ORTHOFIX MODULSYSTEM recallZ-2731-2026 · Orthofix U.S. LLC · initiated Jun 10, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Orthofix U.S. LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Unilateral external fixation devices may be incorrectly labeled. Minirail M101 was labeled as M103, Minirail M103 was labeled as M101, and Minirail M122 was labeled as M111
FDA root cause category: Process change control
What the company is doing
On June 8, 2026, Orthofix issued a Urgent Medical Device Recall Notification, via E.Mail. Orthofix asked consignees to take the following actions: 1.Discontinue use of the affected product lots immediately. 2.Immediately examine your inventory and quarantine any affected product and return them to Orthofix. 3. Complete the attached response form and return it to Orthofix. 4. This notice should be shared with all relevant personnel within your organization and forwarded to any other organization or facility where potentially affected devices may have been transferred.
Recall details
- Recall number
- Z-2731-2026
- Status
- Open, Classified
- Initiated by the company
- Jun 10, 2026
- Posted by the FDA
- Jul 14, 2026
- Recalling firm
- Orthofix U.S. LLC
- Quantity
- 32 units
- Product code
- KTT
- FDA submissions
- K113770
- Distribution
- US: AZ CA CO CT FL GA ID IL IN LA MD MI MO MS MT NC NY OH OR SC SD TX UT VA WA WI OUS: None
Product description
Orthofix ARTICULATED MINIRAIL FIXATOR HORIZONTAL AXIS REF: M111, Unilateral orthopedic/trauma external fixation device
Code information (models, lots, serial numbers)
All Lots; GTIN# 18032568866681
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ORTHOFIX MODULSYSTEM 1Orthofix Srl · Pin, Fixation, Threaded1 recall0 reports
Product lines
Other devices under the same FDA submission
4 devices cleared under K113770 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Galaxy Fixation System 1Orthofix Srl · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component3 reports
- ORTHOFIX GALAXY FIXATION SYSTEM 1Orthofix Srl · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component2 reports
- ORTHOFIX GALAXY FIXATION SYSTEM 1Orthofix Srl · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component1 report
- ORTHOFIX GALAXY FIXATION SYSTEM 1Orthofix Srl · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 99225).
- Z-2729-2026 · Open, ClassifiedOrthofix MINIRAIL, STANDARD LENGTHENER Model/Catalog Number: M101, Unilateral orthopedic/trauma external fixation device
- Z-2730-2026 · Open, ClassifiedOrthofix MINIRAIL, SHORT, LENGTHENER REF: M103, Unilateral orthopedic/trauma external fixation device
- Z-2732-2026 · Open, ClassifiedOrthofix MINIRAIL, ARTICULATED, FIXATOR, VERTICAL AXIS REF: M122, Unilateral orthopedic/trauma external fixation device
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page