Recall Response Form and email back to [email protected]

Recall details

Recall number
Z-2526-2026
Status
Open, Classified
Initiated by the company
Apr 21, 2026
Posted by the FDA
Jun 23, 2026
Recalling firm
Pfm Medical Inc.
Quantity
32
Product code
DXE
FDA submissions
K011783
Distribution
US Nationwide distribution in the states of NY, MD.

Product description

Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310

Code information (models, lots, serial numbers)

REF/UDI-DI/Lot(Expiration): 147305/04042301061829/1049868(10/14/2028); 147310/04042301033567/1049960(10/20/28)

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.