Brand Name: Dexcom G7 Continuous Glucose Monitoring System recallZ-2574-2026 · Dexcom, Inc. · initiated May 27, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Dexcom, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Certain lots of Dexcom G7 sensors originally designated as scrap and intended for destruction were stolen during the destruction process. The sensors were later sold to pharmacies and distributors by Pharmsource, LLC. Lot 1725204004 were designated for destruction because the G7 sensors were not properly sterilized and have an increased risk of skin infection if used. Lot 1725069002 were designated for destruction due to an increased risk of having no sensor readings available or experiencing sensor inaccuracy that includes the potential for missed detection of a hyperglycemic event or incorrect treatment decisions that can result in patient injury, including medical intervention to prevent permanent injury or impairment. Additionally, due to the unknown storage and handling and storage conditions these sensors may have undergone during the distribution, these sensors may present additional risk.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On 05/26-27/2026, the firm issued a Press Release and sent an "URGENT: MEDICAL DEVICE CORRECTION NOTIFICATION" informing customers and the public that Dexcom is aware that two lots (1725204004 and 1725069002) of Dexcom G7 sensors, originally designated as scrap and intended for destruction, were stolen and distributed. For Patients/Healthcare Providers: 1. Check lot number on G7 sensor outer box and compare with the listed lot numbers in this notification. If the lot number is listed, the device is affected. If affected, skip to step 3 below to document affected serial number(s). 2. Determine if current or unused sensor(s) is affected based on serial number by visiting www.dexcom.com/theft-check and enter serial number into the serial number look-up. Reference https://www.dexcom.com/faqs/how-to-find-dexcom-g7-serial-number for instructions. 3. If you have affected product(s), immediately discontinue use and dispose of them. Contact Dexcom Technical Support at 1-844-478-1600 with questions or to request a no-cost replacement. Important: Keep record of affected serial number for traceability purposes This is required to confirm no cost replacement. 4. Replace the affected sensor with an unaffected sensor for glucose monitoring. If you do not have an unaffected G7 sensor, use a blood glucose meter for treatment decisions until you receive a replacement sensor. 5. Contact healthcare provider if you have symptoms of infection or inflammation redness, swelling, or pain at the insertion site. Independent Pharmacies: Check inventory if purchased G7 sensors from Pharmsource, LLC. If you have product from the lots listed, immediately dispose of the product. Contact Dexcom with any related inquiries at [email protected].
Recall details
- Recall number
- Z-2574-2026
- Status
- Open, Classified
- Initiated by the company
- May 27, 2026
- Posted by the FDA
- Jun 24, 2026
- Recalling firm
- Dexcom, Inc.
- Product code
- QBJ
- FDA submissions
- K240902
- Distribution
- Dexcom does not have distribution records for this product due to its unauthorized release. However Dexcom is able to identify sensor session starts for the affected lots through the Dexcom G7 CGM App user performance data. Based on App user performance data, Lot 1725204004 has sensor session starts in the United States. Based on App user performance data, Lot 1725069002 has sensor session starts in the United States and the countries of Australia, Austria, Canada, Czech Republic, Finland, France, Germany, Hong Kong, Israel, Italy, Japan, South Korea, Kuwait, Netherlands, Poland, Portugal, Qatar, Singapore, Spain, Switzerland, Turkey, United Arab Emirates,and the United Kingdom.
Product description
Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US Retail Sensor/Transmitter 1-pack Model/Catalog Number: STP-AT-012 Software Version: N/A Product Description: Dexcom G7 US Retail Sensor/Transmitter 1-pack Component: Dexcom G7 Continuous Glucose Monitoring System
Code information (models, lots, serial numbers)
Lot Code: UDI-DI: 00386270004109 Lot Numbers: 1725204004 and 1725069002
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Dexcom G7 GSS (161) STP-AT-012Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall5 deaths48,180 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.