M5 Corpus Power Wheelchair REF,120032 recallZ-2805-2026 · Permobil Inc · initiated Jun 19, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Permobil Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

This recall is being reported retroactively. Power Platform firmware is missing specific drive inhibit logic. This issue may allow the chair to drive under restricted conditions, and may create an unstable condition in certain circumstances.

FDA root cause category: Process design

What the company is doing

On June 10, 2025 Permobil issued a technical bulletin (recall notice) via Email to affected consignees. Permobil informed consignees of the following: 1. Cellular connection is required to receive and install the firmware update. 2. It is important to not use the wheelchair when the firmware update is in process (as indicated on the display of the input device). 3. Specific Remote Update Instructions are available if needed in addition to the on-screen information. 4. A confirmation of the completed update will be shown on the joystick display when the update successfully completes. 5. If you have any questions or concerns, please contact the Permobil Customer Experience team at (800) 736-0925. 1.

Recall details

Recall number
Z-2805-2026
Status
Open, Classified
Initiated by the company
Jun 19, 2025
Posted by the FDA
Jul 24, 2026
Recalling firm
Permobil Inc
Quantity
715 wheelchairs
Product code
ITI
FDA submissions
K123290
Distribution
Worldwide - US: Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CANADA.

Product description

M5 Corpus Power Wheelchair REF,120032

Code information (models, lots, serial numbers)

Power Platform firmware versions 1.27 through 1.57 UDI-DI 17330818334562

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

2 devices cleared under K123290 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page