Skip to content

Brand Name: Xhibit recallZ-2916-2026 · Spacelabs Healthcare, Inc. · initiated Jun 4, 2026

Class II Open, Classified All recalls posted in August 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Spacelabs Healthcare, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to off-the-shelf display port component that has audio/video failures

FDA root cause category: Unknown/Undetermined by firm

What the company is doing

On 06/04/2026, an "URGENT MEDICAL DEVICE CORRECTION" Letter was emailed to customers informing them that the use of DisplayPort Splitter (PN 010-1975-00) may potentially result in a loss of audio, video, or both audio and video at the local Xhibit Central Station and/or extended displays. Customers are instructed to: 1. Discontinue use Discontinue use of the DisplayPort Splitter, Model number 010-1975-00, or any unapproved splitters that may be used. 2. Notify Inform all relevant personnel within your organization, including any facilities to which you have distributed the affected products. 3. Acknowledge Confirm receipt of this notification by scanning the QR code below and completing the form. Include the model number of splitter that is utilized at the facility, and if it was purchased through Spacelabs Healthcare. (PDF version of the acknowledgement form can be provided upon request) 4. Disposal Dispose of all affected components. 5. Replace Replace the affected DisplayPort Splitters following the proper installation requirements outlined in section 6 and 7 of CSN 077-0722-00. 6. Fill out the Customer Response Form using the link or QR code provided below. Spacelabs has partnered with IQVIA MedTech to coordinate the recall response process. " Link: https://iqvia-response.my.site.com/mt/fca?cid=spacelabsxhibit-av For questions or support - contact Customer Support at 1-800-522-7025 or via website https://spacelabshealthcare.com/support/

Recall details

Recall number
Z-2916-2026
Recall class
Class II
Status
Open, Classified
Initiated by the company
Jun 4, 2026
Posted by the FDA
Aug 7, 2026
Recalling firm
Spacelabs Healthcare, Inc.
Quantity
6074
Product code
MHX
FDA submissions
K122146
Distribution
Worldwide - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WY and the countries of CANADA, FRANCE, ITALY, MALAYSIA, THAILAND.

Product description

Brand Name: Xhibit Product Name: Xhibit Central Station Model/Catalog Number: 96102 Software Version: N/A Product Description: The Spacelabs Healthcare (Spacelabs) Xhibit Central Station, Model 96102. Xhibit offers remote surveillance of patient data for those patients connected to a Spacelabs Healthcare bedside monitors and telemetry systems. The DisplayPort audio/video splitter allows users to extend the audio/video source from the Xhibit to additional external displays or monitors. This extension of audio/video is for viewing only and does not allow interaction with the Xhibit. Component: The DisplayPort audio/video splitter is a 3rd party external accessory to Xhibit.

Code information (models, lots, serial numbers)

Model/Catalog Number: 96102 DisplayPort Splitter Part Number: 010-1975-00 UDI-DI codes: 1084152212706 10841522100345 Serial Numbers: Pending

Other devices under the same FDA submission

3 devices cleared under K122146 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page