GSP Neonatal GALT Kit. Product Codes: 3303-001U, 3303-0010.… recallZ-2713-2026 · Wallac Oy · initiated May 18, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Wallac Oy or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Reduction in the on-board stability of the GSP GALT assay. This may increase the likelihood of run-QC failures and consequently delay the reporting of patient results.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Wallac Oy, a Revvity company, notified consignees on about 05/18/2026 via emailed "URGENT: MEDCIAL DEVICE RECALL" letter. Consignees were instructed to inspect inventory for the affected lots and dispose of affected lots according to local requirements. If consignees do not have another GSP Neonatal GALT kit lot at hand, the newborn screening for classical galactosemia may be continued, with following instructions: - Reconstitute GALT Substrate Reagent by adding 2.8 mL of GALT Assay Buffer and invert gently for 20-30 minutes (note: slightly extended dissolving time). - Reagent cassettes should preferably be used within 12 hours (note: reduced on-board storage time). - Load reagent cassette to GSP without additional delay after reconstitution of Substrate Reagent (one reagent cassette is sufficient for measurement of four full plates (384 wells). - If possible, prioritize GALT plates first if multiple analytes/kits are run within the same instrument. Avoid exceeding the on-board storage time of 12 hours for the reagent cassette. - Run calibration curve on each plate and evaluate the results against the plate-specific curve (note: do not use stored calibration curve). These precautions are aimed at minimizing the fluctuation of the GALT activity results between assays/plates. Consignees were also requested to complete and return the response form provided. Revvity will replace the quantity of excessive kit usage and the affected GSP Neonatal GALT kits disposed from inventory upon receipt of the response form.
Recall details
- Recall number
- Z-2713-2026
- Status
- Open, Classified
- Initiated by the company
- May 18, 2026
- Posted by the FDA
- Jul 9, 2026
- Recalling firm
- Wallac Oy
- Quantity
- 3,274 units
- Product code
- KQP
- FDA submissions
- K100101
- Distribution
- Worldwide - US Nationwide distribution in the states and territories of AL, AR, AZ, CO, FL, GA, IL, IN, KY, LA, MA, MD, MN, MO, NJ, OR, PR, SC, TN, TX, VA, WV and the countries of Austria, Belgium, Egypt, France, Germany, Hong Kong, Iraq, Italy, Netherlands, Oman, Philippines, Saudi Arabia, Singapore, Spain, Thailand, United Arab Emirates, Vietnam.
Product description
GSP Neonatal GALT Kit. Product Codes: 3303-001U, 3303-0010. in vitro diagnostic. Used in FLUORESCENT PROC. (QUAL), GALACTOSE-1-PHOSPHATE URIDYL TRANSFERASE.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
1. Product Code: 3303-001U. UDI-DI: 06438147321704. UDI-PI (PACKING LOT): (01)06438147321704(17)260531(10)763638 (1076363801); (01)06438147321704(17)260531(10)764085 (1076408501); (01)06438147321704(17)260531(10)765033 (1076503301); (01)06438147321704(17)260731(10)765231 (1076523101); (01)06438147321704(17)260731(10)765316 (1076531601); (01)06438147321704(17)261031(10)766420 (1076642001); (01)06438147321704(17)261031(10)766894 (1076689401); (01)06438147321704(17)261130(10)767873 (1076763801); (01)06438147321704(17)261130(10)768361 (1076836101); (01)06438147321704(17)261130(10)768703 (1076870301); (01)06438147321704(17)270228(10)769355 (1076935501); (01)06438147321704(17)270228(10)769542 (1076954201); (01)06438147321704(17)270228(10)769632 (1076963201); (01)06438147321704(17)270228(10)769671 (1076967101); (01)06438147321704(17)270228(10)769782 (1076978201); (01)06438147321704(17)270228(10)769871 (1076987101); (01)06438147321704(17)270228(10)769975 (1076997501); (01)06438147321704(17)270228(10)770078 (1077007801); (01)06438147321704(17)270228(10)770086 (1077008601); (01)06438147321704(17)270228(10)770351 (1077035101). 2. Product Code: 3303-0010. UDI-DI: 06438147311347. UDI-PI (PACKING LOT): (01)06438147311347(17)260531(10)763396 (1076339601); (01)06438147311347(17)260531(10)763396 (1076339602); (01)06438147311347(17)260531(10)764209 (1076420901); (01)06438147311347(17)260531(10)764209 (1076420902); (01)06438147311347(17)260731(10)765317 (1076531702); (01)06438147311347(17)260731(10)7653177 (1076531703); (01)06438147311347(17)261031(10)766378 (1076637801); (01)06438147311347(17)261031(10)766943 (1076694301); (01)06438147311347(17)261031(10)767323 (1076732301); (01)06438147311347(17)261130(10)767955 (1076795501); (01)06438147311347(17)261031(10)767956 (1076795601); (01)06438147311347(17)261130(10)769214 (1076921401); (01)06438147311347(17)261231(10)769967 (1076996701).
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- GSP Neonatal GALT kit 3303-001UWallac Oy · Fluorescent Proc. (Qual.), Galactose-1-Phosphate Uridyl Transferase1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page