Medline Convenience Kits recallZ-2453-2026 · Medline Industries, LP · initiated May 6, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Medline Convenience Kits contain Arrow Kits and Sets that include Lidocaine Hydrochloride Injection, USP and Bupivacaine Hydrochloride in Dextrose Injection, USP which were recalled due to quality issues identified during a recent FDA inspection of the supplier.

FDA root cause category: Nonconforming Material/Component

What the company is doing

An URGENT MEDICAL DEVICE RECALL notification letter dated 5/6/26 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-078-FG-FGX3 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].

Recall details

Recall number
Z-2453-2026
Status
Open, Classified
Initiated by the company
May 6, 2026
Posted by the FDA
Jun 24, 2026
Recalling firm
Medline Industries, LP
Quantity
6775 units
Product code
OEZ
Distribution
US Nationwide.

Product description

Medline Convenience Kits: 1) OPEN HEART CDS, Model Number: CDS840144Q; 2) OPEN HEART, Model Number: CDS840150W; 3) OPEN HEART ADULT, Model Number: CDS840396AA; 4) CABG A CDS, Model Number: CDS982269N; 5) OPEN HEART ANESTHESIA PACK, Model Number: CDS983849G; 6) OPEN HEART ACCESSORY CDS, Model Number: CDS984289L; 7) OPEN HEART A, Model Number: CDS985105Q; 8) OPEN HEART PART 2, Model Number: DYNJ901075R; 9) VPH OPEN HEART, Model Number: DYNJ902664L; 10) HEART A, Model Number: DYNJ903465I; 11) HEART, Model Number: DYNJ903822F; 12) OPEN HEART BASIC B, Model Number: DYNJ904311F; 13) SM OPEN HEART PACK A&B, Model Number: DYNJ904749J; 14) OPEN HEART ANESTHESIA ICC-B, Model Number: DYNJ905567B; 15) SILVER CROSS OPEN HEART, Model Number: DYNJ906102K; 16) ANESTHESIA OPEN HEART, Model Number: DYNJ906225I; 17) OPEN HEART COMBO, Model Number: DYNJ907400F; 18) CARDIAC CABG B, Model Number: DYNJ908121C; 19) FS OPEN HEART, Model Number: DYNJ908308A; 20) OPEN HEART SUPPLEMENT, Model Number: DYNJ909116D; 21) MEM CARDIAC, Model Number: DYNJ9622573AP; 22) OPEN HEART A & B PACK, Model Number: DYNJ9718008AT

Code information (models, lots, serial numbers)

Show the full code information (26k characters)

1) CDS840144Q, UDI-DI: 10195327263102(each), 40195327263103(case), Lot Number: 24LBH805; 2) CDS840144Q, UDI-DI: 10195327263102(each), 40195327263103(case), Lot Number: 24HBM100; 3) CDS840144Q, UDI-DI: 10195327263102(each), 40195327263103(case), Lot Number: 25ABR195; 4) CDS840144Q, UDI-DI: 10195327263102(each), 40195327263103(case), Lot Number: 25EBS810; 5) CDS840144Q, UDI-DI: 10195327263102(each), 40195327263103(case), Lot Number: 25FBS936; 6) CDS840144Q, UDI-DI: 10195327263102(each), 40195327263103(case), Lot Number: 25GBX096; 7) CDS840144Q, UDI-DI: 10195327263102(each), 40195327263103(case), Lot Number: 25IBT859; 8) CDS840144Q, UDI-DI: 10195327263102(each), 40195327263103(case), Lot Number: 24KBE879; 9) CDS840144R, UDI-DI: 10198459561627(each), 40198459561628(case), Lot Number: 26DBF234; 10) CDS840144R, UDI-DI: 10198459561627(each), 40198459561628(case), Lot Number: 26ABG176; 11) CDS840144R, UDI-DI: 10198459561627(each), 40198459561628(case), Lot Number: 26ABK484; 12) CDS840150W, UDI-DI: 10195327648046(each), 40195327648047(case), Lot Number: 24HBQ201; 13) CDS840150X, UDI-DI: 10198459112386(each), 40198459112387(case), Lot Number: 25CBD369; 14) CDS840150X, UDI-DI: 10198459112386(each), 40198459112387(case), Lot Number: 25DBL498; 15) CDS840150X, UDI-DI: 10198459112386(each), 40198459112387(case), Lot Number: 24LBD200; 16) CDS840150X, UDI-DI: 10198459112386(each), 40198459112387(case), Lot Number: 24LBD223; 17) CDS840150X, UDI-DI: 10198459112386(each), 40198459112387(case), Lot Number: 24LBR405; 18) CDS840150X, UDI-DI: 10198459112386(each), 40198459112387(case), Lot Number: 24JBX455; 19) CDS840150Y, UDI-DI: 10198459303029(each), 40198459303020(case), Lot Number: 25FBD604; 20) CDS840150Y, UDI-DI: 10198459303029(each), 40198459303020(case), Lot Number: 25FBT739; 21) CDS840150Y, UDI-DI: 10198459303029(each), 40198459303020(case), Lot Number: 25HBT720; 22) CDS840150Y, UDI-DI: 10198459303029(each), 40198459303020(case), Lot Number: 25JBE191; 23) CDS840150Y, UDI-DI: 10198459303029(each), 40198459303020(case), Lot Number: 25JBW220; 24) CDS840150Y, UDI-DI: 10198459303029(each), 40198459303020(case), Lot Number: 26ABP964; 25) CDS840150Y, UDI-DI: 10198459303029(each), 40198459303020(case), Lot Number: 26BBD028; 26) CDS840150Y, UDI-DI: 10198459303029(each), 40198459303020(case), Lot Number: 26BBF355; 27) CDS840150Y, UDI-DI: 10198459303029(each), 40198459303020(case), Lot Number: 26CBB249; 28) CDS840396AA, UDI-DI: 10198459348686(each), 40198459348687(case), Lot Number: 25FBD477; 29) CDS840396AA, UDI-DI: 10198459348686(each), 40198459348687(case), Lot Number: 25HBM750; 30) CDS840396AB, UDI-DI: 10198459468742(each), 40198459468743(case), Lot Number: 25IBD204; 31) CDS840396AB, UDI-DI: 10198459468742(each), 40198459468743(case), Lot Number: 25JBQ605; 32) CDS840396AB, UDI-DI: 10198459468742(each), 40198459468743(case), Lot Number: 25KBK109; 33) CDS840396AB, UDI-DI: 10198459468742(each), 40198459468743(case), Lot Number: 26ABD873; 34) CDS840396AB, UDI-DI: 10198459468742(each), 40198459468743(case), Lot Number: 26ABT718; 35) CDS840396AC, UDI-DI: 10198459639494(each), 40198459639495(case), Lot Number: 26DLA888; 36) CDS840396Y, UDI-DI: 10195327574932(each), 40195327574933(case), Lot Number: 25BBG163; 37) CDS840396Y, UDI-DI: 10195327574932(each), 40195327574933(case), Lot Number: 25BBP788; 38) CDS840396Y, UDI-DI: 10195327574932(each), 40195327574933(case), Lot Number: 25CBD493; 39) CDS840396Y, UDI-DI: 10195327574932(each), 40195327574933(case), Lot Number: 25CBK781; 40) CDS840396Y, UDI-DI: 10195327574932(each), 40195327574933(case), Lot Number: 24KBK977; 41) CDS840396Y, UDI-DI: 10195327574932(each), 40195327574933(case), Lot Number: 24IBK539; 42) CDS980952AN, UDI-DI: 10198459022531(each), 40198459022532(case), Lot Number: 25BBB644; 43) CDS980952AN, UDI-DI: 10198459022531(each), 40198459022532(case), Lot Number: 25BBR673; 44) CDS980952AN, UDI-DI: 10198459022531(each), 40198459022532(case), Lot Number: 24LBM011; 45) CDS980952AN, UDI-DI: 10198459022531(each), 40198459022532(case), Lot Number: 24LBR411; 46) CDS980952AN, UDI-DI: 10198459022531(each), 40198459022532(case), Lot Number: 24HBW376; 47) CDS980952AN, UDI-DI: 10198459022531(each), 40198459022532(case), Lot Number: 24JBW880; 48) CDS980952AO, UDI-DI: 10198459245237(each), 40198459245238(case), Lot Number: 25DBP914; 49) CDS980952AO, UDI-DI: 10198459245237(each), 40198459245238(case), Lot Number: 25EBO343; 50) CDS980952AO, UDI-DI: 10198459245237(each), 40198459245238(case), Lot Number: 25FBN895; 51) CDS980952AO, UDI-DI: 10198459245237(each), 40198459245238(case), Lot Number: 25FBO022; 52) CDS980952AO, UDI-DI: 10198459245237(each), 40198459245238(case), Lot Number: 25FBS633; 53) CDS980952AP, UDI-DI: 10198459370991(each), 40198459370992(case), Lot Number: 25HBD136; 54) CDS980952AQ, UDI-DI: 10198459478420(each), 40198459478421(case), Lot Number: 25IBN076; 55) CDS980952AQ, UDI-DI: 10198459478420(each), 40198459478421(case), Lot Number: 25IBN339; 56) CDS980952AR, UDI-DI: 10198459520440(each), 40198459520441(case), Lot Number: 25LBC894; 57) CDS980952AR, UDI-DI: 10198459520440(each), 40198459520441(case), Lot Number: 26ABP940; 58) CDS980952AR, UDI-DI: 10198459520440(each), 40198459520441(case), Lot Number: 26CBI600; 59) CDS982269N, UDI-DI: 10195327352998(each), 40195327352999(case), Lot Number: 25HBH657; 60) CDS982269N, UDI-DI: 10195327352998(each), 40195327352999(case), Lot Number: 25IBM477; 61) CDS982269N, UDI-DI: 10195327352998(each), 40195327352999(case), Lot Number: 26BBG451; 62) CDS983849G, UDI-DI: 10195327650148(each), 40195327650149(case), Lot Number: 25ABO737; 63) CDS983849G, UDI-DI: 10195327650148(each), 40195327650149(case), Lot Number: 25BBM406; 64) CDS983849G, UDI-DI: 10195327650148(each), 40195327650149(case), Lot Number: 25CBL909; 65) CDS983849G, UDI-DI: 10195327650148(each), 40195327650149(case), Lot Number: 25EBN768; 66) CDS983849G, UDI-DI: 10195327650148(each), 40195327650149(case), Lot Number: 25GBC295; 67) CDS983849G, UDI-DI: 10195327650148(each), 40195327650149(case), Lot Number: 25HBJ455; 68) CDS983849G, UDI-DI: 10195327650148(each), 40195327650149(case), Lot Number: 25JBJ254; 69) CDS983849G, UDI-DI: 10195327650148(each), 40195327650149(case), Lot Number: 25LBJ730; 70) CDS983849G, UDI-DI: 10195327650148(each), 40195327650149(case), Lot Number: 25LBQ452; 71) CDS983849G, UDI-DI: 10195327650148(each), 40195327650149(case), Lot Number: 24LBB436; 72) CDS983849G, UDI-DI: 10195327650148(each), 40195327650149(case), Lot Number: 26ABV988; 73) CDS983849G, UDI-DI: 10195327650148(each), 40195327650149(case), Lot Number: 26CBR016; 74) CDS984289L, UDI-DI: 10195327439064(each), 40195327439065(case), Lot Number: 24FBT233; 75) CDS984289L, UDI-DI: 10195327439064(each), 40195327439065(case), Lot Number: 24HBJ043; 76) CDS984289M, UDI-DI: 10198459118548(each), 40198459118549(case), Lot Number: 24KBN423; 77) CDS984289M, UDI-DI: 10198459118548(each), 40198459118549(case), Lot Number: 24LBA755; 78) CDS984289M, UDI-DI: 10198459118548(each), 40198459118549(case), Lot Number: 24IBH401; 79) CDS984289M, UDI-DI: 10198459118548(each), 40198459118549(case), Lot Number: 24JBM195; 80) CDS984289M, UDI-DI: 24KBA684(each), 10198459118548(case), Lot Number: 40198459118549; 81) CDS984289N, UDI-DI: 25BBR524(each), 10198459256769(case), Lot Number: 40198459256760; 82) CDS984289N, UDI-DI: 10198459256769(each), 40198459256760(case), Lot Number: 25CBG855; 83) CDS984289N, UDI-DI: 10198459256769(each), 40198459256760(case), Lot Number: 25FBD048; 84) CDS984289O, UDI-DI: 10198459418143(each), 40198459418144(case), Lot Number: 25GBY100; 85) CDS984289O, UDI-DI: 10198459418143(each), 40198459418144(case), Lot Number: 25IBB599; 86) CDS984289O, UDI-DI: 10198459418143(each), 40198459418144(case), Lot Number: 25IBC732; 87) CDS984289O, UDI-DI: 10198459418143(each), 40198459418144(case), Lot Number: 25IBL186; 88) CDS984289O, UDI-DI: 10198459418143(each), 40198459418144(case), Lot Number: 25KBB457; 89) CDS984355I, UDI-DI: 10195327569518(each), 40195327569519(case), Lot Number: 24ELA326; 90) CDS984355J, UDI-DI: 10198459096228(each), 40198459096229(case), Lot Number: 25ALA777; 91) CDS984355J, UDI-DI: 10198459096228(each), 40198459096229(case), Lot Number: 25BLA015; 92) CDS984355J, UDI-DI: 10198459096228(each), 40198459096229(case), Lot Number: 25CLA060; 93) CDS984355J, UDI-DI: 10198459096228(each), 40198459096229(case), Lot Number: 24LLA698; 94) CDS984355K, UDI-DI: 10198459258848(each), 40198459258849(case), Lot Number: 25JLA937; 95) CDS984355K, UDI-DI: 10198459258848(each), 40198459258849(case), Lot Number: 26ALA525; 96) CDS984355K, UDI-DI: 10198459258848(each), 40198459258849(case), Lot Number: 26BLA695; 97) CDS984355K, UDI-DI: 10198459258848(each), 40198459258849(case), Lot Number: 26CLA630; 98) CDS985105Q, UDI-DI: 10198459197925(each), 40198459197926(case), Lot Number: 25ABC137; 99) CDS985105Q, UDI-DI: 10198459197925(each), 40198459197926(case), Lot Number: 25BBD104; 100) CDS985105Q, UDI-DI: 10198459197925(each), 40198459197926(case), Lot Number: 25CBD560; 101) CDS985105Q, UDI-DI: 10198459197925(each), 40198459197926(case), Lot Number: 25DBS683; 102) CDS985105Q, UDI-DI: 10198459197925(each), 40198459197926(case), Lot Number: 25FBI319; 103) CDS985105Q, UDI-DI: 10198459197925(each), 40198459197926(case), Lot Number: 25GBJ383; 104) CDS985105Q, UDI-DI: 10198459197925(each), 40198459197926(case), Lot Number: 25HBJ140; 105) CDS985105R, UDI-DI: 10198459501661(each), 40198459501662(case), Lot Number: 26DBE048; 106) CDS985105R, UDI-DI: 10198459501661(each), 40198459501662(case), Lot Number: 26DBP065; 107) CDS985105R, UDI-DI: 10198459501661(each), 40198459501662(case), Lot Number: 25JBT358; 108) CDS985105R, UDI-DI: 10198459501661(each), 40198459501662(case), Lot Number: 25LBE124; 109) CDS985105R, UDI-DI: 10198459501661(each), 40198459501662(case), Lot Number: 26ABO369; 110) DYNJ901075R, UDI-DI: 10198459047947(each), 40198459047948(case), Lot Number: 25CBE435; 111) DYNJ901075R, UDI-DI: 10198459047947(each), 40198459047948(case), Lot Number: 25EBS829; 112) DYNJ901075R, UDI-DI: 10198459047947(each), 40198459047948(case), Lot Number: 25IBG152; 113) DYNJ901075R, UDI-DI: 10198459047947(each), 40198459047948(case), Lot Number: 24KBJ878; 114) DYNJ901075R, UDI-DI: 10198459047947(each), 40198459047948(case), Lot Number: 24JBD229; 115) DYNJ901075R, UDI-DI: 10198459047947(each), 40198459047948(case), Lot Number: 26CBF249; 116) DYNJ902664L, UDI-DI: 10195327537883(each), 40195327537884(case), Lot Number: 25CLA878; 117) DYNJ902664L, UDI-DI: 10195327537883(each), 40195327537884(case), Lot Number: 24ELA957; 118) DYNJ902664L, UDI-DI: 10195327537883(each), 40195327537884(case), Lot Number: 24GLA394; 119) DYNJ902664L, UDI-DI: 10195327537883(each), 40195327537884(case), Lot Number: 24ILA830; 120) DYNJ902664M, UDI-DI: 10198459349126(each), 40198459349127(case), Lot Number: 25ILA177; 121) DYNJ902664M, UDI-DI: 10198459349126(each), 40198459349127(case), Lot Number: 25KLA500; 122) DYNJ902664N, UDI-DI: 10198459675287(each), 40198459675288(case), Lot Number: 26BLA783; 123) DYNJ902879T, UDI-DI: 10195327461294(each), 40195327461295(case), Lot Number: 25BBL963; 124) DYNJ902879T, UDI-DI: 10195327461294(each), 40195327461295(case), Lot Number: 25CBI916; 125) DYNJ902879T, UDI-DI: 10195327461294(each), 40195327461295(case), Lot Number: 25CBO345; 126) DYNJ902879T, UDI-DI: 10195327461294(each), 40195327461295(case), Lot Number: 25DBF270; 127) DYNJ902879T, UDI-DI: 10195327461294(each), 40195327461295(case), Lot Number: 25FBQ497; 128) DYNJ902879T, UDI-DI: 10195327461294(each), 40195327461295(case), Lot Number: 25HBN030; 129) DYNJ902879T, UDI-DI: 10195327461294(each), 40195327461295(case), Lot Number: 25IBI197; 130) DYNJ902879T, UDI-DI: 10195327461294(each), 40195327461295(case), Lot Number: 25IBL774; 131) DYNJ902879T, UDI-DI: 10195327461294(each), 40195327461295(case), Lot Number: 25JBH529; 132) DYNJ902879T, UDI-DI: 10195327461294(each), 40195327461295(case), Lot Number: 25KBB119; 133) DYNJ902879T, UDI-DI: 10195327461294(each).40195327461295(case), Lot Number: 25KBH512; 134) DYNJ902879T, UDI-DI: 10195327461294(each), 40195327461295(case), Lot Number: 24KBJ173; 135) DYNJ902879T, UDI-DI: 10195327461294(each), 40195327461295(case), Lot Number: 24LBN817; 136) DYNJ902879T, UDI-DI: 10195327461294(each), 40195327461295(case), Lot Number: 24GBL698; 137) DYNJ902879T, UDI-DI: 10195327461294(each), 40195327461295(case), Lot Number: 24HBN007; 138) DYNJ902879T, UDI-DI: 10195327461294(each), 40195327461295(case), Lot Number: 24IBE657; 139) DYNJ902879T, UDI-DI: 10195327461294(each), 40195327461295(case), Lot Number: 24IBQ479; 140) DYNJ902879T, UDI-DI: 10195327461294(each), 40195327461295(case), Lot Number: 24JBM236; 141) DYNJ902879T, UDI-DI: 10195327461294(each), 40195327461295(case), Lot Number: 24JBQ858; 142) DYNJ902879U, UDI-DI: 10198459629525(each), 40198459629526(case), Lot Number: 25LBK265; 143) DYNJ902879U, UDI-DI: 10198459629525(each), 40198459629526(case), Lot Number: 26ABI879; 144) DYNJ902879U, UDI-DI: 10198459629525(each), 40198459629526(case), Lot Number: 26ABK604; 145) DYNJ902879U, UDI-DI: 10198459629525(each), 40198459629526(case), Lot Number: 26CBJ681; 146) DYNJ903465I, UDI-DI: 10195327678395(each), 40195327678396(case), Lot Number: 24HDA682; 147) DYNJ903465I, UDI-DI: 10195327678395(each), 40195327678396(case), Lot Number: 24JDA680; 148) DYNJ903465I, UDI-DI: 10195327678395(each), 40195327678396(case), Lot Number: 24KDB293; 149) DYNJ903465I, UDI-DI: 10195327678395(each), 40195327678396(case), Lot Number: 24LDB086; 150) DYNJ903465J, UDI-DI: 10198459255045(each), 40198459255046(case), Lot Number: 25BDB377; 151) DYNJ903465K, UDI-DI: 10198459292743(each), 40198459292744(case), Lot Number: 26DDA021; 152) DYNJ903465K, UDI-DI: 10198459292743(each), 40198459292744(case), Lot Number: 25CDA729; 153) DYNJ903465K, UDI-DI: 10198459292743(each), 40198459292744(case), Lot Number: 25EDA907; 154) DYNJ903465K, UDI-DI: 10198459292743(each), 40198459292744(case), Lot Number: 25GDB257; 155) DYNJ903465K, UDI-DI: 10198459292743(each), 40198459292744(case), Lot Number: 25JDC094; 156) DYNJ903465K, UDI-DI: 10198459292743(each), 40198459292744(case), Lot Number: 25KDA540; 157) DYNJ903465K, UDI-DI: 10198459292743(each), 40198459292744(case), Lot Number: 26ADA464; 158) DYNJ903465K, UDI-DI: 10198459292743(each), 40198459292744(case), Lot Number: 26BDA587; 159) DYNJ903822F, UDI-DI: 10195327355692(each), 40195327355693(case), Lot Number: 25BBG220; 160) DYNJ903822F, UDI-DI: 10195327355692(each), 40195327355693(case), Lot Number: 25DBB263; 161) DYNJ903822F, UDI-DI: 10195327355692(each), 40195327355693(case), Lot Number: 25EBR367; 162) DYNJ903822F, UDI-DI: 10195327355692(each), 40195327355693(case), Lot Number: 25GBI751; 163) DYNJ903822F, UDI-DI: 10195327355692(each), 40195327355693(case), Lot Number: 25HBU086; 164) DYNJ903822F, UDI-DI: 10195327355692(each), 40195327355693(case), Lot Number: 25IBH108; 165) DYNJ903822F, UDI-DI: 10195327355692(each), 40195327355693(case), Lot Number: 25JBL344; 166) DYNJ903822F, UDI-DI: 10195327355692(each), 40195327355693(case), Lot Number: 25KBD169; 167) DYNJ903822F, UDI-DI: 10195327355692(each), 40195327355693(case), Lot Number: 24KBI650; 168) DYNJ903822F, UDI-DI: 10195327355692(each), 40195327355693(case), Lot Number: 24LBP781; 169) DYNJ903822I, UDI-DI: 10198459702860(each), 40198459702861(case), Lot Number: 26CBU152; 170) DYNJ903822I, UDI-DI: 10198459702860(each), 40198459702861(case), Lot Number: 26DBI319; 171) DYNJ903822I, UDI-DI: 10198459702860(each), 40198459702861(case), Lot Number: 26CBR069; 172) DYNJ904311F, UDI-DI: 10195327297077(each), 40195327297078(case), Lot Number: 25GBM249; 173) DYNJ904311F, UDI-DI: 10195327297077(each), 40195327297078(case), Lot Number: 25GBO426; 174) DYNJ904311F, UDI-DI: 10195327297077(each), 40195327297078(case), Lot Number: 25GBW218; 175) DYNJ904311F, UDI-DI: 10195327297077(each), 40195327297078(case), Lot Number: 25LBG055; 176) DYNJ904311F, UDI-DI: 10195327297077(each), 40195327297078(case), Lot Number: 25LBG655; 177) DYNJ904311F, UDI-DI: 10195327297077(each), 40195327297078(case), Lot Number: 25LBH679; 178) DYNJ904311F, UDI-DI: 10195327297077(each), 40195327297078(case), Lot Number: 24FBP491; 179) DYNJ904311F, UDI-DI: 10195327297077(each), 40195327297078(case), Lot Number: 24HBC264; 180) DYNJ904311F, UDI-DI: 10195327297077(each), 40195327297078(case), Lot Number: 24IBI533; 181) DYNJ904311F, UDI-DI: 10195327297077(each), 40195327297078(case), Lot Number: 24JBR690; 182) DYNJ904311F, UDI-DI: 10195327297077(each), 40195327297078(case), Lot Number: 26ABD798; 183) DYNJ904311F, UDI-DI: 10195327297077(each), 40195327297078(case), Lot Number: 26CBC603; 184) DYNJ904749J, UDI-DI: 10195327511142(each), 40195327511143(case), Lot Number: 26DBI408; 185) DYNJ904749J, UDI-DI: 10195327511142(each), 40195327511143(case), Lot Number: 25ABE593; 186) DYNJ904749J, UDI-DI: 10195327511142(each), 40195327511143(case), Lot Number: 25BBG734; 187) DYNJ904749J, UDI-DI: 10195327511142(each), 40195327511143(case), Lot Number: 25EBU720; 188) DYNJ904749J, UDI-DI: 10195327511142(each), 40195327511143(case), Lot Number: 25HBI847; 189) DYNJ904749J, UDI-DI: 10195327511142(each), 40195327511143(case), Lot Number: 25IBQ209; 190) DYNJ904749J, UDI-DI: 10195327511142(each), 40195327511143(case), Lot Number: 25IBU994; 191) DYNJ904749J, UDI-DI: 10195327511142(each), 40195327511143(case), Lot Number: 25KBC533; 192) DYNJ904749J, UDI-DI: 10195327511142(each), 40195327511143(case), Lot Number: 24KBJ796; 193) DYNJ904749J, UDI-DI: 10195327511142(each), 40195327511143(case), Lot Number: 24LBP688; 194) DYNJ904749J, UDI-DI: 10195327511142(each), 40195327511143(case), Lot Number: 26ABM369; 195) DYNJ905567B, UDI-DI: 10198459151163(each), 40198459151164(case), Lot Number: 26DBE950; 196) DYNJ905567B, UDI-DI: 10198459151163(each), 40198459151164(case), Lot Number: 25ABR149; 197) DYNJ905567B, UDI-DI: 10198459151163(each), 40198459151164(case), Lot Number: 25BBE959; 198) DYNJ905567B, UDI-DI: 10198459151163(each), 40198459151164(case), Lot Number: 25CBG599; 199) DYNJ905567B, UDI-DI: 10198459151163(each), 40198459151164(case), Lot Number: 25GBG395; 200) DYNJ905567B, UDI-DI: 10198459151163(each), 40198459151164(case), Lot Number: 25HBT255; 201) DYNJ905567B, UDI-DI: 10198459151163(each), 40198459151164(case), Lot Number: 25IBK707; 202) DYNJ905567B, UDI-DI: 10198459151163(each), 40198459151164(case), Lot Number: 25KBH514; 203) DYNJ905567B, UDI-DI: 10198459151163(each), 40198459151164(case), Lot Number: 25LBU205; 204) DYNJ906102K, UDI-DI: 10198459079733(each), 40198459079734(case), Lot Number: 24KBJ996; 205) DYNJ906102L, UDI-DI: 10198459242564(each), 40198459242565(case), Lot Number: 25GBE498; 206) DYNJ906102M, UDI-DI: 10198459406409(each), 40198459406400(case), Lot Number: 25JBE315; 207) DYNJ906102M, UDI-DI: 10198459406409(each), 40198459406400(case), Lot Number: 25KBI700; 208) DYNJ906102M, UDI-DI: 10198459406409(each), 40198459406400(case), Lot Number: 25KBM827; 209) DYNJ906102M, UDI-DI: 10198459406409(each), 40198459406400(case), Lot Number: 25LBB028; 210) DYNJ906102M, UDI-DI: 10198459406409(each), 40198459406400(case), Lot Number: 26ABH106; 211) DYNJ906102M, UDI-DI: 10198459406409(each), 40198459406400(case), Lot Number: 26CBQ206; 212) DYNJ906225I, UDI-DI: 10195327658328(each), 40195327658329(case), Lot Number: 26DDA781; 213) DYNJ906225I, UDI-DI: 10195327658328(each), 40195327658329(case), Lot Number: 25BDB488; 214) DYNJ906225I, UDI-DI: 10195327658328(each), 40195327658329(case), Lot Number: 25DDA872; 215) DYNJ906225I, UDI-DI: 10195327658328(each), 40195327658329(case), Lot Number: 25DDB818; 216) DYNJ906225I, UDI-DI: 10195327658328(each), 40195327658329(case), Lot Number: 25EDB259; 217) DYNJ906225I, UDI-DI: 10195327658328(each), 40195327658329(case), Lot Number: 25FDA502; 218) DYNJ906225I, UDI-DI: 10195327658328(each), 40195327658329(case), Lot Number: 25FDA669; 219) DYNJ906225I, UDI-DI: 10195327658328(each), 40195327658329(case), Lot Number: 25FDA885; 220) DYNJ906225I, UDI-DI: 10195327658328(each), 40195327658329(case), Lot Number: 25FDB470; 221) DYNJ906225I, UDI-DI: 10195327658328(each), 40195327658329(case), Lot Number: 25GDB124; 222) DYNJ906225I, UDI-DI: 10195327658328(each), 40195327658329(case), Lot Number: 25IDA895; 223) DYNJ906225I, UDI-DI: 10195327658328(each), 40195327658329(case), Lot Number: 25KDA349; 224) DYNJ906225I, UDI-DI: 10195327658328(each), 40195327658329(case), Lot Number: 26ADA530; 225) DYNJ906225I, UDI-DI: 10195327658328(each), 40195327658329(case), Lot Number: 26CDA583; 226) DYNJ906225I, UDI-DI: 10195327658328(each), 40195327658329(case), Lot Number: 26CDB312; 227) DYNJ907400F, UDI-DI: 10195327694579(each), 40195327694570(case), Lot Number: 25BBN388; 228) DYNJ907400F, UDI-DI: 10195327694579(each), 40195327694570(case), Lot Number: 25EBA952; 229) DYNJ907400F, UDI-DI: 10195327694579(each), 40195327694570(case), Lot Number: 25EBP655; 230) DYNJ907400F, UDI-DI: 10195327694579(each), 40195327694570(case), Lot Number: 25FBG002; 231) DYNJ907400F, UDI-DI: 10195327694579(each), 40195327694570(case), Lot Number: 25GBC367; 232) DYNJ907400G, UDI-DI: 10198459440656(each), 40198459440657(case), Lot Number: 26DBM514; 233) DYNJ907400G, UDI-DI: 10198459440656(each), 40198459440657(case), Lot Number: 25HBO685; 234) DYNJ907400G, UDI-DI: 10198459440656(each), 40198459440657(case), Lot Number: 25IBB659; 235) DYNJ907400G, UDI-DI: 10198459440656(each), 40198459440657(case), Lot Number: 25IBH530; 236) DYNJ908121C, UDI-DI: 10198459024863(each), 40198459024864(case), Lot Number: 26DBP621; 237) DYNJ908121C, UDI-DI: 10198459024863(each), 40198459024864(case), Lot Number: 25CBQ702; 238) DYNJ908121C, UDI-DI: 10198459024863(each), 40198459024864(case), Lot Number: 25DBN708; 239) DYNJ908121C, UDI-DI: 10198459024863(each), 40198459024864(case), Lot Number: 25EBH917; 240) DYNJ908121C, UDI-DI: 10198459024863(each), 40198459024864(case), Lot Number: 25GBH817; 241) DYNJ908121C, UDI-DI: 10198459024863(each), 40198459024864(case), Lot Number: 25JBM106; 242) DYNJ908121C, UDI-DI: 10198459024863(each), 40198459024864(case), Lot Number: 25KBG673; 243) DYNJ908121C, UDI-DI: 10198459024863(each), 40198459024864(case), Lot Number: 26ABD610; 244) DYNJ908121C, UDI-DI: 10198459024863(each), 40198459024864(case), Lot Number: 26CBM190; 245) DYNJ908308A, UDI-DI: 10198459287213(each), 40198459287214(case), Lot Number: 25DBG598; 246) DYNJ908308A, UDI-DI: 10198459287213(each), 40198459287214(case), Lot Number: 25FBU833; 247) DYNJ908308A, UDI-DI: 10198459287213(each), 40198459287214(case), Lot Number: 25HBP099; 248) DYNJ909116D, UDI-DI: 10198459055355(each), 40198459055356(case), Lot Number: 25JBM466; 249) DYNJ9622573AP, UDI-DI: 10193489462821(each), 40193489462822(case), Lot Number: 25CBJ587; 250) DYNJ9622573AP, UDI-DI: 10193489462821(each), 40193489462822(case), Lot Number: 25DBV374; 251) DYNJ9622573AP, UDI-DI: 10193489462821(each), 40193489462822(case), Lot Number: 25FBJ227; 252) DYNJ9622573AP, UDI-DI: 10193489462821(each), 40193489462822(case), Lot Number: 25GBF521; 253) DYNJ9622573AP, UDI-DI: 10193489462821(each), 40193489462822(case), Lot Number: 25HBU890; 254) DYNJ9622573AP, UDI-DI: 10193489462821(each), 40193489462822(case), Lot Number: 25IBN994; 255) DYNJ9622573AP, UDI-DI: 10193489462821(each), 40193489462822(case), Lot Number: 25JBM121; 256) DYNJ9622573AP, UDI-DI: 10193489462821(each), 40193489462822(case), Lot Number: 25JBP614; 257) DYNJ9622573AP, UDI-DI: 10193489462821(each), 40193489462822(case), Lot Number: 25LBM800; 258) DYNJ9622573AP, UDI-DI: 10193489462821(each), 40193489462822(case), Lot Number: 24LBO075; 259) DYNJ9622573AP, UDI-DI: 10193489462821(each), 40193489462822(case), Lot Number: 24JBW896; 260) DYNJ9622573AP, UDI-DI: 10193489462821(each), 40193489462822(case), Lot Number: 24JBY038; 261) DYNJ9622573AP, UDI-DI: 10193489462821(each), 40193489462822(case), Lot Number: 24KBE874; 262) DYNJ9622573AQ, UDI-DI: 10198459682421(each), 40198459682422(case), Lot Number: 26CBR084; 263) DYNJ9718008AS, UDI-DI: 10193489263121(each), 40193489263122(case), Lot Number: 24GLB178; 264) DYNJ9718008AS, UDI-DI: 10193489263121(each), 40193489263122(case), Lot Number: 24ILA162; 265) DYNJ9718008AS, UDI-DI: 10193489263121(each), 40193489263122(case), Lot Number: 24ILA474; 266) DYNJ9718008AT, UDI-DI: 10198459133978(each), 40198459133979(case), Lot Number: 24LLA953; 267) DYNJ9718008AT, UDI-DI: 10198459133978(each), 40198459133979(case), Lot Number: 25ALA273; 268) DYNJ9718008AT, UDI-DI: 10198459133978(each), 40198459133979(case), Lot Number: 25ALA353; 269) DYNJ9718008AT, UDI-DI: 10198459133978(each), 40198459133979(case), Lot Number: 25BLA453; 270) DYNJ9718008AT, UDI-DI: 10198459133978(each), 40198459133979(case), Lot Number: 25CLA742; 271) DYNJ9718008AT, UDI-DI: 10198459133978(each), 40198459133979(case), Lot Number: 25CLA982; 272) DYNJ9718008AT, UDI-DI: 10198459133978(each), 40198459133979(case), Lot Number: 24KLA590; 273) DYNJ9718008AU, UDI-DI: 10198459291579(each), 40198459291570(case), Lot Number: 26DLA248; 274) DYNJ9718008AU, UDI-DI: 10198459291579(each), 40198459291570(case), Lot Number: 26DLB039; 275) DYNJ9718008AU, UDI-DI: 10198459291579(each), 40198459291570(case), Lot Number: 25FLA478; 276) DYNJ9718008AU, UDI-DI: 10198459291579(each), 40198459291570(case), Lot Number: 25FLA656; 277) DYNJ9718008AU, UDI-DI: 10198459291579(each), 40198459291570(case), Lot Number: 25GLA353; 278) DYNJ9718008AU, UDI-DI: 10198459291579(each), 40198459291570(case), Lot Number: 25GLA499; 279) DYNJ9718008AU, UDI-DI: 10198459291579(each), 40198459291570(case), Lot Number: 25HLA068; 280) DYNJ9718008AU, UDI-DI: 10198459291579(each), 40198459291570(case), Lot Number: 26ALA729; 281) DYNJ9718008AU, UDI-DI: 10198459291579(each), 40198459291570(case), Lot Number: 26BLA613; 282) DYNJ9718008AU, UDI-DI: 10198459291579(each), 40198459291570(case), Lot Number: 26CLA783; 283) DYNJ9855016AA, UDI-DI: 10198459687617(each), 40198459687618(case), Lot Number: 26CBD676; 284) DYNJ9855016AA, UDI-DI: 10198459687617(each), 40198459687618(case), Lot Number: 26CBT699

Devices named in this recall

36 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 99043).