Halyard EYE PACK. Model/Catalog Number: LVEY27 recallZ-2763-2026 · AVID Medical, Inc. · initiated May 7, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from AVID Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Incorrect concentration of Betadine solution contained within certain AVID convenience kits.
FDA root cause category: Process control
What the company is doing
AVID Medical notified the single consignee on about 05/07/2026 via email. The consignee was instructed to identify, segregate, and quarantine all impacted products, add the provided warning labels to affected AVID kits to ensure all users are aware of the issue, and discard all impacted components upon opening the Eye Pack. Customers and end users were instructed that if they received affected products from a distributor, to please contact that distributor of sales representatives with questions. Distributors were instructed that if any of the affected kits identified were further distributed, to ensure all end-users are appropriately notified and maintain records of effectiveness. Recipients were requested to complete and return the provided Recall Response Form by 05/22/2026.
Recall details
- Recall number
- Z-2763-2026
- Status
- Open, Classified
- Initiated by the company
- May 7, 2026
- Posted by the FDA
- Jul 20, 2026
- Recalling firm
- AVID Medical, Inc.
- Quantity
- 260 kits
- Product code
- OJK
- Distribution
- US Nationwide distribution in the state of California.
Product description
Halyard EYE PACK. Model/Catalog Number: LVEY27 Convenience kit.
Code information (models, lots, serial numbers)
Model/Catalog Number: LVEY27. UDI-DI: 10809160499528. Lot Numbers (Expiration Date): 1680076 (12/31/2026), 1680590 (6/30/2027).
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Halyard LVEY27_0003Avid Medical, Inc. · Eye tray1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page