Skip to content

Medline Kits containing Zimmer Biomet mixing bowls and… recallZ-2737-2026 · Medline Industries, LP · initiated Jun 10, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Kits contain Zimmer Biomet mixing bowls and spatulas, which were recalled due to seal nonconformities which present risk of sterility breach. Health consequences from use of product include clinically insignificant extension of surgery to find another readily available device.

FDA root cause category: Under Investigation by firm

What the company is doing

On June 10, 2026, the firm began notifying customers via First-Class Mail and email. Customers were instructed to request stickers to over-label the affected convenience kits with a label indicating to remove the affected component and replace it with product from supply. No kits are to be returned.

Recall details

Recall number
Z-2737-2026
Status
Open, Classified
Initiated by the company
Jun 10, 2026
Posted by the FDA
Jul 15, 2026
Recalling firm
Medline Industries, LP
Quantity
163 kits
Product code
OJH
Distribution
US Nationwide distribution in the states of CA and UT.

Product description

Medline Kits containing Zimmer Biomet mixing bowls and spatulas TOTAL KNEE PACK-LF DYNJ61651J

Code information (models, lots, serial numbers)

UDI-DI 10195327388447 Kit Lot 26DME172

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 99318).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page