Phantom Hindfoot TTC/TC Nail System recallZ-2758-2026 · Paragon 28, Inc. · initiated Jun 17, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Paragon 28, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Paragon 28 received two (2) complaints were received involving breaches in the sterile barrier of the pouch packaging.
FDA root cause category: Under Investigation by firm
What the company is doing
Paragon 28 issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 6/22/2026 via email and courier. The notice explained the issue, risks associated with the use of the affected product, and requested the following: "Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, immediately locate and quarantine affected product in your inventory. a. Your Paragon 28 sales representative may remove and return the affected product from your facility on your behalf. b. Alternatively, you may directly return all affected product from your facility. i. Complete Attachment 1 Certificate of Acknowledgement for each return and send to [email protected]. ii. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. iii. Mark RECALL on the outside of the returned cartons 3. If the product has been further distributed, provide your customers with this notice and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call 888-728-1888. Alternatively, your questions may be emailed to [email protected]. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. 4. Retain a copy of
Recall details
- Recall number
- Z-2758-2026
- Status
- Open, Classified
- Initiated by the company
- Jun 17, 2026
- Posted by the FDA
- Jul 17, 2026
- Recalling firm
- Paragon 28, Inc.
- Quantity
- 514 units
- Product code
- HXI
- Distribution
- US nationwide, Australia, Canada, The Netherlands, and New Zealand
Product description
P28 PHANTON INTRAMEDULLARY NAIL SYSTEM, Phantom Hindfoot TTC/TC Ball Tipped Guide Rod, SIZE: 0.30 mm, LENGTH: 800 mmm, ITem number P31-955-1800-S; provides stabilization of the hindfoot and ankle.
Code information (models, lots, serial numbers)
UDI (01)00889795061577(17)300122(10)501167202225, Lot Number 501167202225; UDI (01)00889795061577(17)300122(10)501167302225, Lot Number 501167302225; UDI (01)00889795061577(17)300527(10)501209514725, Lot Number 501209514725; UDI (01)00889795061577(17)301006(10)501259627925, Lot Number 501259627925; UDI (01)00889795061577(17)301118(10)501275832225, Lot Number 501275832225.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Phantom Hindfoot TTC/TC Nail System P31-955-1800-SParagon 28, Inc. · PASSER, WIRE, ORTHOPEDIC1 recall4 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page