Skip to content

Advance Aluminum Chloride Gel Clear Hemostatic Gel Kit recallZ-2400-2026 · Inter-Med Llc · initiated May 27, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Inter-Med Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Some units of Aluminum Chloride Gel contain the package insert for Ferric Sulfate Gel. The incorrect insert misidentifies the product and contains different first aid instructions, most critically in that the Ferric Sulfate insert instructs users to induce vomiting in the case of ingestion, while the correct Aluminum Chloride insert instructs users not to induce vomiting.

FDA root cause category: Labeling mix-ups

What the company is doing

On 5/27/2026, firm notified consignees via email. Consignees were instructed to return all units from Lot 2026-0204 to Inter-Med, Inc. Inter-Med, Inc. will provide prepaid return shipping labels. All returned product will be replaced with correctly packaged product at no charge, or a credit may be issued at the customer's preference. Customers must also notify any downstream dental practices or end users who received product from this lot and instruct them to return product.

Recall details

Recall number
Z-2400-2026
Status
Open, Classified
Initiated by the company
May 27, 2026
Posted by the FDA
Jun 10, 2026
Recalling firm
Inter-Med Llc
Quantity
500
Product code
MVL
FDA submissions
K190220
Distribution
US Nationwide distribution in the states of Tennessee and Missouri.

Product description

Advance Aluminum Chloride Gel Clear Hemostatic Gel Kit, Catalog number 504600-28

Code information (models, lots, serial numbers)

Model No. 504600-28; UDI-DI 00371347008104; Lot Number 2026-0204