AIRSEAL recallZ-2473-2026 · ConMed Corporation · initiated May 14, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from ConMed Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for overpressure alerts.
FDA root cause category: Process change control
What the company is doing
ConMed notified consignees on about 05/14/2026 via emailed and mailed letter titled URGENT: Medical Device Correction. Consignees were made aware of the issue and risk associated with the issue. They were instructed that a replacement AirSeal disposable should be available prior to procedures in order to minimize potential delays. Before initiating insufflation, the user should ensure the tubing is properly connected between the AirSeal tube set and Airseal Access Ports. If the connection cannot be securely tightened and the overpressure alert cannot be resolved, please replace the affected device. Detailed instructions were provided on how to connect the AirSeal tube set and Airseal Access Ports. Consignees were also instructed to review inventory for affected units, place a copy of the provided Instructions for Use, Part No. AS-DIS-IFU, Rev. B with any inventory in possession with an active expiration date, contact all applicable departments and facilities that may have received affected products for awareness, and complete and return the medical device response form. If further information relevant to this issue becomes necessary CONMED will communicate as such. On about 06/04/2026 ConMed expanded their recall to include internationally distributed units. The letter sent included the same instructions.
Recall details
- Recall number
- Z-2473-2026
- Status
- Open, Classified
- Initiated by the company
- May 14, 2026
- Posted by the FDA
- Jun 17, 2026
- Recalling firm
- ConMed Corporation
- Quantity
- 71,151 packs (426,908 eaches)
- Product code
- HIF
- FDA submissions
- K143404
- Distribution
- Nationwide distribution. International distribution to Australia, Austria, Belgium, Bosnia and Herzegovina, Canada, CHILE, CHINA, COLOMBIA, Croatia, Denmark, DOMINICAN REPUBLIC, ECUADOR, Egypt, Estonia, Finland, France, Germany, Greece, HONG KONG, India, INDONESIA, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Libya, Luxembourg, MALAYSIA, Malta, Montenegro, Netherlands, NEW ZEALAND, Norway, Oman, Palestine, PANAMA, PHILIPPINES, Poland, Qatar, Romania, Russia, Saudi Arabia, Serbia, SINGAPORE, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, TAIWAN,PROVINCE OF CHINA, THAILAND, Turkey, Ukraine, UNITED ARAB EMIRATES, United Kingdom. .
Product description
ConMed AirSeal 12mm Access Port and Low Profile Obturator with Bladeless Optical Tip, 120mm Length. Catalog Number: iAS12-120LPI. 6 units per pack. Disposable Endoscopic Trocar and Cannula; Carbon Dioxide Insufflator for Laparoscopy.
Code information (models, lots, serial numbers)
Show the full code information (6k characters)
Catalog Number: iAS12-120LPI. UDI-DI: 10845854045688. Lot Numbers: 202506164, 202507144, 202506274, 202506136, 202506124, 2025071010, 202507114, 202506244, 202505194, 202505164, 202505064, 202505144, 202509114, 202508204, 202509124, 202510184, 202509274, 202511064, 202509254, 202509264, 202510254, 202510274, 202508294, 202509104, 202508214, 202509154, 202508084, 202507264, 202507254, 2025080611, 202507284, 202512124, 202511204, 202511214, 202511244, 202512134, 2025100310, 202510034, 202510154, 202511065, 202511075, 202511074, 202510244, 202510174, 202502214, 202503204, 202503075, 202503064, 202504244, 202504025, 2025040310, 2025040411, 202504046, 202503215, 202503244, 202504044, 202501164, 202501244, 202501235, 202506134, 202406204, 202406245, 202408154, 202502194, 202501145, 202505054, 202504036, 202412134, 202501274, 202408095, 202407316, 202410305, 202410244, 202506254, 202508274, 202509244, 202410234, 202411045, 202409035, 202407106, 202506304, 202512104, 202510204, 202501225, 202406214, 2025040210, 2025042110, 2025080810, 202508264, 202408135, 202410315, 202410295, 202408214, 202408294, 202507234, 202505304, 202505294, 202407155, 202408204, 202408206, 202406195, 202503065, 202502244, 202501095, 202504305, 202511086, 2024111911, 202501144, 2025110410, 2025110510, 202510164, 202511104, 2025122310, 202504304, 202501166, 202502184, 202407104, 202406215, 202406206, 202506264, 202410255, 202508284, 202507244, 202408015, 202505154, 202407305, 202411236, 2024112210, 202412105, 2025110512, 202501224, 202512154, 202407114, 202504294, 2025111910, 202407105, 202406266, 202501104, 202501275, 202503086, 2025040211, 2025111810, 2025041610, 2025110511, 202411054, 202408164, 202507104, 202408065, 202501286, 202503114, 202503214, 202412235, 202506024, 2024112111, 2024112010, 202505274, 202410304, 202411015, 202410225, 202408125, 202408205, 202409034, 202408165, 202510035, 202408155, 202408296, 202407154, 202407124, 202408064, 202406284, 202406205, 202601064, 2025122210, 2025111410, 202511066, 202411265, 202411225, 202411224, 202506104, 202408176, 202410294, 202512234, 202407015, 202407014, 202407125, 202504224, 202505014, 202411195, 202502064, 202411014, 202408094, 202411216, 202411214, 202504214, 202507094, 202406285, 202406244, 202502205, 202503205, 2024112610, 202412144, 202410235, 202411255, 202501246, 202408106, 202408224, 202407254, 202407024, 202504056, 202503245, 202501245, 202502075, 202501234, 202502114, 2024112011, 202411206, 202412135, 202406275, 202501024, 202411276, 202412124, 2024112211, 2024121210, 2025102310, 202512114, 202601055, 2025080710, 202506114, 202505284, 202505176, 202407314, 202408144, 202408134, 202410245, 202411194, 202411254, 2024121310, 202501025, 202501155, 202412114, 202408084, 202501086, 202502204, 202501034, 202412226, 202406254, 202509014, 2025100210, 202511076, 202503195, 202412164, 202407256, 202407116, 202411226, 202406286, 202407136, 202408194, 2025081110, 202411205, 202408146, 202411264, 202412234, 202411215, 202406226, 2025040710, 202501256, 202502055, 202410314, 202410284, 202408295, 202408145, 202408136, 202501154, 202502105, 202501236, 202411266, 202408124, 2024121311, 202503105, 2025040311, 202406255, 202407095, 202412125, 2024112510, 2025043010, 2024112110, 2024112511, 202408156, 202501226, 202408216, 202501085, 202407026, 2026010511, 2024071212, 202407126, 202410285, 202408166, 202406256, 202504174, 2024121211. *OUS added 06/10/2026 expansion: 202411204, 202511075, 202511244, 202509154, 202509104, 202509264, 202508084, 2025080611, 202507284, 202507254, 202507264, 202504174, 202506274, 2025040310, 202504244, 202505144, 202501085, 202501094, 202504164, 202502214, 2024112110, 202410245, 202410285, 202411044, 2024112011, 202412105, 202411266, 202510274, 202510184, 202406226, 202406216, 202407024, 202407136, 202408066, 2024071212, 202511064, 202511065, 202509274, 202508214, 202510254, 202506124, 202503204, 202504044, 202505064, 202503075, 202501244, 202501154, 202407026, 202408194, 202407116, 202511214, 2025071010, 202505164, 202410315, 202501166, 202512124, 202408304, 202410225, 202408135, 202511074, 202509124, 202510244, 2025100310, 202507144, 202505194, 2025041710, 202501084, 202406255, 202407126, 202408195, 202411214, 202408084, 202508204, 2025040410, 202411225, 202509114, 202503064, 202406256, 202407245, 202412124, 202511204, 202509254, 202510154, 202510034, 202504046, 2025040411, 202501235, 202501256, 202510035, 202512154, 202503065, 202407105, 202408144, 202506114, 202510204, 202512114, 202503245, 202501034, 202501224, 202411215, 202507244, 202509014, 202509244, 202508284, 202511086, 202511104, 202601064, 2024121210, 202408216, 202412234, 202502114, 202504056, 202411194, 2025043010, 202506134, 202507114, 202406245, 202408164, 2025080810, 202409034, 202508264, 202406214, 202408295, 202410294, 202504304, 202506164, 202506304, 202502105, 2024121311, 202410305, 202408146, 202504214, 202501164, 202601124, 202507234, 2025041610, 202501145, 202410284, 202408064, 202408136, 202406204, 202501024, 202501025, 2024121310, 202505284, 202501234, 202411195, 202411205, 202408166, 202408176, 202505054, 202410295, 202408155, 202407154, 202407114, 202407125, 202508294, 2025110512, 202507104, 2025042110, 202502075, 202506024, 202501245, 202410304, 202410314, 2024112210, 202512234, 202505014, 202510164, 202406254, 202505304, 202502055, 2025110510, 202506136, 202506244, 202408156, 202508274, 202506254, 202504224, 202501236, 2024112211, 202408124, 202503215, 202501286, 202412144, 2024112511, 202408145, 202408065, 202501226, 202412125, 2025110410, 202510174, 202504025, 202503244, 202504036, 2025040210, 202408165. Manufactured from 14 June 2024 to 12 January 2026.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- AIRSEAL IAS12-120LPIConmed Corporation · Laparoscope, general & plastic surgery1 recall19 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 98854).
- Z-2464-2026 · Open, ClassifiedConMed AirSeal 5 mm Smooth Access Port and Obturator with Blunt Tip, 150mm Length. Catalog Number: iASB5-150. 6 units per pack. Disposable Endoscopic Trocar and Cannula; Carbon Dio…
- Z-2465-2026 · Open, ClassifiedConMed AirSeal 12 mm Access Port and Obturator with Blunt Tip, 120mm Length. Catalog Number: iASB12-120. 6 units per pack. Disposable Endoscopic Trocar and Cannula; Carbon Dioxide …
- Z-2466-2026 · Open, ClassifiedConMed AirSeal 12mm Access Port and Obturator with Blunt Tip, 100mm Length. Catalog Number: iASB12-100. 6 units per pack. Disposable Endoscopic Trocar and Cannula; Carbon Dioxide I…
- Z-2467-2026 · Open, ClassifiedConMed AirSeal 8mm Access Port and Low Profile Obturator with Bladeless Optical Tip, 120mm Length. Catalog Number: iAS8-120LP. 6 units per pack. Disposable Endoscopic Trocar and Ca…
- Z-2468-2026 · Open, ClassifiedConMed AirSeal 8mm Access Port and Low Profile Obturator with Bladeless Optical Tip, 100mm Length. Catalog Number: iAS8-100LP. 6 units per pack. Disposable Endoscopic Trocar and Ca…
- Z-2469-2026 · Open, ClassifiedConMed AirSeal 5mm Access Port and Low Profile Obturator with Bladeless Optical Tip, 75mm Length. Catalog Number: iAS5-75LP. 6 units per pack. Disposable Endoscopic Trocar and Cann…
- Z-2470-2026 · Open, ClassifiedConMed AirSeal 5mm Access Port and Low Profile Obturator with Bladeless Optical Tip, 120mm Length. Catalog Number: iAS5-120LP. 6 units per pack. Disposable Endoscopic Trocar and Ca…
- Z-2471-2026 · Open, ClassifiedConMed AirSeal 5mm Access Port and Low Profile Obturator with Bladeless Optical Tip, 100mm Length. Catalog Number: iAS5-100LP. 6 units per pack. Disposable Endoscopic Trocar and Ca…
- Z-2472-2026 · Open, ClassifiedConMed AirSeal 12mm Access Port and Palm Grip Obturator with Bladeless Optical Tip, 150mm Length. Catalog Number: iAS12-150LPI. 6 units per pack. Disposable Endoscopic Trocar and C…
- Z-2474-2026 · Open, ClassifiedConMed AirSeal 12mm Access Port and Palm Grip Obturator with Bladeless Optical Tip, 100mm Length. Catalog Number: iAS12-100LPI. 6 units per pack. Disposable Endoscopic Trocar and C…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page