Skip to content

ABIOMED Impella CP Set, Pump Set Code 0048-0014 recallZ-2689-2026 · Abiomed, Inc. · initiated Jun 11, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abiomed, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.

FDA root cause category: Under Investigation by firm

What the company is doing

An URGENT MEDICAL DEVICE RECALL (CORRECTION) notice dated 06/10/2026 was mailed to consignees. The notice directs consignees to maintain awareness of this recall among practitioners and use appropriate clinical awareness, monitoring, and patient support as needed. Further clinical direction is provided in the notice. The notification is to be forwarded to users and to facilities where product was further distributed. The notice should be visible for awareness. Consignees with any questions can contact [email protected].

Recall details

Recall number
Z-2689-2026
Status
Open, Classified
Initiated by the company
Jun 11, 2026
Posted by the FDA
Jul 10, 2026
Recalling firm
Abiomed, Inc.
Quantity
35,430 units
Product code
DYB
FDA submissions
K192769, K201116
Distribution
Worldwide distribution. US Nationwide, Australia (AU), Austria (AT), Belgium (BE), Brazil (BR), Brunei (BN), Canada (CA), China (CN), Croatia (HR), Czech Republic (CZ), Denmark (DK), Finland (FI), France (FR), Germany (DE), Greece (GR), Hong Kong (HK), Ireland (IE), Israel (IL), Italy (IT), Japan (JP), Kuwait (KW), Luxembourg (LU), Malaysia (MY), Mexico (MX), Netherlands (NL), Norway (NO), Panama (PA), Poland (PL), Saudi Arabia (SA), Serbia (RS), Slovenia (SI), South Korea (KR), Spain (ES), Sweden (SE), Switzerland (CH), Taiwan (TW), Thailand (TH), United Arab Emirates (AE), and United Kingdom (GB).

Product description

ABIOMED Impella CP Set, Pump Set Code 0048-0014; containing Introducer Product Code 0052-3046. heart pump and accessories

Code information (models, lots, serial numbers)

Pump Set Code 0048-0014; Introducer Product Code 0052-3046. GTIN: 00813502012873, 00813502013344, 00813502013368, 04260113630280. All current batches in field.

Other devices under the same FDA submission

8 devices cleared under K192769, K201116 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 98893).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page