Skip to content

Zeiss Intrabeam Spherical Applicator 1.5 cm recallZ-2649-2026 · Carl Zeiss Meditec AG · initiated Jun 1, 2026

Class II Open, Classified All recalls posted in July 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Carl Zeiss Meditec AG or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Spherical applicators (radiation therapy accessory) failed to meet manufacturing specifications.

FDA root cause category: Employee error

What the company is doing

On June 12, 2026, Carl Zeiss Meditech issued a Urgent Medical Device Recall to affected consignees via FedEx Priority. Carl Zeiss ask consignees to take the following actions: 1. Immediately discontinue use of any affected applicator identified in this notice and segregate/quarantine it to prevent further use. 2. Complete, sign, and return the attached Acknowledgement and Receipt Form to CZMUI in accordance with the instructions provided on the form. 3. Once the Acknowledgement and Receipt Form has been received by CZMUI, then a member of our service team will contact you to schedule the replacement of your applicator and collect the recalled applicator from you. 4. This recall letter should be shared with anyone who needs to be aware within your organization and forwarded to any organization where potentially affected devices have been transferred.

Recall details

Recall number
Z-2649-2026
Recall class
Class II
Status
Open, Classified
Initiated by the company
Jun 1, 2026
Posted by the FDA
Jul 1, 2026
Recalling firm
Carl Zeiss Meditec AG
Quantity
13 units
Product code
JAD
FDA submissions
K121653
Distribution
Worldwide - US Nationwide distribution in the state and territories of Florida and Puerto Rico. The countries of China, Germany, Ecuador, Spain, France, Malaysia, Philippines, Poland, and Romania.

Product description

Zeiss Intrabeam Spherical Applicator 1.5 cm REF:304534-6000-557

Code information (models, lots, serial numbers)

UDI: (01)04049539101563 Batch no: A0002 Serial numbers: US - A63141, A63086 OUS (Outside the US) - A63085, A63096, A63101, A63102, A63105, A63107, A63108, A63109, A63110, A63112, A6313

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page