Skip to content

IntelliCuff, Portable automatic cuff pressure controller recallZ-2625-2026 · Hamilton Medical AG · initiated May 22, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Hamilton Medical AG or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to reported complaints, the cuff device may alarm with a "Cuff System Leakage" error. As a result, the motor may pump continuously and fail to maintain the cuff pressure set by the user.

FDA root cause category: Process design

What the company is doing

On May 22, 2026, Hamilton Medical issued a Urgent Medical Device Recall to affected consignees. Hamilton Medical ask consignees to take the following actions: 1. Identify affected devices (consider devices in use, in stock, spare devices etc.) and make them available for exchange of cuff connectors. 2. Continue to use IntelliCuff devices according to their labelling. Mitigate potential risks by following the actions described in point B. Type of Action to mitigate the risk . 3. Please fill in and sign Customer Reply Form (page 5) and send it to your Hamilton Medical distribution partner or subsidiary. This must be conducted as quickly as possible, but no later than 30 calendar days after receiving the Field Safety Notice.

Recall details

Recall number
Z-2625-2026
Status
Open, Classified
Initiated by the company
May 22, 2026
Posted by the FDA
Jul 1, 2026
Recalling firm
Hamilton Medical AG
Quantity
962 units
Product code
BSK
FDA submissions
K150893
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, HI, IL, KY, MA, MI, ND, NE, NE, NY, OH, PA, PR, SD, TX, VA, WA, WI, WV, WY and the country of Canada.

Product description

IntelliCuff, Portable automatic cuff pressure controller REF: 951001

Code information (models, lots, serial numbers)

GTIN: 07630002800839/ Lot # range: 00001 -19732

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page