Emano Flow with software version v2.0.0 recallZ-2777-2026 · Emano Metrics, Inc. · initiated May 29, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Emano Metrics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Problem description/Reason for recall: During the period March 22, 2026, to May 29, 2026, recorded urination audio was captured at 48,000 Hz while the model was trained and validated on 44,100 Hz audio. The models depend on the sampling so every recording in the window was therefore processed against a time/frequency scale the models did not expect, producing systematically biased Qmax, Qmean, and voided-volume estimates. The problem usually over-reports flow rate and volume; the firm have made corrected results available to care teams.
FDA root cause category: Software Design Change
What the company is doing
On 06/10/2026, a "Medical Device Correction" Letter was emailed to customers informing them that it would found that Emano Flow software versions v2.2.0-v2.3.4 processed audio at a sample rate that did not match the sample rate expected by the flow prediction algorithms (model version v2.0.0), affecting results generated between 03/22/2026 to 05/29/2026. Customer were also informed that, the correction (updated software) was deployed on May 29, 2026. An updated customer notice was emailed on 06/23/2026. Instructions provided to Customer included: 1. Review the corrected results for patients with Emano Flow data from the affected period, prioritizing cases where the data informed a clinical decision. 2. If customers have previously downloaded patient reports generated between 03/22/2026 and 05/29/2026, replace them with the corrected versions available in the Emano Flow clinic portal. A spreadsheet with secure links to all corrected reports is attached. 3. Please reply to this email to confirm receipt of this notice.
Recall details
- Recall number
- Z-2777-2026
- Status
- Open, Classified
- Initiated by the company
- May 29, 2026
- Posted by the FDA
- Jul 22, 2026
- Recalling firm
- Emano Metrics, Inc.
- Quantity
- 685
- Product code
- EXY
- Distribution
- U.S. Nationwide distribution in the states of AR, AZ, CA, CO, FL. IA, ID, NC, NJ, PA, TX and WI.
Product description
Emano Flow with software version v2.0.0 GS1 GTIN code: 199874903484 Emano Flow is a smartphone-based acoustic uroflowmetry system. Software-only device that measures urinary flow rate and voided volume from the acoustic signal of urination. No physical components. Distributed digitally via mobile application.
Code information (models, lots, serial numbers)
Software Version - 2.0.0 GS1 GTIN code: 199874903484 UDI code: GUDID registration is in progress
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Emano FlowEmano Metrics Inc. · Uroflowmeter1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page