U2 Total Knee System recallZ-2883-2026 · United Orthopedic Corp. · initiated May 20, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from United Orthopedic Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail.
FDA root cause category: Process design
What the company is doing
On 03/11/2026, the firm emailed an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that due the fact that the PSA Box Trial cannot be properly assembled with the Hinge Trial in the affected devices because of incomplete processing of the screw components, which may result in interference during assembly with the corresponding device. On 05/20/2026, the firm emailed an updated "URGENT: MEDCIAL DEVICE RECALL" Letter Customers are instructed: 1. Examine your inventory immediately to locate the affected product whose quantities and lot numbers are listed above. Please note the product may have already been returned. 2. Cease all use of these devices immediately and quarantine all affected devices in your territory for immediate return. 3. Customer Service will be contacting you regarding replacements. If you still have the product, you are requested to immediately return the affected product using the prepaid airbills on RMA XXXX . a. Write the RMA number on the outside of the box clearly and legibly. b. Ship recalled product via Fed Ex standard overnight to: UOC USA INC. Attention: RMA# XXXX 16601 Hale Ave. Irvine, CA 92606 (Attached Shipping Label) 4. Additionally, if your firm has further distributed the recall product, please provide that customer with a copy of this recall notification and assist them with the return of the product. 5. Complete and return the following page according to the instructions. For Question - contact UOC Sr. Director of Sales Operations at 949.328.3366.
Recall details
- Recall number
- Z-2883-2026
- Status
- Open, Classified
- Initiated by the company
- May 20, 2026
- Posted by the FDA
- Aug 5, 2026
- Recalling firm
- United Orthopedic Corp.
- Quantity
- 2
- Product code
- JWH
- FDA submissions
- K082424
- Distribution
- US Nationwide distribution in the states of Texas and Michigan.
Product description
Brand Name: U2 Total Knee System Product Name: PSA Box Trial, L, #1-6 and PSA Box Trial, R, #1-6 Model/Catalog Number: 9323-5511 thru 9323-5516 and 9323-5521 thru 9323-5526 Software Version: N/A Product Description: Knee femur prosthesis trial, reusable Component: No, N/A
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Product Number; UDI-DI Code; Lot Code: Product Number 9323-5511; UDI-DI 04719886906929; Lot number 25FB54 Product Number 9323-5512; UDI-DI 04719886906936; Lot number 25FB55 Product Number 9323-5513; UDI-DI 04719886906943; Lot number 25FB56 Product Number 9323-5514; UDI-DI 04719886906950; Lot number 25FB57 Product Number 9323-5515; UDI-DI 04719886906967; Lot number 25FB58 Product Number 9323-5516; UDI-DI 04719886906424; Lot number 25FB59 Product Number 9323-5521; UDI-DI 04719886906974; Lot number 25FB29 Product Number 9323-5522; UDI-DI 04719886906981; Lot number 25FB30 Product Number 9323-5523; UDI-DI 04719886906998; Lot number 25FB31 Product Number 9323-5524; UDI-DI 04719886907001; Lot number 25FB32 Product Number 9323-5525; UDI-DI 04719886907018; Lot number 25FB33 Product Number 9323-5526; UDI-DI 04719886907025; Lot number 25FB34
Devices named in this recall
12 devices in the FDA device registry whose identifier appears in the recall's code information.
- U2 Total Knee System 9323-5516United Orthopedic Corp. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall0 reports
- U2 Total Knee System 9323-5511United Orthopedic Corp. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall0 reports
- U2 Total Knee System 9323-5512United Orthopedic Corp. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall0 reports
- U2 Total Knee System 9323-5513United Orthopedic Corp. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall0 reports
- U2 Total Knee System 9323-5514United Orthopedic Corp. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall0 reports
- U2 Total Knee System 9323-5515United Orthopedic Corp. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall0 reports
- U2 Total Knee System 9323-5521United Orthopedic Corp. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall0 reports
- U2 Total Knee System 9323-5522United Orthopedic Corp. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall0 reports
- U2 Total Knee System 9323-5523United Orthopedic Corp. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall0 reports
- U2 Total Knee System 9323-5524United Orthopedic Corp. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall0 reports
- U2 Total Knee System 9323-5525United Orthopedic Corp. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall0 reports
- U2 Total Knee System 9323-5526United Orthopedic Corp. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 99026).
- Z-2884-2026 · Open, ClassifiedBrand Name: USTAR II Knee System Product Name: Hinge Box Trial, L, #1-6 and Hinge Box Trial, R, #1-6 Model/Catalog Number: 9325-5711 thru 9325-5716 and 9325-5721 thru 9325-5726 Sof…
- Z-2885-2026 · Open, ClassifiedBrand Name: U2 Total Knee System Product Name: Femoral valgus adaptor, left and right Model/Catalog Number: 9323-5333 and 9323-5334 Software Version: N/A Product Description: Manua…
- Z-2886-2026 · Open, ClassifiedBrand Name: U2 Total Knee System Product Name: Femoral valgus offset adaptor, 2mm, 4mm, 6mm, left and Femoral valgus offset adaptor, 2mm, 4mm, 6mm, right Model/Catalog Number: 9323…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page