BIOFIRE Respiratory Panel 2.1 (RP2.1) REF 423742 recallZ-2936-2026 · BioFire Diagnostics, LLC · initiated Jun 24, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from BioFire Diagnostics, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to increase in false negative results
FDA root cause category: Under Investigation by firm
What the company is doing
On 06/24/2026, an "URGENT MEDICAL DEVICE CORRECTION" Letter was emailed to customers informing them that bioMrieux received complaints of false-negative results associated with BIOFIRE RP2.1 kit lot 1872125 / pouch lot 3VLB25. An internal investigation confirmed the presence of incorrect PCR1 primers in a subset of pouches within BIOFIRE RP2.1 kit lot 1872125 introduced during reagent manufacturing. This field safety notice was issued because the incorrect PCR1 primers will lead to pouch pathogen assay negative results with passing control assays. Customers are instructed to: "Examine your inventory for the lot identified in this field safety notice. "Cease use and discard any remaining product from this lot in your possession. "Confirm the quantity of kits that have been destroyed, previously used, or returned on the attached Acknowledgement of Receipt Form. "If you have further distributed this product, please identify any recipients and notify them at once. "Complete the accompanying Acknowledgement of Receipt Form and return to bioMrieux so that bioMrieux may acknowledge your receipt of this notification. "bioMerieux will be providing pouch credit for affected products. For Questions - contact local Customer Service Representative at (800) 682-2666. For non-urgent inquiries, contact the Technical Support Center by email at [email protected].
Recall details
- Recall number
- Z-2936-2026
- Status
- Open, Classified
- Initiated by the company
- Jun 24, 2026
- Posted by the FDA
- Aug 12, 2026
- Recalling firm
- BioFire Diagnostics, LLC
- Quantity
- 2,370 pouches (79 kits)
- Product code
- QOF
- Distribution
- U.S. Nationwide distribution in the states of AL, AR, IA, IL, KS, LA, MN, MO, TX, and WI.
Product description
BIOFIRE Respiratory Panel 2.1 (RP2.1) REF 423742 UDI code: 00815381020529 PCR-based multiplexed nucleic acid test, in vitro diagnostic
Code information (models, lots, serial numbers)
Part Number: 423742 UDI-DI code: 00815381020529 Pouch Lot: 3VLB25 Kit Lot: 1872125
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- BioFire® Respiratory Panel 2.1 423742Biofire Diagnostics, LLC · Multi-Target Respiratory Specimen Nucleic Acid Test Including Sars-Cov-2 And Other Microbial Agents3 recalls4 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page