Philips Avalon Fetal Monitor, FM 30 Part numberM2703A recallZ-2523-2026 · Philips North America Llc · initiated May 26, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Monitor has incorrect assembly of the speaker connector which can cause device cables to be pulled out of the housing and increase the risk of intermittent or permanent loss of speaker output.
FDA root cause category: Process control
What the company is doing
Philips issued an URGENT Medical Device Correction notice to its consignees on 5/26/2026 via USPS certified mail. The notice explained the issue and how it can be identified, potential risk to the patient should the loss of audio is not detected, and requested the following: "Actions that should be taken by the customer / user in order to prevent risks for patients or users " Ensure fetal heart rate tones are enabled on the Avalon Fetal Monitor until the planned fix/correction is implemented. " Pass this notice to all those who need to be aware within your organization or to any organization where affected product(s) have been potentially transferred. " Place this Urgent Medical Device Correction with the documentation of the Philips Avalon Fetal Monitor and associated devices in a place where it is most likely to be seen and viewed."
Recall details
- Recall number
- Z-2523-2026
- Status
- Open, Classified
- Initiated by the company
- May 26, 2026
- Posted by the FDA
- Jun 22, 2026
- Recalling firm
- Philips North America Llc
- Quantity
- 169 units
- Product code
- HGM
- FDA submissions
- K140535
- Distribution
- Worldwide distribution - US Nationwide and The countries of AE, AT, AU, BE, BG, BH, BR, CA, CH, CN, CR, CZ, DE, ES, FI, FR, GB, GR, HK, ID, IL, IN, IT, JO, JP, KR, KR, LT, LU, LV, MX, MY, NL, NZ, OM, SA, SE, SG, SI, TR, TW, ZA.
Product description
Philips Avalon Fetal Monitor, FM 30 Part numberM2703A
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
UDI/DI 00884838000414, Serial Numbers: DE79646555, DE79646616, DE79646645, DE79646647, DE79646651, DE79646697, DE79646702, DE79646746, DE79646754, DE79646757, DE79646758, DE79646762, DE79646765, DE79646775, DE79646776, DE79646783, DE79646857, DE79646891, DE79646924, DE79646750, DE79646751, DE79646756, DE79646804, DE79646592, DE79646877, DE79646644, DE79646706, DE79646723, DE79646653, DE79646766, DE79646786, DE79646824, DE79646827, DE79646848, DE79646956, DE79646709, DE79646892, DE79646724, DE79646729, DE79646732, DE79646733, DE79646927, DE79646675, DE79646820, DE79646821, DE79646837, DE79646866, DE79646869, DE79646871, DE79646875, DE79646882, DE79646683, DE79646676, DE79646640, DE79646849, DE79646952, DE79646699, DE79646694, DE79646648, DE79646701, DE79646900, DE79647027, DE79647039, DE79646654, DE79646800, DE79646604, DE79646660, DE79646829, DE79646747, DE79646606, DE79646826, DE79646784, DE79646794.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Avalon Fetal Monitor FM30 Avalon FM30 Fetal MonitorPhilips Medizin Systeme Böblingen GmbH · SYSTEM, MONITORING, PERINATAL1 recall11 deaths142 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99084).
- Z-2522-2026 · Open, ClassifiedPhilips Avalon Fetal Monitor, FM20, Part number M2702A, Part M2703A
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page