Brand Name: OXOS Medical, Inc recallZ-2857-2026 · Oxos Medical Inc · initiated Jun 11, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Oxos Medical Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to a defective circuit board that overheats
FDA root cause category: Process change control
What the company is doing
On 06/12/2026, an "URGENT: Medical Device Recall - Limited to Specific MC2 Units" letter was emailed to customers informing them that, due to faulty circuit boards limited units may experience localized overheating. As a result the firm released a software update (version 5.7.0) which will place a unit in a safe state if boards fail. Customers are instructed to: 1. Update units to software version 5.7.0 via over-air updates. Note, it is important to apply the software update to continue using the MC2 units 2. In addition to the software update, OXOS will replace the defective components with the cassette as part of a scheduled procedure. The MC2 must be returned to OXOS headquarters for replacement of affected components. Servicing will coordinate with customers to minimized any inconvenience. 3. Customer need to acknowledge receipt of the recall communication and complete the software update by calling OXOS Support Team at 866-749-4229 to schedule a time to perform the updated by emailing support#oxos.com, or accepting the update on their devices. For questions or assistance - contact Support Team at 866-749-4229 or email [email protected].
Recall details
- Recall number
- Z-2857-2026
- Status
- Open, Classified
- Initiated by the company
- Jun 11, 2026
- Posted by the FDA
- Aug 3, 2026
- Recalling firm
- Oxos Medical Inc
- Quantity
- 77
- Product code
- OWB
- FDA submissions
- K252068
- Distribution
- US Nationwide distribution in the states of FL, GA, IN, SC, TX, VA.
Product description
Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System Model/Catalog Number: MC2, MC2.1 Software Version: 5.7.0 Product Description: The MC2 Portable X-Ray System ( MC2 System or MC2 ) is a portable and handheld x-ray system designed to aid clinicians with point of care visualization through diagnostic x-rays of extremities and shoulders. The device allows a clinician to select desired technique factors best suited for their patient anatomy. The MC2 consists of two major system components: the emitter and the cassette. The MC2 emitter and cassette are battery-powered and can be charged with either a wired or wireless charger. The system is intended to interface wirelessly to an external tablet or display when used with the OXOS Device App. The MC2 utilizes an infrared Tracking System and Viewfinder that allows the operator to position the emitter within a safe source-to-image distance (SID) and allows the operator to visualize the anatomy and x-ray field projection prior to initiating an x-ray acquisition. The MC2 also utilizes a LIDAR system to ensure patient safety by maintaining a safe source-to-skin distance. Component: The MC2 Portable X-Ray System is not a component.
Code information (models, lots, serial numbers)
Lot Code: MC2; UDI 00860008892782; Serial Numbers 1523, 1527, 1488, 1528, 1529, 1544 MC2.1; UDI 00850079606043; Serial Numbers: 1461, 1462, 1547 Full UDI for affected devices: (01)00860008892782(11)260107(21)1523 (01)00860008892782(11)260129(21)1527 (01)00860008892782(11)260109(21)1488 (01)00860008892782(11)260123(21)1528 (01)00860008892782(11)260130(21)1529 (01)00860008892782(11)260227(21)1544 (01)00850079606043(11)260407(21)1461 (01)00850079606043(11)260407(21)1462 (01)00850079606043(11)260408(21)1547
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- MC2 Portable X-ray System MC2.1OXOS Medical, Inc. · Interventional Fluoroscopic X-Ray System1 recall0 reports
- MC2 Portable X-ray SystemOXOS Medical, Inc. · Interventional Fluoroscopic X-Ray System1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page