ProNephro AKI (NGAL) Reagent Kit (100 tests). IVD Acute… recallZ-2937-2026 · Bioporto Diagnostics A/S · initiated May 18, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Bioporto Diagnostics A/S or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The use of a non-validated rework process for assay reagents which could lead to missed diagnosis of renal disease, inappropriate treatment decisions, or inappropriate monitoring of renal disease in intended users.

FDA root cause category: Nonconforming Material/Component

What the company is doing

A Quality Notification dated 6/3/26 was sent to consignees via email notifying them of the recall. Consignees in the US are to coordinate replacement of devices. ROW Consignees are to scrap unopened kits in accordance with local requirements and request replacements.

Recall details

Recall number
Z-2937-2026
Status
Open, Classified
Initiated by the company
May 18, 2026
Posted by the FDA
Aug 12, 2026
Recalling firm
Bioporto Diagnostics A/S
Quantity
400 units
Product code
PIG
FDA submissions
K232761
Distribution
Worldwide - The affected product distribution included the United States and the countries of European Union, Chile, and Hong Kong. No affected product was distributed to Canada; however, Health Canada was notified because the product is registered in Canada.

Product description

ProNephro AKI (NGAL) Reagent Kit (100 tests). IVD Acute Kidney Injury Test System.

Code information (models, lots, serial numbers)

Model number: ST101UB / 08787174 190; GTIN code: 5711435000156; Lot-number: 04909125 and 04912125.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.