Medline Convenience Kits recallZ-2695-2026 · Medline Industries, LP · initiated May 28, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Convenience Kits contain Covidien Mon-a-Therm General Purpose Temperature Probes Size 9 Fr. Medtronic has issued a recall of specific production lots of the Covidien Mon-a-Therm general purpose temperature probe-size 9 due to an error in its distribution control process.
FDA root cause category: Under Investigation by firm
What the company is doing
An URGENT MEDICAL DEVICE RECALL notification letter dated 5/28/26 was sent to customers. REQUIRED ACTION: 1.Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected production inventory, please complete and submit the form. Website link: https://recalls.medline.comRecall Reference #: R-26-085-FGX1 Recall Code: 3.Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4.If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form ,you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-2695-2026
- Status
- Open, Classified
- Initiated by the company
- May 28, 2026
- Posted by the FDA
- Jul 8, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 1400 kits
- Product code
- CAI
- Distribution
- US Nationwide distribution.
Product description
Medline Convenience Kits: 1) ANESTHESIA KIT PEDS 90-LF, Model Number: DYNJAP10125A
Code information (models, lots, serial numbers)
1) DYNJAP10125A, UDI-DI: 10198459112911(each), 40198459112912(case), Lot Number: 24JBR072
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJAP10125AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99196).
- Z-2696-2026 · Open, ClassifiedMedline Convenience Kits: 1) OPEN HEART PEDS PACK, Model Number: DYNJ88569D; 2) ACCESSORY KIT, Model Number: DYNJ901348F
- Z-2697-2026 · Open, ClassifiedMedline Convenience Kits: 1) OPEN HEART PEDS PACK, Model Number: DYNJ88569; 2) OPEN HEART PEDS PACK, Model Number: DYNJ88569A; 3) OPEN HEART PEDS PACK, Model Number: DYNJ88569B; …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page